FULL-ENERGY PEAK EFFICIENCY OF ASYMMETRICAL POLYHEDRON GERMANIUM DETECTOR

被引:8
|
作者
Abbas, Mahmoud I. [1 ]
Yoseph, Sara [1 ]
El-Khatib, Ahmed M. [1 ]
Badawi, Mohamed S. [1 ,2 ]
Gouda, Mona M. [1 ]
Thabet, Abouzeid A. [3 ]
机构
[1] Alexandria Univ, Fac Sci, Phys Dept, Alexandria, Egypt
[2] Beirut Arab Univ, Fac Sci, Dept Phys, Beirut, Lebanon
[3] Pharos Univ Alexandria, Fac Allied Med Sci, Dept Med Equipment Technol, Alexandria, Egypt
来源
关键词
polyhedron detector; full-energy peak efficiency; gamma-ray; GAMMA-RAY DETECTORS; EXTENDED SOURCES; NUMERICAL-METHOD; HPGE DETECTORS; CALIBRATION; CALCULATE; ARRAYS;
D O I
10.2298/NTRP1802150A
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
To understand the nuclear structure for most elements, it is essential to investigate the nuclear excitations by using high precision gamma-ray spectroscopy in which intensive measurements should be carried out. This is becoming a new challenge for the radiation scientific community nowadays, where the instrumentations and technical advances must be developed to be used in a wide range of applications. To discover the weakest nuclear reaction, the maximum probability of the detection system of the total energy of any released individual photon must be determined. In this work, a new mathematical method to calculate the absolute full-energy peak efficiency of asymmetrical polyhedron germanium detector is presented. This type of detector can be arranged in array, forming "complex detectors of encapsulated germanium crystals", with the solid angle reaching 82 % of total solid angle coverage, i.e., with the highest possible efficiency and with a good quality of spectral response. In addition, the photon path length was enclosed in the mathematical method to determine its attenuation through different materials such as, the detector active medium and any other material in-between source-detector system during the measuring process. The comparison between the efficiency calculated in this work and that of the published Monte Carlo simulation showed a good agreement and a small variation. However, the method discussed in the current work can be useful in nuclear safeguards, in overcoming the huge difficulties in identification of the energy range of radioactive isotopes and their quantities in nuclear waste.
引用
收藏
页码:150 / 158
页数:9
相关论文
共 50 条
  • [1] Methods and software for predicting germanium detector absolute full-energy peak efficiencies
    K. R. Jackman
    S. R. Biegalski
    Journal of Radioanalytical and Nuclear Chemistry, 2009, 279
  • [2] Methods and software for predicting germanium detector absolute full-energy peak efficiencies
    Jackman, K. R.
    Biegalski, S. R.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2009, 279 (01) : 355 - 360
  • [3] HPGe detector full-energy peak efficiency calculation for inhomogeneously activated samples
    Gacnik, Lojze
    Jacimovic, Radojko
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2017, 311 (02) : 1315 - 1321
  • [4] HPGe detector full-energy peak efficiency calculation for inhomogeneously activated samples
    Lojze Gačnik
    Radojko Jaćimović
    Journal of Radioanalytical and Nuclear Chemistry, 2017, 311 : 1315 - 1321
  • [5] Absolute full-energy peak efficiency calibration of a Clover-BGO detector system
    Elekes, Z
    Belgya, T
    Molnár, GL
    Kiss, AZ
    Csatlós, M
    Gulyás, J
    Krasznahorkay, A
    Máté, Z
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 503 (03): : 580 - 588
  • [6] Investigation hybrid MCNP/Angle model for calculating the absolute full-energy peak efficiency of HPGe detector
    Salman, A.
    Ahmed, Z.
    Allam, Kh. A.
    El-Sharkawy, S.
    APPLIED RADIATION AND ISOTOPES, 2019, 150 : 57 - 62
  • [7] Full-Energy peak efficiency of an NaI(Tl) detector with coincidence summing correction showing the effect of the source-to-detector distance
    El-Khatib, Ahmed M.
    Salem, Bohaysa A.
    Badawi, Mohamed S.
    Gouda, Mona M.
    Thabet, Abouzeid A.
    Abbas, Mahmoud I.
    CHINESE JOURNAL OF PHYSICS, 2017, 55 (02) : 478 - 489
  • [8] Full-Energy Peak Efficiency Calibration of a Metallic Magnetic Calorimeter Detector for Photon Spectrometry Below 100 keV
    Riham Mariam
    Matias Rodrigues
    Martin Loidl
    Journal of Low Temperature Physics, 2018, 193 : 1269 - 1275
  • [9] HPGe detector absolute full-energy peak efficiency calibration including coincidence correction for circular disc sources
    Abbas, Mahmoud I.
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2006, 39 (18) : 3952 - 3958
  • [10] FULL-ENERGY PEAK EFFICIENCY CALIBRATION OF A WELL-TYPE GE(LI) GAMMA-RAY DETECTOR
    VERPLANKE, JC
    NUCLEAR INSTRUMENTS & METHODS, 1971, 96 (04): : 557 - +