Comparison of Visante and slit-lamp anterior segment optical coherence tomography in imaging the anterior chamber angle

被引:31
|
作者
Sakata, L. M.
Wong, T. T. L.
Wong, H-T [2 ]
Kumar, R. S.
Htoon, H. M.
Aung, H. T.
He, M. [3 ]
Aung, T. [1 ,4 ]
机构
[1] Singapore Natl Eye Ctr, Singapore Eye Res Inst, Singapore 168751, Singapore
[2] Tan Tock Seng Hosp, Singapore, Singapore
[3] Zhongshan Ophthalm Ctr, Guangzhou, Guangdong, Peoples R China
[4] Natl Univ Singapore, Yong Loo Lin Sch Med, Singapore 117595, Singapore
关键词
angle closure glaucoma; angle closure; closed angle; optical coherence tomography; ULTRASOUND BIOMICROSCOPY; WIDTH;
D O I
10.1038/eye.2009.141
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aims To compare the diagnostic performance of two anterior segment optical coherence tomography (AS-OCT) devices in assessing the anterior chamber angle (ACA). Methods Visante-OCT and slit-lamp-OCT (SL-OCT) were performed on 101 patients by a single operator. The AS-OCT images were processed by customised 'dewarping' software and assessed by two glaucoma specialists masked to clinical findings. A closed ACA was defined by the presence of contact between the iris and angle anterior to the scleral spur. Measurements of the ACA, anterior chamber depth (ACD), and pupil diameter were analysed. Gonioscopy was performed by another examiner masked to AS-OCT findings. Results Qualitative analysis could be carried out in 83 (83%) eyes and quantitative analysis in 61 (60%) eyes. A closed angle in at least one quadrant of the eye was observed in 30 eyes with gonioscopy; Visante-OCT imaging identified 29 of 30 (97%) and SL-OCT imaging identified 27 of 30 (90%) of these eyes (P = 0.50, McNemar test). Visante-OCT detected more eyes with at least one closed quadrant than SL-OCT (55 vs 46 eyes, respectively, P = 0.01). Overall, SL-OCT had better agreement with gonioscopy than with Visante-OCT. Both AS-OCTs showed good agreement for ACD measurements; however, SL-OCT tended to provide consistently higher ACA measurements and smaller pupil diameters than did Visante-OCT. Conclusions Both AS-OCT devices detected most of the eyes with closed ACA on gonioscopy. However, Visante-OCT detected more closed ACAs than did SL-OCT. The better agreement between SL-OCT and gonioscopy is likely because of the use of visible light during both examinations. The ACA measurements obtained with each device are not interchangeable. Eye (2010) 24, 578-587; doi:10.1038/eye.2009.141; published online 12 June 2009
引用
收藏
页码:578 / 587
页数:10
相关论文
共 50 条
  • [31] Evaluation of anterior segment parameters with two anterior segment optical coherence tomography systems: Visante and Casia, in primary angle closure disease
    Angmo, Dewang
    Singh, Reena
    Chaurasia, Shweta
    Yadav, Suresh
    Dada, Tanuj
    [J]. INDIAN JOURNAL OF OPHTHALMOLOGY, 2019, 67 (04) : 500 - 504
  • [32] Real time optical coherence tomography of the anterior segment using hand-held and slit-lamp adapted systems.
    Radhakrishnan, S
    Patil, CA
    Rollins, AM
    Reeves, D
    Lass, JH
    Bardenstein, DS
    Westphal, V
    Roth, JE
    Izatt, JA
    [J]. COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICAL SCIENCE AND CLINICAL APPLICATIONS VI, 2002, 4619 : 227 - 229
  • [33] Measuring Anterior Chamber Flare using Novel Slit-Lamp Imaging Technique
    Halim, Muhammad
    Chakravarthy, Adithi
    Onghanseng, Neil
    Hassan, Muhammad
    Ormaechea, Maria
    Hasanreisoglu, Murat
    Matsumiya, Wataru
    Uludag, Gunay
    Tran, Anh
    Sepah, Yasir
    Subramaniam, Mahadevan
    Nguyen, Quan
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2021, 62 (08)
  • [34] Measuring Anterior Chamber Flare using Novel Slit-Lamp Imaging Technique
    Halim, Muhammad Sohail
    Chakravarthy, Adithi Deborah
    Onghanseng, Neil L.
    Hassan, Muhammad
    Ormaechea, Maria Soledad
    Huynh Van
    Hasanreisoglu, Murat
    Hien Luong Doan
    Uludag, Gunay
    Anh Ngoc Tram Tran
    Sepah, Yasir Jamal
    Subramaniam, Mahadevan
    Quan Dong Nguyen
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (07)
  • [35] Anterior segment optical coherence tomography for imaging the anterior chamber after laser peripheral iridotomy
    Memarzadeh, Farnaz
    Li, Yan
    Chopra, Vikas
    Varma, Rohit
    Francis, Brian A.
    Huang, David
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 143 (05) : 877 - 879
  • [36] Comparing Gonioscopy With Visante and Cirrus Optical Coherence Tomography for Anterior Chamber Angle Assessment in Glaucoma Patients
    Doss, Mallika K.
    Blieden, Lauren S.
    Lee, David A.
    Bell, Nicholas P.
    Feldman, Robert M.
    [J]. JOURNAL OF GLAUCOMA, 2017, 26 (03) : E125 - E125
  • [37] Comparing Gonioscopy With Visante and Cirrus Optical Coherence Tomography for Anterior Chamber Angle Assessment in Glaucoma Patients
    Hu, Cindy X.
    Mantravadi, Anand
    Zangalli, Camila
    Ali, Mohsin
    Faria, Bruno M.
    Richman, Jesse
    Wizov, Sheryl S.
    Razeghinejad, M. Reza
    Moster, Marlene R.
    Katz, L. Jay
    [J]. JOURNAL OF GLAUCOMA, 2016, 25 (02) : 177 - 183
  • [38] Effect of Body Position on Anterior Chamber Angle Dimensions Assessed by Anterior Segment Optical Coherence Tomography
    Penteado, Rafaella
    Xu, Benjamin
    Yang, Diya
    Proudfoot, James
    Susanna, Remo
    Weinreb, Robert N.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [39] Optical coherence tomography in anterior segment imaging
    Kalev-Landoy, Maya
    Day, Alexander C.
    Cordeiro, M. Francesca
    Migdal, Clive
    [J]. ACTA OPHTHALMOLOGICA SCANDINAVICA, 2007, 85 (04): : 427 - 430
  • [40] Comparison of Scheimpflug imaging and spectral domain anterior segment optical coherence tomography for detection of narrow anterior chamber angles
    Grewal, D. S.
    Brar, G. S.
    Jain, R.
    Grewal, S. P. S.
    [J]. EYE, 2011, 25 (05) : 603 - 611