Fluid composition influences wear testing of artificial knee implants

被引:0
|
作者
Schwenke, T. [1 ]
Wimmer, M. A. [1 ]
Kaddick, C. [1 ]
机构
[1] Rush Univ, Med Ctr, Chicago, IL 60612 USA
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Multi-axial simulator testing is a standard tool in development and evaluation of artificial hip and knee joints. Close to in-vivo conditions are desirable in terms of implant set-up, input parameters, and environment. Thus, specimens are subjected to loads and displacements measured for patients with artificial joints, and a lubrication medium is provided to simulate human synovial fluid. Implant wear is then commonly assessed through periodic gravimetric measurements. However, ultra-high weight polyethylene (UHMWPE), the typical material for tibial plateaus, has been shown to gain weight through fluid absorption, which influences the results of such tests. A recent study performed on a 4-DOF knee joint simulator reveals that composition variations in the lubricating medium highly influences fluid uptake of UHMWPE, and alters the wear behavior of tested specimen. Two lubricants, both within boundaries of ISO and ASTM testing standards but with different compositions, have been compared. They showed significant deviations in fluid soak and wear results. It is suggested that wear measurements are only comparable if testing parameters are kept in very tight tolerances and identical lubrication media are used.
引用
收藏
页码:173 / 177
页数:5
相关论文
共 50 条
  • [41] Nano-hardness Testing of Wear Particles in Sheep Knee Joints
    Allen, Christopher
    Peng, Zhongxiao
    Yin, Ling
    Carmichael, Emma
    PRICM 7, PTS 1-3, 2010, 654-656 : 2253 - 2256
  • [42] Understanding the differences in wear testing method standards for total knee replacement
    Abdelgaied, A.
    Fisher, J.
    Jennings, L. M.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2022, 132
  • [43] Methodology for long-term wear testing of total knee replacements
    Walker, PS
    Blunn, GW
    Perry, JP
    Bell, CJ
    Sathasivam, S
    Andriacchi, TP
    Paul, JP
    Haider, H
    Campbell, PA
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2000, (372) : 290 - 301
  • [44] Tribological performance of the biological components of synovial fluid in artificial joint implants
    Ghosh, Subir
    Choudhury, Dipankar
    Roy, Taposh
    Moradi, Ali
    Masjuki, H. H.
    Pingguan-Murphy, Belinda
    SCIENCE AND TECHNOLOGY OF ADVANCED MATERIALS, 2015, 16 (04)
  • [45] OpenSim as a preliminary kinematic testing platform for the development of total knee arthroplasty implants
    Chan, Hao Yang
    Walker, Peter S.
    JOURNAL OF BIOMECHANICS, 2018, 76 : 53 - 60
  • [46] Analysis on Wear Phenomenon of Artificial Knee Joint Based on FEM and Mechanical Test
    Yeo-Kyung Kang
    Mun-Hee Kim
    Jae-Won Kim
    Trinh Ngoc Tien
    Do-hyung Lim
    Heoung-Jae Chun
    International Journal of Precision Engineering and Manufacturing, 2018, 19 : 1211 - 1217
  • [47] Wear Debris Characterization and Corresponding Biological Response: Artificial Hip and Knee Joints
    Nine, Md J.
    Choudhury, Dipankar
    Hee, Ay Ching
    Mootanah, Rajshree
    Abu Osman, Noor Azuan
    MATERIALS, 2014, 7 (02) : 980 - 1016
  • [48] Analysis on Wear Phenomenon of Artificial Knee Joint Based on FEM and Mechanical Test
    Kang, Yeo-Kyung
    Kim, Mun-Hee
    Kim, Jae-Won
    Trinh Ngoc Tien
    Lim, Do-hyung
    Chun, Heoung-Jae
    INTERNATIONAL JOURNAL OF PRECISION ENGINEERING AND MANUFACTURING, 2018, 19 (08) : 1211 - 1217
  • [49] The influence of femoral condylar lift-off on the wear of artificial knee joints
    Jennings, L. M.
    Bell, C. J.
    Ingham, E.
    Komistek, R. D.
    Stone, M. H.
    Fisher, J.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H3) : 305 - 314
  • [50] In vitro response of the natural cadaver knee to the loading profiles specified in a standard for knee implant wear testing
    Sutton, Levi G.
    Werner, Frederick W.
    Haider, Hani
    Hamblin, Tracy
    Clabeaux, Jonathan J.
    JOURNAL OF BIOMECHANICS, 2010, 43 (11) : 2203 - 2207