Intermediate analysis of magnesium alloy covered stent for a lateral aneurysm model in the rabbit common carotid artery

被引:5
|
作者
Cui, Hong-kai [1 ,2 ]
Li, Fen-Bao [1 ,2 ]
Guo, Ying-Chang [1 ,2 ]
Zhao, Yong-Li [1 ,2 ]
Yan, Rui-Fang [1 ,2 ]
Wang, Wu [3 ,4 ]
Li, Yong-Dong [1 ,2 ,3 ,4 ]
Wang, Yong-Li [3 ,5 ]
Yuan, Guang-Yin [6 ,7 ,8 ]
机构
[1] Xinxiang Med Univ, Affiliated Hosp 1, Dept Intervent Radiol, Xinxiang, Henan, Peoples R China
[2] 88 Jian Kang Rd, Jian 453100, Henan, Peoples R China
[3] Shanghai Jiao Tong Univ, Affiliated Peoples Hosp 6, Dept Intervent Radiol, Fengxian Hosp, Shanghai 200233, Peoples R China
[4] 600 Yi Shan Rd, Shanghai 200233, Peoples R China
[5] 6600 Nan Feng Rd, Shanghai 210499, Peoples R China
[6] Shanghai Jiao Tong Univ, Natl Engn Res Ctr Light Alloy Net Forming, Shanghai 200240, Peoples R China
[7] Shanghai Jiao Tong Univ, State Key Lab Met Matrix Composite, Shanghai 200240, Peoples R China
[8] 800 Dong Chuan Rd, Shanghai 200240, Peoples R China
基金
中国国家自然科学基金;
关键词
Mg alloy stent; Common carotid artery; Degradation; Intravascular ultrasound (IVUS); Covered stent; GUGLIELMI DETACHABLE COILS; UNRUPTURED CEREBRAL ANEURYSMS; IN-VITRO DEGRADATION; INTRACRANIAL ANEURYSMS; CORONARY-ARTERIES; SHEAR-STRESS; SUBARACHNOID HEMORRHAGE; FLOW-DIVERSION; TRIAL; IMPLANTATION;
D O I
10.1007/s00330-016-4715-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To analyze the outcomes of a magnesium alloy covered stent (MACS) for a lateral aneurysm model in common carotid artery (CCA). In 32 rabbits, a MACS (group A, n = 17) or a Willis covered stent (WCS; group B, n = 15) was inserted and the rabbits were sacrificed 1, 3, 6, or 12 months after stenting. Angiography and intravascular ultrasound (IVUS) were performed at 3, 6, and 12 months. Scanning electron microscopy was performed for six stents in each group at 1, 3, and 6 months, and histopathology and histomorphology were conducted at 3 (n = 4), 6 (n = 4), and 12 (n = 12) months. Final angiography showed complete occlusion of the aneurysms in 12 cases. IVUS at 6 and 12 months revealed a significant increase in mean lumen area of the stented CCA in group A and also showed greater mean lumen area in group A than in group B. The endothelialization process was quicker in group A than in group B. MACS is effective for occlusion of lateral aneurysms and is superior to WCS in growth of the stented CCA and endothelialization. Further work is needed to make this device available for human use. aEuro cent The MACS is an effective approach for occlusion of a lateral aneurysm. aEuro cent IVUS showed that the CCA could grow following degradation of the MACS. aEuro cent The lumen area of the stented CCA was excellent in MACS. aEuro cent HE staining displayed the degradation of the magnesium alloy stent. aEuro cent Combination of IVUS and DSA were applied in this study.
引用
收藏
页码:3694 / 3702
页数:9
相关论文
共 50 条
  • [1] Intermediate analysis of magnesium alloy covered stent for a lateral aneurysm model in the rabbit common carotid artery
    Hong-kai Cui
    Fen-Bao Li
    Ying-Chang Guo
    Yong-Li Zhao
    Rui-Fang Yan
    Wu Wang
    Yong-Dong Li
    Yong-Li Wang
    Guang-Yin Yuan
    European Radiology, 2017, 27 : 3694 - 3702
  • [2] Magnesium alloy covered stent for treatment of a lateral aneurysm model in rabbit common carotid artery: An in vivo study
    Wu Wang
    Yong-Li Wang
    Mo Chen
    Liang Chen
    Jian Zhang
    Yong-Dong Li
    Ming-Hua Li
    Guang-Yin Yuan
    Scientific Reports, 6
  • [3] Magnesium alloy covered stent for treatment of a lateral aneurysm model in rabbit common carotid artery: An in vivo study
    Wang, Wu
    Wang, Yong-Li
    Chen, Mo
    Chen, Liang
    Zhang, Jian
    Li, Yong-Dong
    Li, Ming-Hua
    Yuan, Guang-Yin
    SCIENTIFIC REPORTS, 2016, 6
  • [4] Covered Stent Treatment for an Aneurysm of a Saphenous Vein Graft to the Common Carotid Artery
    Nishinari, Kenji
    Wolosker, Nelson
    Yazbek, Guilherme
    Bernardi, Christiano Vinicius
    Zottele Bomfim, Guilherme Andre
    ANNALS OF VASCULAR SURGERY, 2010, 24 (07) : 954.e9 - 954.e12
  • [5] Treatment of an Acute Mycotic Aneurysm of the Common Carotid Artery with a Covered Stent-Graft
    Lee, Suk Hoon
    Cho, Young Kwon
    Park, Jong Moo
    Chung, Curie
    Kim, Hyun Suk
    Woo, Jeong Joo
    YONSEI MEDICAL JOURNAL, 2012, 53 (01) : 224 - 227
  • [6] Occlusion of lateral aneurysm in common carotid artery of rabbits with flow-diverting stent
    He, Daohua
    Yang, Ming
    Cai, Mingjun
    Liu, Peng
    Pan, Li
    Diao, Bo
    Feng, Lei
    Feng, Guang
    Ma, Lianting
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2016, 9 (06): : 10304 - 10310
  • [7] Fusiform Aneurysm Model in Rabbit Carotid Artery
    Reinald, Nicoleta
    Fournier, Benjamin
    Naveau, Adrien
    Couty, Ludovic
    Lemitre, Mathilde
    Seguier, Sylvie
    Coulomb, Bernard
    Gogly, Bruno
    Lafont, Antoine
    Durand, Eric
    JOURNAL OF VASCULAR RESEARCH, 2010, 47 (01) : 61 - 68
  • [8] A case of giant common carotid artery aneurysm associated with vascular Behqet disease: successfully treated with a covered stent
    Ohshima, Tomotaka
    Miyachi, Shigeru
    Hattori, Ken-ichi
    Iizuka, Hiroshi
    Izumi, Takashi
    Nakane, Yukimi
    Aimi, Yuri
    Yoshida, Jun
    SURGICAL NEUROLOGY, 2008, 69 (03): : 297 - 301
  • [9] Use of covered stent grafts to treat an extracranial carotid artery aneurysm due to carotid stent deformation: A case report
    Fujita, Ohju
    Hayashi, Shigemasa
    Oshima, Ryosuke
    Matsuyama, Tomoki
    Tamai, Hiroaki
    CLINICAL NEUROLOGY AND NEUROSURGERY, 2024, 241
  • [10] DEVELOPMENT OF A FUSIFORM ANEURYSM MODEL IN RABBIT CAROTID ARTERY
    Couty, Ludovic
    Reinald, Nicoleta
    Fournier, Benjamin
    Lemitre, Mathilde
    Seguier, Sylvie
    Coulomb, Bernard
    Gogly, Bruno
    Lafont, Antoine
    Durand, Eric
    WOUND REPAIR AND REGENERATION, 2009, 17 (04) : A64 - A64