Quantifying non-contact tip-sample thermal exchange parameters for accurate scanning thermal microscopy with heated microprobes

被引:20
|
作者
Wilson, Adam A. [1 ]
Borca-Tasciuc, Theodorian [1 ]
机构
[1] Rensselaer Polytech Inst, Dept Mech Aerosp & Nucl Engn, 110 8th St, Troy, NY 12180 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2017年 / 88卷 / 07期
关键词
TRANSPORT-PROPERTIES; CONDUCTIVITY; AIR; CALIBRATION; MECHANISMS; NANOWIRES; PROBE; FILMS;
D O I
10.1063/1.4991017
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Simplified heat-transfer models are widely employed by heated probe scanning thermal microscopy techniques for determining thermal conductivity of test samples. These parameters have generally been assumed to be independent of sample properties; however, there has been little investigation of this assumption in non-contact mode, and the impact calibration procedures have on sample thermal conductivity results has not been explored. However, there has been little investigation of the commonly used assumption that thermal exchange parameters are sample independent in non-contact mode, or of the impact calibration procedures have on sample thermal conductivity results. This article establishes conditions under which quantitative, localized, non-contact measurements using scanning thermal microscopy with heated microprobes may be most accurately performed. The work employs a three-dimensional finite element (3DFE) model validated using experimental results and no fitting parameters, to determine the dependence of a heated microprobe thermal resistance as a function of sample thermal conductivity at several values of probe-to-sample clearance. The two unknown thermal exchange parameters were determined by fitting the 3DFE simulated probe thermal resistance with the predictions of a simplified probe heat transfer model, for two samples with different thermal conductivities. This calibration procedure known in experiments as the intersection method was simulated for sample thermal conductivities in the range of 0.1-50 W m(-1) K-1 and clearance values in the 260-1010 nm range. For a typical Wollaston wire microprobe geometry as simulated here, both the thermal exchange radius and thermal contact resistance were found to increase with the sample thermal conductivity in the low thermal conductivity range while they remained approximately constant for thermal conductivities > 1 W m(-1) K-1, with similar trends reported for all clearance values investigated. It is shown that versatile sets of calibration samples for the intersection method should employ either medium range (1 W m(-1) K-1) and (2 W m(-1) K-1) thermal conductivities, or wide range (0.5 W m(-1) K-1) and (50 W m(-1) K-1). The medium range yielded results within 1.5%-20.4% of the expected values of thermal conductivity for specimens with thermal conductivity within 0.1-10 W m(-1) K-1, while the wide range yielded values within 0.5%-19.4% in the same range. Published by AIP Publishing.
引用
收藏
页数:10
相关论文
共 33 条
  • [1] Tip-sample Heat Transfer in Non-contact mode Scanning Thermal Microscopy with Wollaston Microprobes
    Zhang, Yanliang
    Han, Liang
    Borca-Tasciuc, Theodorian
    PROCEEDINGS OF THE ASME INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, 2011, VOL 10, PTS A AND B, 2012, : 405 - 406
  • [2] Monolithically fabricated sample for the calibration of the tip-sample thermal conductance in scanning thermal microscopy
    Umatova, Zarina
    Zhang, Yuan
    Lambert, Rory
    Dobson, Phillip S.
    Weaver, Jonathan M. R.
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 2020, 38 (04):
  • [3] Analysis of heat transfer in the water meniscus at the tip-sample contact in scanning thermal microscopy
    Assy, Ali
    Lefevre, Stephane
    Chapuis, Pierre-Olivier
    Gomes, Severine
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2014, 47 (44)
  • [4] Strong tip-sample coupling in thermal radiation scanning tunneling microscopy
    Joulain, K.
    Ben-Abdallah, P.
    Chapuis, P. -O.
    De Wilde, Y.
    Babuty, A.
    Henkel, C.
    JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 2014, 136 : 1 - 15
  • [5] Quantitative analysis of tip-sample interaction in non-contact scanning force spectroscopy
    Palacios-Lidon, Elisa
    Colchero, Jaime
    NANOTECHNOLOGY, 2006, 17 (21) : 5491 - 5500
  • [6] AC scanning thermal microscopy:: Tip-sample interaction and buried defects modellings
    Gomès, S
    Trannoy, N
    Depasse, F
    Grossel, P
    INTERNATIONAL JOURNAL OF THERMAL SCIENCES, 2000, 39 (04) : 526 - 531
  • [7] Reduction of tip-sample contact using dielectrophoretic force scanning probe microscopy
    Hilton, AM
    Lynch, BP
    Simpson, GJ
    ANALYTICAL CHEMISTRY, 2005, 77 (24) : 8008 - 8012
  • [8] Tip-sample distance dependency of non-contact atomic force microscopy images on a GaAs(110) surface
    Uehara, N
    Hosoi, H
    Sueoka, K
    Mukasa, K
    NANOTECHNOLOGY, 2004, 15 (02) : S97 - S100
  • [9] THE CONVOLUTION OF THE TIP-SAMPLE CONTACT STIFFNESS AND THE CANTILEVER STIFFNESS IN SCANNING FORCE AND FRICTION MICROSCOPY
    BURNHAM, NA
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1993, 205 : 348 - COLL
  • [10] Imaging contrast and tip-sample interaction of non-contact amplitude modulation atomic force microscopy with Q-control
    Shi, Shuai
    Guo, Dan
    Luo, Jianbin
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2017, 50 (41)