Empirical modeling of the percent depth dose for megavoltage photon beams

被引:2
|
作者
Li, Xiao-Jun [1 ,2 ,3 ]
Ye, Yan-Cheng [1 ,2 ,3 ]
Zhang, Yan-Shan [1 ,2 ,3 ]
Wu, Jia-Ming [1 ,2 ,3 ,4 ,5 ]
机构
[1] Wuwei Canc Hosp, Heavy Ion Ctr, Wuwei City, Gansu, Peoples R China
[2] Gansu Wuwei Acad Med Sci, Wuwei City, Gansu, Peoples R China
[3] Gansu Wuwei Tumor Hosp, Wuwei City, Gansu, Peoples R China
[4] Chengde Med Univ, Dept Med Phys, Chengde City, Hebei, Peoples R China
[5] Yee Zen Gen Hosp, Dept Radiat Oncol, Taoyuan, Taiwan
来源
PLOS ONE | 2022年 / 17卷 / 01期
关键词
RATIO;
D O I
10.1371/journal.pone.0261042
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Introduction This study presents an empirical method to model the high-energy photon beam percent depth dose (PDD) curve by using the home-generated buildup function and tail function (buildup-tail function) in radiation therapy. The modeling parameters n and p of buildup-tail function can be used to characterize the Collimator Scatter Factor (So either in a square field or in the different individual upper jaw and lower jaw setting separately for individual monitor unit check. Methods and materials The PDD curves for four high-energy photon beams were modeled by the buildup and tail function in this study. The buildup function was a quadratic function in the form of d/root d(2)+n with the main parameter of d (depth in water) and n, while the tail function was in the form of e(-mu d) and was composed by an exponential function with the main parameter of d and mu. The PDD was the product of buildup and tail function, PDD = d/root d(2)+n . e(-mu d). The PDD of four-photon energies was characterized by the buildup-tail function by adjusting the parameters n and mu. The S-c of 6 MV and 10 MV can then be expressed simply by the modeling parameters n and mu. Results The main parameters n increases in buildup-tail function when photon energy increased. The physical meaning of the parameter n expresses the beam hardening of photon energy in PDD. The fitting results of parameters n in the buildup function are 0.17, 0.208, 0.495, 1.2 of four-photon energies, 4 MV, 6 MV, 10 MV, 18 MV, respectively. The parameter mu can be treated as attenuation coefficient in tail function and decreases when photon energy increased. The fitting results of parameters p in the tail function are 0.065, 0.0515, 0.0458, 0.0422 of four-photon energies, 4 MV, 6 MV, 10 MV, 18 MV, respectively. The values of n and mu obtained from the fitted buildup-tail function were applied into an analytical formula of S-c = n(E) (S)(0.63)mu to get the collimator to scatter factor S-c for 6 and 10 MV photon beam, while n(E), mu(E), S denotes n, mu at photon energy E of field size S, respectively. The calculated S-c were compared with the measured data and showed agreement at different field sizes to within +/- 1.5%. Conclusions We proposed a model incorporating a two-parameter formula which can improve the fitting accuracy to be better than 1.5% maximum error for describing the PDD in different photon energies used in clinical setting. This model can be used to parameterize the S-c factors for some clinical requirements. The modeling parameters n and mu can be used to predict the S-c in either square field or individual jaws opening asymmetrically for treatment monitor unit double-check in dose calculation. The technique developed in this study can also be used for systematic or random errors in the QA program, thus improves the clinical dose computation accuracy for patient treatment.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] Measurement of ionization chamber absorbed dose kQ factors in megavoltage photon beams
    McEwen, Malcolm R.
    MEDICAL PHYSICS, 2010, 37 (05) : 2179 - 2193
  • [22] Tumour dose enhancement using modified megavoltage photon beams and contrast media
    Robar, J
    Riccio, S
    Martin, M
    MEDICAL PHYSICS, 2003, 30 (07) : 1952 - 1953
  • [23] Reduction of the Skin Surface Dose for Megavoltage Photon Beams Passing Carbon Couch
    Tamura, M.
    Monzen, H.
    Matsumoto, K.
    Okumura, M.
    Nishimura, Y.
    MEDICAL PHYSICS, 2017, 44 (06) : 2945 - 2945
  • [24] Tumour dose enhancement using modified megavoltage photon beams and contrast media
    Robar, JL
    Riccio, SA
    Martin, MA
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (14): : 2433 - 2449
  • [25] BUILDUP REGION AND DEPTH OF DOSE MAXIMUM OF MEGAVOLTAGE X-RAY-BEAMS
    SIXEL, KE
    PODGORSAK, EB
    MEDICAL PHYSICS, 1994, 21 (03) : 411 - 416
  • [26] EFFECT OF ANGULATION ON THE PERCENT DEPTH DOSE OF ELECTRON-BEAMS
    BIGGS, PJ
    MEDICAL PHYSICS, 1983, 10 (04) : 528 - 528
  • [27] Depth for dose calibration in high energy photon beams
    Sjogren, R
    Karlsson, MG
    Karlsson, M
    Svensson, H
    RADIOTHERAPY AND ONCOLOGY, 1997, 43 (03) : 311 - 313
  • [28] Investigation of the effects of a carbon-fiber tabletop on the surface dose and attenuation dose for megavoltage photon beams
    Gursoy, G.
    Eser, E.
    Yigitoglu, I
    Koc, H.
    Kahraman, F. C.
    Yamcicier, S.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2018, 16 (02): : 235 - 241
  • [29] Evaluation of uncertainty predictions and dose output for model-based dose calculations for megavoltage photon beams
    Olofsson, Jorgen
    Nyholm, Tufve
    Georg, Dietmar
    Ahnesjo, Anders
    Karlsson, Mikael
    MEDICAL PHYSICS, 2006, 33 (07) : 2548 - 2556
  • [30] Impact of transverse magnetic fields on dose response of a nanoDot OSLD in megavoltage photon beams
    Ito, Shotaro
    Araki, Fujio
    Hoshida, Kento
    Ohno, Takeshi