3-DOF Force-Sensing Motorized Micro-Forceps for Robot-Assisted Vitreoretinal Surgery

被引:59
|
作者
Gonenc, Berk [1 ]
Chamani, Alireza [1 ]
Handa, James [2 ]
Gehlbach, Peter [2 ]
Taylor, Russell H. [1 ]
Iordachita, Iulian [1 ]
机构
[1] Johns Hopkins Univ, CISST ERC, Baltimore, MD 21218 USA
[2] Johns Hopkins Sch Med, Wilmer Eye Inst, Baltimore, MD 21287 USA
基金
美国国家卫生研究院;
关键词
Fiber Bragg grating; force sensing; micro-forceps; IATROGENIC RETINAL BREAKS; DIABETIC-RETINOPATHY; UNITED-STATES; MICROSURGICAL INSTRUMENT; EPIRETINAL MEMBRANES; STRAIN SENSOR; PREVALENCE; FEEDBACK; TEMPERATURE; TISSUE;
D O I
10.1109/JSEN.2017.2694965
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In vitreoretinal surgery, membrane peeling is a prototypical task where a layer of fibrous tissue is delaminated off the retina with a micro-forceps by applying very fine forces that are mostly imperceptible to the surgeon. Previously, we developed sensitized ophthalmic surgery tools based on fiber Bragg grating strain sensors, which were shown to precisely detect forces at the instrument's tip in two degrees of freedom perpendicular to the tool axis. This paper presents a new design that employs an additional sensor to capture also the tensile force along the tool axis. The grasping functionality is provided via a compact motorized unit. To compute forces, we investigate two distinct fitting methods: a linear regression and a nonlinear fitting based on second-order Bernstein polynomials. We carry out experiments to test the repeatability of sensor outputs, calibrate the sensor, and validate its performance. Results demonstrate sensor wavelength repeatability within 2 pm. Although the linear method provides sufficient accuracy in measuring transverse forces, in the axial direction, it produces a root mean square (rms) error over 3 mN even for a confined magnitude and direction of forces. On the other hand, the nonlinear method provides a more consistent and accurate measurement of both the transverse and axial forces for the entire force range (0-25 mN). Validation, including random samples, shows that our tool with the nonlinear force computation method can predict 3-D forces with an rms error under 0.15 mN in the transverse plane and within 2 mN accuracy in the axial direction.
引用
收藏
页码:3526 / 3541
页数:16
相关论文
共 35 条
  • [1] Design of 3-DOF Force Sensing Micro-Forceps for Robot Assisted Vitreoretinal Surgery
    Gonenc, Berk
    Handa, James
    Gehlbach, Peter
    Taylor, Russell H.
    Iordachita, Iulian
    2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 5686 - 5689
  • [2] 3-DOF Force-Sensing Micro-Forceps for Robot-Assisted Membrane Peeling: Intrinsic Actuation Force Modeling
    Gao, Anzhu
    Gonenc, Berk
    Guo, Jiangzhen
    Liu, Hao
    Gehlbach, Peter
    Iordachita, Iulian
    2016 6TH IEEE INTERNATIONAL CONFERENCE ON BIOMEDICAL ROBOTICS AND BIOMECHATRONICS (BIOROB), 2016, : 489 - 494
  • [3] Towards Robot-Assisted Vitreoretinal Surgery: Force-Sensing Micro-Forceps Integrated with a Handheld Micromanipulator
    Gonenc, Berk
    Feldman, Ellen
    Gehlbach, Peter
    Handa, James
    Taylor, Russell H.
    Iordachita, Iulian
    2014 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION (ICRA), 2014, : 1399 - 1404
  • [4] Force Sensing Micro-Forceps for Robot Assisted Retinal Surgery
    Kuru, Ismail
    Gonenc, Berk
    Balicki, Marcin
    Handa, James
    Gehlbach, Peter
    Taylor, Russell H.
    Iordachita, Iulian
    2012 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2012, : 1401 - 1404
  • [5] Force sensing micro-forceps with integrated fiber Bragg grating for vitreoretinal surgery
    He, Xingchi
    Balicki, Marcin A.
    Kang, Jin U.
    Gehlbach, Peter L.
    Handa, James T.
    Taylor, Russell H.
    Iordachita, Iulian I.
    OPTICAL FIBERS AND SENSORS FOR MEDICAL DIAGNOSTICS AND TREATMENT APPLICATIONS XII, 2012, 8218
  • [6] Motorized Force-Sensing Micro-Forceps with Tremor Cancelling and Controlled Micro-Vibrations for Easier Membrane Peeling
    Gonenc, Berk
    Gehlbach, Peter
    Handa, James
    Taylor, Russell H.
    Iordachita, Iulian
    2014 5TH IEEE RAS & EMBS INTERNATIONAL CONFERENCE ON BIOMEDICAL ROBOTICS AND BIOMECHATRONICS (BIOROB), 2014, : 244 - 251
  • [7] A force-sensing retractor for robot-assisted transoral surgery
    Lifeng Zhu
    Shuyan Yang
    Jiangwei Shen
    Cheng Wang
    Aiguo Song
    International Journal of Computer Assisted Radiology and Surgery, 2022, 17 : 2001 - 2010
  • [8] A force-sensing retractor for robot-assisted transoral surgery
    Zhu, Lifeng
    Yang, Shuyan
    Shen, Jiangwei
    Wang, Cheng
    Song, Aiguo
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2022, 17 (11) : 2001 - 2010
  • [9] A Laminated Continuum Robot With Gripping Force-Sensing Forceps for Robot-Assisted Minimally Invasive Surgeries
    Ju, Feng
    Hong, Yan
    Bai, Dongming
    Yin, Zhanpeng
    Zhu, Chengjun
    Yue, Wenchao
    IEEE SENSORS JOURNAL, 2023, 23 (01) : 766 - 774
  • [10] FBG-Based Transverse and Axial Force-Sensing Micro-Forceps for Retinal Microsurgery
    Gonene, Berk
    Iordachita, Iulian
    2016 IEEE SENSORS, 2016,