Derivation of Indices of Left Ventricular Contractility in the Setting of Continuous-Flow Left Ventricular Assist Device Support

被引:4
|
作者
Gupta, Sunil [1 ,2 ]
Muthiah, Kavitha [1 ,2 ,3 ]
Woldendorp, Kei [1 ,2 ]
Robson, Desiree [1 ]
Jansz, Paul [1 ]
Hayward, Christopher S. [1 ,2 ,3 ]
机构
[1] St Vincents Hosp, Heart Failure & Transplant Unit, Sydney, NSW 2010, Australia
[2] Univ New S Wales, Fac Med, Sydney, NSW, Australia
[3] Victor Chang Cardiac Res Inst, Sydney, NSW, Australia
关键词
Contractility; Left ventricular assist device; Continuous flow; Heart failure; RECOVERY; BRIDGE;
D O I
10.1111/aor.12292
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It is important to accurately monitor residual cardiac function in patients under long-term continuous-flow left ventricular assist device (cfLVAD) support. Two new measures of left ventricular (LV) chamber contractility in the cfLVAD-unloaded ventricle include I-Q, a regression coefficient between maximum flow acceleration and flow pulsatility at different pump speeds; and K, a logarithmic relationship between volumes moved in systole and diastole. We sought to optimize these indices. We also propose RIQ, a ratio between maximum flow acceleration and flow pulsatility at baseline pump speed, as an alternative to I-Q. Eleven patients (mean age 49 +/- 11 years) were studied. The K index was derived at baseline pump speed by defining systolic and diastolic onset as time points at which maximum and minimum volumes move through the pump. I-Q across the full range of pump speeds was markedly different between patients. It was unreliable in three patients with underlying atrial fibrillation (coefficient of determination R-2 range: 0.38-0.74) and also when calculated without pump speed manipulation (R-2 range: 0.01-0.74). The K index was within physiological ranges, but poorly correlated to both I-Q (P=0.42) and RIQ (P=0.92). In four patients there was excellent correspondence between RIQ and I-Q, while four other patients showed a poor relationship between these indices. As RIQ does not require pump speed changes, it may be a more clinically appropriate measure. Further studies are required to determine the validity of these indices.
引用
收藏
页码:1029 / 1034
页数:6
相关论文
共 50 条
  • [1] Impact of continuous-flow left ventricular assist device support on right ventricular function
    Morgan, Jeffrey A.
    Paone, Gaetano
    Nemeh, Hassan W.
    Murthy, Raghav
    Williams, Celeste T.
    Lanfear, David E.
    Tita, Cristina
    Brewer, Robert J.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2013, 32 (04): : 398 - 403
  • [2] Evaluation of Risk Indices in Continuous-Flow Left Ventricular Assist Device Patients
    Schaffer, Justin M.
    Allen, Jeremiah G.
    Weiss, Eric S.
    Patel, Nishant D.
    Russell, Stuart D.
    Shah, Ashish S.
    Conte, John V.
    [J]. ANNALS OF THORACIC SURGERY, 2009, 88 (06): : 1889 - 1896
  • [3] Continuous-Flow Left Ventricular Assist Device REPLY
    Slaughter, Mark S.
    Rogers, Joseph G.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2010, 362 (12): : 1149 - 1149
  • [4] Determination of Left Ventricular Preload and Contractility Using Waveform Analysis in Continuous-Flow Left Ventricular Assist Device Patients
    Ramanayake, A.
    Robson, D.
    Gunawan, A.
    Krishnaswamy, R. J.
    Muthiah, K.
    Adji, A.
    Hayward, C. S.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2022, 41 (04): : S228 - S229
  • [5] Ventricular Arrhythmias in Patients With a Continuous-Flow Left Ventricular Assist Device
    Yap, Sing-Chien
    Ramjankhan, Faiz
    de Jonge, Rahat Muslem Nicolaas
    Kirkels, Hans J.
    Akin, Sakir
    Manintveld, Olivier C.
    Birim, Ozcan
    Szili-Torok, Tamas
    Caliskan, Kadir
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 68 (03) : 323 - 325
  • [6] Left Ventricular Recovery with Explantation of Continuous-Flow Left Ventricular Assist Device after 5 Years of Support
    Critsinelis, Andre
    Kurihara, Chitaru
    Kawabori, Masashi
    Sugiura, Tadahisa
    Civitello, Andrew B.
    Morgan, Jeffrey A.
    [J]. ANNALS OF THORACIC AND CARDIOVASCULAR SURGERY, 2021, 27 (03) : 211 - 214
  • [7] Tolerance of Sustained Ventricular Fibrillation During Continuous-Flow Left Ventricular Assist Device Support
    Baldwin, Andrew C. W.
    Gemmato, Courtney J.
    Sandoval, Elena
    Cohn, William E.
    Morgan, Jeffrey A.
    Frazier, O. H.
    [J]. TEXAS HEART INSTITUTE JOURNAL, 2017, 44 (05): : 357 - 360
  • [8] Measurement of Blood Pressure during Support with a Continuous-Flow Left Ventricular Assist Device in the Outpatient Setting
    Coyle, L. A.
    Gallagher, C.
    Maier, J.
    Bhat, G.
    Pappas, P.
    Tatooles, A.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2013, 32 (04): : S235 - S235
  • [9] Spontaneous Oscillatory Left Ventricular-Aortic Uncoupling Under Continuous-Flow Left Ventricular Assist Device Support
    Jain, Pankaj
    Meredith, Thomas
    Adji, Audrey
    Schnegg, Bruno
    Hayward, Christopher S.
    [J]. CIRCULATION-HEART FAILURE, 2021, 14 (02) : E007658
  • [10] Physiology of Continuous-flow Left Ventricular Assist Device Therapy
    Rosenbaum, Andrew N.
    Antaki, James F.
    Behfar, Atta
    Villavicencio, Mauricio A.
    Stulak, John
    Kushwaha, Sudhir S.
    [J]. COMPREHENSIVE PHYSIOLOGY, 2022, 12 (01) : 2731 - 2767