Reproducibility of RTVue retinal nerve fiber layer thickness and optic nerve head measurements in normal and glaucoma eyes

被引:10
|
作者
Li Ji-peng [1 ]
Wang Xiao-zhen [1 ]
Fu Jing [1 ]
Li Shu-ning [1 ]
Wang Ning-li [1 ]
机构
[1] Capital Med Univ, Beijing Tongren Hosp, Tongren Eye Ctr, Ophthalmol & Visual Sci Key Lab, Beijing 100730, Peoples R China
关键词
diagnostic techniques; ophthalmological; instrumentation; optical coherence tomography; standards; reproducibility; COHERENCE TOMOGRAPHY; STRATUSOCT; DOMAIN; DAMAGE;
D O I
10.3760/cma.j.issn.0366-6999.2010.14.013
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background RTVue spectral-domain optical coherence tomography (OCT) is a new, ultra high-speed and high-resolution instrument, potentially to measure the presence of glaucoma or its progression accurately. The objective of this study was to evaluate its reproducibility of retinal nerve fiber layer (RNFL) thickness and optic nerve head (ONH) measurements in normal and glaucoma eyes. Methods This study was an observational clinical study. One eye was selected randomly from each of 89 normal individuals and 63 glaucoma patients in a range of severity. RNFL thickness and ONH were measured 3 times on the same day to determine intrasession variability. The same instrument was used by the same operator for all scans. lntrasession within-subject standard deviation (Sw), precision (1.96xSw), coefficient of variation (CVw, 100xSw/overall mean), and intraclass coefficient (ICC) were calculated to evaluate reproducibility. Results RTVue OCT demonstrated double hump patterns in the RNFL profiles. High reproducibility was observed in all ONH parameters. For normal eyes, the value of ICC ranged between 0.98 and 1.00. For eyes with different extent of glaucoma, it ranged between 0.94 and 1.00. High reproducibility was also observed in RNFL thickness measurements. The values of ICC for averaged RNFL thickness ranged between 0.95 and 1.00 in all cases. For regional parameters, it ranged from 0.94 to 0.98 for normal eyes, 0.94 to 1.00 for mild glaucoma eyes, 0.87 to 1.00 for moderate glaucoma eyes, and 0.77 to 0.97 for severe glaucoma eyes. The nasal regions of severe glaucoma appeared to be most variable, as nasal lower region and inferior nasal region had the ICC values of 0.77 and 0.87. Conclusion Reproducibility of RTVue RNFL and ONH measurements was excellent in normal and glaucoma groups. Chin Med J 2010;123(14):1898-1903
引用
收藏
页码:1898 / 1903
页数:6
相关论文
共 50 条
  • [21] Optic nerve head, retinal nerve fiber layer and macular thickness measurements in patients with restless legs syndrome
    Koskderelioglu, A.
    Kusbeci, T.
    Kusbeci, O.
    Gedizlioglu, M.
    MOVEMENT DISORDERS, 2015, 30 : S487 - S487
  • [22] Time-Course Changes in Optic Nerve Head Blood Flow and Retinal Nerve Fiber Layer Thickness in Eyes with Open-angle Glaucoma
    Kiyota, Naoki
    Shiga, Yukihiro
    Omodaka, Kazuko
    Pak, Kyongsun
    Nakazawa, Toru
    OPHTHALMOLOGY, 2021, 128 (05) : 663 - 671
  • [23] Optic disc rim area to retinal nerve fiber layer thickness correlation: comparison of diabetic and normal tension glaucoma eyes
    Min Hee Suh
    Seok Hwan Kim
    Ki Ho Park
    Hyeong Gon Yu
    Jang Won Huh
    Dong Myung Kim
    Japanese Journal of Ophthalmology, 2013, 57 : 156 - 165
  • [24] Optic disc rim area to retinal nerve fiber layer thickness correlation: comparison of diabetic and normal tension glaucoma eyes
    Suh, Min Hee
    Kim, Seok Hwan
    Park, Ki Ho
    Yu, Hyeong Gon
    Huh, Jang Won
    Kim, Dong Myung
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 2013, 57 (02) : 156 - 165
  • [25] Reproducibility of nerve fiber layer thickness measurements
    Weinreb, RN
    Zangwill, L
    OPHTHALMOLOGY, 1997, 104 (10) : 1530 - 1531
  • [26] Imaging the optic nerve head and nerve fiber layer in glaucoma
    Caprioli, J
    Quigley, H
    GLAUCOMA IN THE NEW MILLENNIUM, PROCEEDINGS, 2003, : 129 - 132
  • [27] NORMAL RETINAL NERVE-FIBER LAYER THICKNESS IN HUMAN EYES
    DICHTL, A
    JONAS, JB
    HOLBACH, L
    NAUMANN, GOH
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1995, 36 (04) : S627 - S627
  • [28] Retinal nerve fiber layer thickness in normal human eyes.
    Skaf, M
    Barron, E
    Heller, K
    Varma, R
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 5043 - 5043
  • [29] Reproducibility of retinal nerve fiber layer thickness measurements using scanning laser polarimetry in pseudophakic eyes
    Rhee, DJ
    Greenfield, DS
    Chen, PP
    Schiffman, J
    OPHTHALMIC SURGERY AND LASERS, 2002, 33 (02): : 117 - 122
  • [30] OCT 3 measurement reproducibility of nerve fiber layer thickness, macula thickness map and optic nerve head analysis on normal subjects
    Paunescu, LA
    Price, LL
    Stark, PC
    Beaton, S
    Ishikawa, H
    Wollstein, G
    Fujimoto, JG
    Schuman, JS
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 : U370 - U370