Bismuth breast-shield use in chest computed tomography for efficient dose reduction and sufficient image quality

被引:5
|
作者
Ko, Chih-Hsiang [1 ,2 ]
Lee, Shiou-Ping [2 ]
Hsieh, Yi-Chien [1 ,2 ]
Lee, Yuan-Hao [1 ]
Yao, Melissa Min-Szu [1 ,2 ]
Chan, Wing P. [1 ,2 ]
机构
[1] Taipei Med Univ, Wan Fang Hosp, Dept Radiol, 111 Hsing Long Rd,Sec 3, Taipei 116, Taiwan
[2] Taipei Med Univ, Coll Med, Sch Med, Dept Radiol, Taipei, Taiwan
关键词
breast shield; computed tomography scan; image quality; radiation dose; safety; TUBE CURRENT MODULATION; THORACIC CT; FEMALE BREAST; RADIATION-EXPOSURE; MDCT;
D O I
10.1097/MD.0000000000026277
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Radiosensitivity in the breasts increases the risk of carcinogenesis from exposure to the ionizing radiation of computed tomography (CT) administered in the course of medical attention. Bismuth shielding techniques have been used to reduce radiation, but image noise increased, degrading image quality. Purpose: The aim of this study was to investigate how the use of iterative reconstruction (IR) combined with bismuth shielding influences image quality. Materials and Methods: Women aged at least 20 years with body mass indexes <28 were recruited and randomly assigned to 1 of 3 CT scanning protocols without shielding, with a bismuth breast shield before the scout view, or with a bismuth breast shield after the scout view. All obtained images were reconstructed using an IR algorithm. To evaluate radiation dose, 2 Gafchromic films were placed over the clothes, 1 near each nipple. Results: Average dose reduction was significant (27.99%, P < .05) when bismuth shielding was applied after the scout view. Using the contrast-to-noise ratio, the image quality was found to be superior when the IR algorithm was applied. Using quantitative evaluations by 2 radiologists applying a 4-point Likert scale, significant differences in image quality were not found among the 3 protocols. Conclusion: Bismuth breast shields, particularly when used after acquiring scout images, are effective at reducing radiation dose without undermining the diagnostic value of the images when the IR technique is applied.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Evaluation of the impact of organ-specific dose reduction on image quality in pediatric chest computed tomography
    Johannes Boos
    Patric Kröpil
    Dirk Klee
    Philipp Heusch
    Lars Schimmöller
    Jörg Schaper
    Gerald Antoch
    Rotem S. Lanzman
    Pediatric Radiology, 2014, 44 : 1065 - 1069
  • [2] Evaluation of the impact of organ-specific dose reduction on image quality in pediatric chest computed tomography
    Boos, Johannes
    Kroepil, Patric
    Klee, Dirk
    Heusch, Philipp
    Schimmoeller, Lars
    Schaper, Joerg
    Antoch, Gerald
    Lanzman, Rotem S.
    PEDIATRIC RADIOLOGY, 2014, 44 (09) : 1065 - 1069
  • [3] Image quality and dose in computed tomography
    Jurik, AG
    Jessen, KA
    Hansen, J
    EUROPEAN RADIOLOGY, 1997, 7 (01) : 77 - 81
  • [4] Image quality and dose in computed tomography
    A. G. Jurik
    K. A. Jessen
    J. Hansen
    European Radiology, 1997, 7 : 77 - 81
  • [5] Image Quality in Computed Tomography Coronary Angiography and Radiation Dose Reduction
    Golubickas, Domas
    Lukosevicius, Saulius
    Tamakauskas, Vytenis
    Dobrovolskiene, Laima
    Baseviciene, Ingrida
    Grib, Liviu
    Ragaisyte, Nijole
    Leonavicius, Rytis
    Medvedev, Viktor
    Veikutis, Vincentas
    INFORMATICA, 2021, 32 (04) : 741 - 757
  • [6] High-Pitch Spiral Computed Tomography Effect on Image Quality and Radiation Dose in Pediatric Chest Computed Tomography
    Lell, Michael M.
    May, Matthias
    Deak, Paul
    Alibek, Sedat
    Kuefner, Michael
    Kuettner, Axel
    Koehler, Henrik
    Achenbach, Stephan
    Uder, Michael
    Radkow, Tanja
    INVESTIGATIVE RADIOLOGY, 2011, 46 (02) : 116 - 123
  • [7] Lens dose reduction with a bismuth shield in neuro cone-beam computed tomography: an investigation on optimum shield device placement conditions
    Kawauchi, Satoru
    Chida, Koichi
    Hamada, Yusuke
    Tsuruta, Wataro
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2022, 15 (01) : 25 - 36
  • [8] Lens dose reduction with a bismuth shield in neuro cone-beam computed tomography: an investigation on optimum shield device placement conditions
    Satoru Kawauchi
    Koichi Chida
    Yusuke Hamada
    Wataro Tsuruta
    Radiological Physics and Technology, 2022, 15 : 25 - 36
  • [9] Radiation dose reduction for musculoskeletal computed tomography of the pelvis with preserved image quality
    Zabak, Elaina
    Omar, Hythem
    Boothe, Ethan
    Tenorio, Lulu
    Guild, Jeffrey
    Abbara, Suhny
    Chhabra, Avneesh
    SKELETAL RADIOLOGY, 2019, 48 (03) : 375 - 385
  • [10] Radiation dose reduction for musculoskeletal computed tomography of the pelvis with preserved image quality
    Elaina Zabak
    Hythem Omar
    Ethan Boothe
    Lulu Tenorio
    Jeffrey Guild
    Suhny Abbara
    Avneesh Chhabra
    Skeletal Radiology, 2019, 48 : 375 - 385