Minimum Volume Optimal Design Method of Hybrid MRI Magnet

被引:0
|
作者
Liu, Quanyue [1 ]
Lee, Sangjin [1 ]
Choi, Kibum [2 ]
Hahn, Seungyong [2 ]
机构
[1] Uiduk Univ, Dept Elect Engn, Gyeongju 38004, Gyeongbuk, South Korea
[2] Seoul Natl Univ, Dept Elect & Comp Engn, Seoul 08826, South Korea
基金
新加坡国家研究基金会;
关键词
Hybrid Magnet; MRI; Field Superposition Technology; Hybrid Optimization method; Actively Shield Technology; FIELD;
D O I
10.1109/TASC.2021.3066325
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The purpose of this paper is to present a minimum volume optimal design method of hybrid magnetic resonance imaging (MRI) magnet using field superposition. The hybrid MRI magnet made by high-temperature superconductor (HTS) and low-temperature superconductor (LTS) is the optimal approach of producing higher magnetic field, which also reduces volume than LTS. In this study, the field superposition theory combined with linear programming and nonlinear optimization is introduced in advance to get minimum volume. Considering the different advantages of LTS magnets and HTS magnets, three hybrid MRI magnet configurations are introduced: (1) HTS insert and LTS outsert: (2) Minimum HTS magnet under LTS magnet: (3) Minimum LTS magnet under HTS magnet. In addition, the homogeneity over the central sphere volume and the scope of 5 Gauss fringe field are considered in actively shielding MRI design. Finally, the three different configurations are compared and a 7 T hybrid actively shielding MRI magnet is presented. The designed magnet homogeneity and 5 Gauss fringe field are validated by finite element simulation method, which illustrates the efficiency and the flexibility of the proposed optimal method.
引用
收藏
页数:5
相关论文
共 50 条
  • [21] A new design method for completely open architecture permanent magnet for MRI
    Yao, YY
    Fang, YT
    Koh, CS
    Ni, GZ
    IEEE TRANSACTIONS ON MAGNETICS, 2005, 41 (05) : 1504 - 1507
  • [22] OPTIMAL-DESIGN OF HYBRID MAGNET IN MAGLEV SYSTEM WITH BOTH PERMANENT AND ELECTROMAGNETS
    ONUKI, T
    TODA, Y
    IEEE TRANSACTIONS ON MAGNETICS, 1993, 29 (02) : 1783 - 1786
  • [23] OPTIMAL MAGNET DESIGN FOR NMR
    GOTTVALD, A
    IEEE TRANSACTIONS ON MAGNETICS, 1990, 26 (02) : 399 - 401
  • [24] 3-D optimal shape design of ferromagnetic pole in MRI magnet of open permanent-magnet type
    Kim, DH
    Kim, BS
    Lee, JH
    Nah, WS
    Park, H
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2002, 12 (01) : 1467 - 1470
  • [25] A hybrid passive shimming method applied to the design of a unilateral NMR magnet
    He, Wei
    He, Zhonghua
    INTERNATIONAL JOURNAL OF APPLIED ELECTROMAGNETICS AND MECHANICS, 2015, 49 (04) : 597 - 606
  • [26] Optimal Design of Permanent Magnet Synchronous Generator for Wind Energy Conversion Considering Annual Energy Input and Magnet Volume
    Faiz, Jawad
    Ebrahimi, Bashir Mahdi
    Rajabi-Sebdani, M.
    Khan, A.
    2009 INTERNATIONAL CONFERENCE ON SUSTAINABLE POWER GENERATION AND SUPPLY, VOLS 1-4, 2009, : 1654 - +
  • [27] A Hybrid Descent Method for Optimal Sigmoid Filter Design
    Chan, Kit Yan
    Nordholm, Sven
    Low, Siow Yong
    Yong, Pei Chee
    Yiu, Ka Fai Cedric
    IEEE SIGNAL PROCESSING LETTERS, 2014, 21 (04) : 478 - 482
  • [28] Minimum Volume, Meticulous Method
    Madhavan, Beenu
    Ahmad, Imran
    REGIONAL ANESTHESIA AND PAIN MEDICINE, 2010, 35 (05) : 480 - 480
  • [29] Design concepts of optimized MRI magnet
    Sinha, Gautam
    Sundararaman, Ravishankar
    Singh, Gurnam
    IEEE TRANSACTIONS ON MAGNETICS, 2008, 44 (10) : 2351 - 2360
  • [30] THE COIL DESIGN OF THE SUPERCONDUCTING MRI MAGNET
    FUJITA, M
    IEEE TRANSACTIONS ON MAGNETICS, 1988, 24 (06) : 2907 - 2909