Development of an automated system for the analysis of inorganic chloramines in swimming pools via multi-syringe chromatography and photometric detection with ABTS

被引:9
|
作者
Chehab, Rana [1 ]
Coulomb, Bruno [1 ]
Boudenne, Jean-Luc [1 ]
Robert-Peillard, Fabien [1 ]
机构
[1] Aix Marseille Univ, CNRS, LCE, Marseille, France
关键词
Multi-syringe chromatography; Inorganic chloramines; Free chlorine; Post-column reaction; Swimming pools; DISINFECTION BY-PRODUCTS; LIQUID-CHROMATOGRAPHY; MONOCHLORAMINE; TRICHLORAMINE; CHLORINATION; WATER; NDMA;
D O I
10.1016/j.talanta.2019.120322
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Inorganic chloramines are disinfection by-products resulting from the unwanted reaction between chlorine used as disinfectant in swimming pools and nitrogenous compounds brought by bathers. This parameter (total chloramines or combined chlorine) is currently measured on site by a colorimetric method that does not allow to measure only inorganic chloramines. In this paper, a multi-syringe chromatography system combined with a post column derivatization is applied for the first time for the specific detection of the three individual inorganic chloramines (monochloramine, dichloramine and trichloramine). These latter ones are separated using a low-pressure monolithic C18 column, and separately detected after a post-column reaction with the chromogenic reagent ABTS (2,2'-azino-bis-(3-ethyl-benzothiazoline)-6-sulfonic acid-diammonium salt). Development of two ABTS reagents provides discrimination of chlorine and monochloramine that are not separated on the column. Optimization of the experimental conditions enables determination of inorganic chloramines with very good detection limits (around 10 mu g eq.Cl-2 L-1) without interferences from other chlorinated compounds such as organic chloramines or free available chlorine. The validation of the whole procedure has been successfully applied to real swimming pools samples.
引用
收藏
页数:8
相关论文
共 5 条
  • [1] Development of a size exclusion chromatography - Electrochemical detection method for the analysis of total organic and inorganic chloramines
    Amiri, Fariba
    Andrews, Susan
    JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2008, 46 (07) : 591 - 595
  • [2] Improving the chemiluminescence-based determination of sulphide in complex environmental samples by using a new, automated multi-syringe flow injection analysis system coupled to a gas diffusion unit
    Maya, Fernando
    Estela, Jose Manuel
    Cerda, Victor
    ANALYTICA CHIMICA ACTA, 2007, 601 (01) : 87 - 94
  • [3] Primary Step Towards In Situ Detection of Chemical Biomarkers in the UNIVERSE via Liquid-Based Analytical System: Development of an Automated Online Trapping/Liquid Chromatography System
    Ribette, Thomas
    Leroux, Bertrand
    Eddhif, Balkis
    Allavena, Audrey
    David, Marc
    Sternberg, Robert
    Poinot, Pauline
    Geffroy-Rodier, Claude
    MOLECULES, 2019, 24 (07)
  • [4] Development of a multi-target screening analysis for 301 drugs using a QTrap liquid chromatography/tandem mass spectrometry system and automated library searching
    Mueller, CA
    Weinmann, W
    Dresen, S
    Schreiber, A
    Gergov, M
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (10) : 1332 - 1338
  • [5] Development of a Quality Evaluation Method for Allii Macrostemonis Bulbus Based on Solid-Phase Extraction-High-Performance Liquid Chromatography-Evaporative Light Scattering Detection Chromatographic Fingerprinting, Chemometrics, and Quantitative Analysis of Multi-Components via a Single-Marker Method
    Wu, Jianfa
    Wang, Lulu
    Cui, Ying
    Liu, Chang
    Ding, Weixing
    Ren, Shen
    Dong, Rui
    Zhang, Jing
    MOLECULES, 2024, 29 (19):