Automated calibration of multiple long gauge blocks using the KRISS linear measurement system

被引:0
|
作者
Kim, Jong-Ahn [1 ]
Lee, Jae Yong [1 ,2 ]
Woo, Jae Heun [1 ]
Eom, Sung Hoon [1 ]
Kim, Myung Soon [1 ]
Kang, Chu-Shik [1 ,2 ]
机构
[1] Korea Res Inst Stand & Sci, Div Phys Metrol, Length Grp, 267 Gajeong Ro, Daejeon 34113, South Korea
[2] Korea Univ Sci & Technol, Dept Sci Measurement, 217 Gajeong Ro, Daejeon 34113, South Korea
关键词
calibration system; long gauge block; laser interferometer; automatic measurement;
D O I
10.1088/1361-6501/abcb23
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
For efficient calibration of multiple long gauge blocks (LGBs), an automatic alignment system, consisting of a multi-axis stage and an angle sensor, was developed and integrated with the Korea Research Institute of Standards and Science linear measurement system. The stage was designed to be installed in the limited space of the linear measurement system, with the ability to translate and align the linear and angular positions of LGBs. The auto-collimation angle sensor detected the tip-tilt angle of an LGB with an angular resolution of 0.11 '', by image processing the intensity profile of a beam reflected from one of the measured faces of an LGB. Using this automatic alignment system, up to five LGBs were calibrated sequentially without any manual operation or thermal instability. Therefore, multiple LGBs can be calibrated in a short time with small alignment errors and under optimal thermal conditions. When four LGBs with nominal lengths of between 250 mm and 1000 mm were measured, the measurement repeatability of each LGB length was less than 22 nm, and the calibration results agreed well with those obtained by an LGB interferometer within the expanded measurement uncertainty.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Variation in Length of Gauge Blocks Calibration Using Interferometry Technique and Machine Vision
    Phuaknoi, Pawat
    Buajarern, Jariya
    Tonmueanwai, Anusorn
    2013 10TH INTERNATIONAL CONFERENCE ON ELECTRICAL ENGINEERING/ELECTRONICS, COMPUTER, TELECOMMUNICATIONS AND INFORMATION TECHNOLOGY (ECTI-CON), 2013,
  • [22] Interferometer for long gauge blocks up to 1000 mm and it's measurement uncertainty
    Narumi, T
    Saruki, Y
    Nakamura, T
    Shutoh, S
    Sawabe, M
    RECENT DEVELOPMENTS IN OPTICAL GAUGE BLOCK METROLOGY, 1998, 3477 : 70 - 75
  • [23] THE INFLUENCE OF THE POSITION OF SUPPORT POSITIONS OF LONG PARALLEL GAUGE BLOCKS ON LENGTH MEASUREMENT
    BOHME, H
    FISCHER, W
    PTB-MITTEILUNGEN, 1986, 96 (03): : 159 - 164
  • [24] A SYSTEM FOR VACUUM GAUGE CALIBRATION USING THE COMPARISON TECHNIQUE
    NASH, PJ
    THOMPSON, TJ
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1983, 1 (02): : 172 - 174
  • [25] Absolute length calibration of gauge blocks using optical comb of a femtosecond pulse laser
    Jin, Jonghan
    Kim, Young-Jin
    Kim, Yunseok
    Kim, Seung-Woo
    OPTICS EXPRESS, 2006, 14 (13): : 5968 - 5974
  • [26] Measurements of long gauge blocks using a fringe-counting interferometer and subgauge blocks with lines
    Natl Research Lab of Metrology, Ibaraki, Japan
    Metrologia, 1 (41-48):
  • [27] AUTOMATED PARTIAL POSE MEASUREMENT SYSTEM FOR MANIPULATOR CALIBRATION EXPERIMENTS
    DRIELS, MR
    SWAYZE, WE
    IEEE TRANSACTIONS ON ROBOTICS AND AUTOMATION, 1994, 10 (04): : 430 - 440
  • [28] Measurements of long gauge blocks using a fringe-counting interferometer and subgauge blocks with lines
    Fujii, Y
    Matsumoto, H
    Fujima, I
    METROLOGIA, 1996, 33 (01) : 41 - 48
  • [29] Absolute length calibration of gauge blocks using optical comb of a femtosecond pulse laser
    Jin, Jonghan
    Kim, Young-Jin
    Kim, Yunseok
    Kang, Chu-Shik
    Kim, Seung-Woo
    INTERFEROMETRY XIII: TECHNIQUES AND ANALYSIS, 2006, 6292
  • [30] Absolute length calibration of gauge blocks using optical comb of a femtosecond pulse laser
    Billionth Uncertainty Precision Engineering Group, Korea Advanced Institute of Science and Technology , Science Town, Daejeon, Korea, Republic of
    不详
    Opt. Express, 2006, 13 (139-145):