Automated calibration of multiple long gauge blocks using the KRISS linear measurement system

被引:0
|
作者
Kim, Jong-Ahn [1 ]
Lee, Jae Yong [1 ,2 ]
Woo, Jae Heun [1 ]
Eom, Sung Hoon [1 ]
Kim, Myung Soon [1 ]
Kang, Chu-Shik [1 ,2 ]
机构
[1] Korea Res Inst Stand & Sci, Div Phys Metrol, Length Grp, 267 Gajeong Ro, Daejeon 34113, South Korea
[2] Korea Univ Sci & Technol, Dept Sci Measurement, 217 Gajeong Ro, Daejeon 34113, South Korea
关键词
calibration system; long gauge block; laser interferometer; automatic measurement;
D O I
10.1088/1361-6501/abcb23
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
For efficient calibration of multiple long gauge blocks (LGBs), an automatic alignment system, consisting of a multi-axis stage and an angle sensor, was developed and integrated with the Korea Research Institute of Standards and Science linear measurement system. The stage was designed to be installed in the limited space of the linear measurement system, with the ability to translate and align the linear and angular positions of LGBs. The auto-collimation angle sensor detected the tip-tilt angle of an LGB with an angular resolution of 0.11 '', by image processing the intensity profile of a beam reflected from one of the measured faces of an LGB. Using this automatic alignment system, up to five LGBs were calibrated sequentially without any manual operation or thermal instability. Therefore, multiple LGBs can be calibrated in a short time with small alignment errors and under optimal thermal conditions. When four LGBs with nominal lengths of between 250 mm and 1000 mm were measured, the measurement repeatability of each LGB length was less than 22 nm, and the calibration results agreed well with those obtained by an LGB interferometer within the expanded measurement uncertainty.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Uncertainty of measurement in calibration of long gauge blocks by mechanical comparison
    Acko, B
    Godina, A
    MEASURE AND QUALITY CONTROL IN PRODUCTION, 2004, 1860 : 471 - 476
  • [2] Measurement Uncertainty Analysis for Calibration of Gauge Blocks
    Godina, Andrej
    Acko, Bojan
    24TH DAAAM INTERNATIONAL SYMPOSIUM ON INTELLIGENT MANUFACTURING AND AUTOMATION, 2013, 2014, 69 : 191 - 198
  • [3] System for contactless gauge blocks measurement
    Buchta, Zdenek
    Rerucha, Simon
    Mikel, Bretislav
    Cizek, Martin
    Lazar, Josef
    Cip, Ondrej
    EIGHTH INTERNATIONAL SYMPOSIUM ON PRECISION ENGINEERING MEASUREMENTS AND INSTRUMENTATION, 2013, 8759
  • [4] MEASUREMENT OF THE LINEAR, THERMIC COEFFICIENT OF EXPANSION OF GAUGE BLOCKS
    DEBLER, E
    BOHME, H
    F&M-FEINWERKTECHNIK & MESSTECHNIK, 1980, 88 (01): : 27 - 30
  • [5] HIGH-PRECISION CALIBRATION OF LONG GAUGE BLOCKS USING THE VACUUM WAVELENGTH COMPARATOR
    DARNEDDE, H
    METROLOGIA, 1992, 29 (05) : 349 - 359
  • [6] Automated Pressure Gauge Calibration System Model 1000
    不详
    CANADIAN CERAMICS QUARTERLY-JOURNAL OF THE CANADIAN CERAMIC SOCIETY, 1996, 65 (03): : 174 - 174
  • [7] Establishment of Measuring Coordinate System for Large Gears Measurement by Using Gauge Blocks
    Li, Shaokang
    Zheng, Zhongpeng
    Wang, Linyan
    TENTH INTERNATIONAL SYMPOSIUM ON PRECISION ENGINEERING MEASUREMENTS AND INSTRUMENTATION, 2019, 11053
  • [8] MEASUREMENT SYSTEM BASED ON MULTI-WAVELENGTH INTERFEROMETRY FOR LONG GAUGE BLOCK CALIBRATION
    Wengierow, Michal
    Salbut, Leszek
    Ramotowski, Zbigniew
    Szumski, Robert
    Szykiedans, Ksawery
    METROLOGY AND MEASUREMENT SYSTEMS, 2013, 20 (03): : 479 - 490
  • [9] Characteristics of Linear Variable Differential Transformer(LVDT) Probe for Gauge Blocks Calibration
    Peingbangyang, S.
    Wongsaroj, J.
    SIAM PHYSICS CONGRESS 2017 (SPC2017), 2017, 901
  • [10] Calibration of integrating sphere system designed for roughness correction in optical calibration of gauge blocks
    Kang, Chu-Shik
    Kim, Jae Wan
    Kim, Jong-Ahn
    Eom, Tae Bong
    ADVANCED CHARACTERIZATION TECHNIQUES FOR OPTICS, SEMICONDUCTORS, AND NANOTECHNOLOGIES III, 2007, 6672