Accurate wavelength calibration method for compact CCD spectrometer

被引:22
|
作者
Sun, Y. C. [1 ]
Huang, C. [1 ]
Xia, G. [1 ]
Jin, S. Q. [1 ]
Lu, H. B. [1 ]
机构
[1] Hefei Univ Technol, Key Lab Special Display Technol, Natl Engn Lab Special Display Technol,Minist Educ, Natl Key Lab Adv Display Technol,Acad Photoelect, Hefei 230009, Peoples R China
关键词
ALGORITHM; NOISE;
D O I
10.1364/JOSAA.34.000498
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Wavelength calibration is an important step in charge-coupled device (CCD) spectrometers. In this paper, an accurate calibration method is proposed. A model of a line profile spectrum is built at the beginning, followed by noise reduction, bandwidth correction, and automatic peak-seeking treatment. Experimental tests are conducted on the USB4000 spectrometer with a mercury-argon calibration light source. Compared with the traditional method, the results show that this wavelength calibration procedure obtains higher accuracy and the deviations are within 0.1 nm. (C) 2017 Optical Society of America
引用
收藏
页码:498 / 505
页数:8
相关论文
共 50 条
  • [21] The photometric calibration of the ISO Short Wavelength Spectrometer
    Schaeidt, SG
    Morris, PW
    Salama, A
    Vandenbussche, B
    Beintema, DA
    Boxhoorn, DR
    Feuchtgruber, H
    Heras, AM
    Lahuis, F
    Leech, K
    Roelfsema, PR
    Valentijn, EA
    Bauer, OH
    vanderBliek, NS
    Cohen, M
    deGraauw, T
    Haser, LN
    vanderHucht, KA
    Huygen, E
    Katterloher, RO
    Kessler, MF
    Koornneef, J
    Luinge, W
    Lutz, D
    Planck, M
    Spoon, H
    Waelkens, C
    Waters, LBFM
    Wieprecht, E
    Wildeman, KJ
    Young, E
    Zaal, P
    ASTRONOMY & ASTROPHYSICS, 1996, 315 (02) : L55 - L59
  • [22] COMPUTERIZED CALIBRATION OF WAVELENGTH DRIVE OF PRISM SPECTROMETER
    BARNHARDT, EA
    APPLIED OPTICS, 1968, 7 (09) : 1861 - +
  • [23] Sub-pixel Wavelength Calibration of Spectrometer
    Luo, Ying
    Feng, Guojin
    2019 INTERNATIONAL CONFERENCE ON OPTICAL INSTRUMENTS AND TECHNOLOGY: OPTOELECTRONIC MEASUREMENT TECHNOLOGY AND SYSTEMS, 2020, 11439
  • [24] ACCURATE RELATIVE CALIBRATION OF A MULTICHANNEL RAMAN SPECTROMETER
    HUGUET, P
    GAUFRES, R
    JOURNAL OF RAMAN SPECTROSCOPY, 1995, 26 (04) : 325 - 326
  • [25] ACCURATE SPECTROMETER CALIBRATION IN ELECTRON-SPECTROSCOPY
    OELHAFEN, P
    FREEOUF, JL
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1983, 1 (01): : 96 - 97
  • [26] Raman shift calibration of a compact multichannel spectrometer
    Vickers, TJ
    Mann, CK
    APPLIED SPECTROSCOPY, 1999, 53 (12) : 1617 - 1622
  • [27] Multi-wavelength fringe measurement with a CCD spectrometer at COAST
    Lawson, PR
    Baldwin, JE
    Warner, PJ
    Boysen, RC
    Haniff, CA
    Rogers, J
    St-Jacques, D
    Wilson, DMA
    Young, JS
    Mackay, CD
    ASTRONOMICAL INTERFEROMETRY, PTS 1 AND 2, 1998, 3350 : 753 - 761
  • [28] A high-precision wavelength calibration method based on Fourier transform imaging spectrometer
    Xu, Yixuan
    Li, Jianxin
    Bai, Caixun
    Liu, Jie
    Zong, Yi
    Duan, Mingliang
    AOPC 2019: OPTICAL SPECTROSCOPY AND IMAGING, 2019, 11337
  • [29] New Techniques for Accurate and Stable Wavelength Calibration
    Schmidt, Tobias M.
    ADVANCES IN OPTICAL AND MECHANICAL TECHNOLOGIES FOR TELESCOPES AND INSTRUMENTATION VI, 2024, 13100
  • [30] Temporal Sensitivity of the Wavelength Calibration of a Photodiode Array Spectrometer
    Martinsen, Paul
    McGlone, V. Andrew
    Jordan, Robert B.
    Gaastra, Paul
    APPLIED SPECTROSCOPY, 2010, 64 (12) : 1325 - 1329