Line-based calibration of a micro -vision motion measurement system

被引:29
|
作者
Li, Hai [1 ]
Zhang, Xianmin [1 ]
Wu, Heng [1 ]
Gan, Jinqiang [1 ]
机构
[1] South China Univ Technol, Sch Mech & Automot Engn, Key Lab Precis Equipment & Mfg Technol, Guangzhou 510640, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
Micro-vision system; Calibration; Line-based method; Resolution target; Microscope; Motion measurement; DISPLACEMENT MEASUREMENT; COMPLIANT MECHANISMS; CAMERA CALIBRATION; OPTICAL MICROSCOPE; LENS DISTORTION; DESIGN; TOPOLOGY; STAGE; MODEL;
D O I
10.1016/j.optlaseng.2016.12.018
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We propose a flexible and practical line-based method for calibrating a micro-vision motion measurement system by using a resolution target, considering the lens distortion. First, centerlines of the stripes in the image of resolution target are extracted with an improved center gravity approach. Then distortion coefficients are obtained through the nonlinear optimization, utilizing straightness of the lines. Finally, these lines are corrected and used for estimating parameters of the system. Two experiments are conducted to validate the effectiveness of the proposed method. The results demonstrate that our method is effective and practical for calibrating the micro-vision measurement system.
引用
收藏
页码:40 / 46
页数:7
相关论文
共 50 条
  • [31] Precise calibration of binocular vision system used for vision measurement
    Cui, Yi
    Zhou, Fuqiang
    Wang, Yexin
    Liu, Liu
    Gao, He
    OPTICS EXPRESS, 2014, 22 (08): : 9134 - 9149
  • [32] Online calibration of stereo vision measurement system
    School of Electromechanical Engineering, Henan University of Science and Technology, Luoyang, Henan 471003, China
    不详
    不详
    Guangxue Xuebao, 2009, 6 (1546-1551):
  • [33] Online calibration for a stereo vision measurement system
    Yu, ZJ
    Ma, SY
    Che, RS
    Wang, QZ
    Tian, W
    Li, ZH
    Li, JZ
    MEASUREMENT TECHNOLOGY AND INTELLIGENT INSTRUMENTS VI, 2005, 295-296 : 723 - 728
  • [34] Calibration and experiment of vision measurement accuracy for motion of rocket nozzle
    Liu B.
    Ye D.
    Chen G.
    Che R.-S.
    Guangxue Jingmi Gongcheng/Optics and Precision Engineering, 2010, 18 (11): : 2513 - 2520
  • [35] Robot line structured light vision measurement system: light strip center extraction and system calibration
    Wan, Zhuoren
    Lai, Leijie
    Yin, Xiaoqia
    Mao, Jian
    Zhu, Limin
    OPTICAL ENGINEERING, 2021, 60 (11)
  • [36] Auto-calibration of the Motion Parameters for a Vision System
    Zhang, Beiwei
    Zhu, Bin
    Li, Y. F.
    2009 IEEE INTERNATIONAL CONFERENCE ON INTELLIGENT COMPUTING AND INTELLIGENT SYSTEMS, PROCEEDINGS, VOL 4, 2009, : 122 - +
  • [37] THE INCREMENTAL RIGIDITY SCHEME FOR STRUCTURE FROM MOTION - THE LINE-BASED FORMULATION
    DUBE, D
    MITICHE, A
    LECTURE NOTES IN COMPUTER SCIENCE, 1990, 427 : 292 - 296
  • [38] Study on Omni- directional Vision System Calibration Method Based on Line
    (School of Information Science & Engineering
    Northeastern University)Wang
    Yu Xu
    Xinhe (The department of automation
    North China Electric Power University)Wang
    Yong (Changchun Institute of technology)
    微计算机信息, 2006, (34) : 262 - 264
  • [39] Fast line-based experiment timing system for LabVIEW
    Owen, SF
    Hall, DS
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (01): : 259 - 265
  • [40] Measurement of surgical motion with a marker free computer vision based system
    Darolti, Cristina
    Bouchard, Ralf
    Farke, Stefan
    Condurache, Alexandru
    Hofmann, Ulrich
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2007, 2 : S213 - S215