Noninvasive assessment of cardiac output in critically ill patients by analysis of the finger blood pressure waveform

被引:33
|
作者
Hirschl, MM
Binder, M
Gwechenberger, M
Herkner, H
Bur, A
Kittler, H
Laggner, AN
机构
[1] UNIV VIENNA,NEW GEN HOSP,DEPT DERMATOL,A-1090 VIENNA,AUSTRIA
[2] UNIV VIENNA,NEW GEN HOSP,DEPT CARDIOL,A-1090 VIENNA,AUSTRIA
关键词
thermodilution method; MODELFLOW computer program; cardiac output; noninvasive; stroke volume; pulmonary artery catheter; critically ill patients; emergency medicine; intensive care medicine; finger blood pressure measurement;
D O I
10.1097/00003246-199711000-00033
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To assess whether the measurement of cardiac output by computer-assisted analysis of the finger blood pressure waveform can substitute for the thermodilution method in critically ill patients. Design: Prospective data collection. Setting: Emergency department in a 2000-bed inner city hospital. Patients: Forty-six critically ill patients requiring invasive monitoring for clinical management were prospectively studied. Interventions: Under local anesthesia a 7-Fr pulmonary artery catheter was inserted via the central subclavian or jugular vein. Cardiac output was determined by the use of a cardiac output computer and injections of 10 mt ice-cold glucose 5%. Noninvasive cardiac output was calculated from the finger blood pressure waveform by the use of the test software program. Measurements and Main Results: Three hundred twenty-three pairs of invasive and noninvasive hemodynamic measurements were collected in intervals of 30 mins from 46 patients (mean age 61.9 +/- 12.4 yrs; 35 male, 11 female). The average cardiac index during the study period was 2.83 L/min/m(2) (range 0.97 to 5.56). The overall discrepancy between both measurements was 0.14 L/min/m(2) (95% confidence interval: 0.10-0.18, p <.001). Seventy-five (23.2%) measurements had an absolute discrepancy >+/-0.50 L/min/m(2). Noninvasive and invasive comparisons of mean differential cardiac output were out of phase for 9.7% of all readings. Conclusion: Computer-assisted analysis of finger blood pressure waveform to assess cardiac output is not a substitute for the thermodilution method due to a high percentage (23.2%) of inaccurate readings; however, it may be a useful tool for the detection of relative hemodynamic trends in critically ill patients.
引用
收藏
页码:1909 / 1914
页数:6
相关论文
共 50 条
  • [41] Bedside and continuous assessment of arterial load in critically-ill patients by combined analysis of esophageal doppler blood flow and arterial pressure waveform: preliminary results
    MI Monge García
    A Santos Oviedo
    P Guijo González
    M Gracia Romero
    A Gil Cano
    Intensive Care Medicine Experimental, 3 (Suppl 1)
  • [42] Evaluation of a noninvasive blood glucose monitoring device for critically ill patients
    A Kononenko
    D Dvir
    B Grunberg
    J Cohen
    E Gabis
    P Singer
    Critical Care, 10 (Suppl 1):
  • [43] Evaluation of a noninvasive blood glucose monitoring device for critically ill patients
    O Amir
    D Dvir
    B Grunberg
    J Cohen
    E Gabis
    P Singer
    Critical Care, 12 (Suppl 2):
  • [44] Noninvasive Ventilation in Critically Ill Patients
    Gregoretti, Cesare
    Pisani, Lara
    Cortegiani, Andrea
    Ranieri, V. Marco
    CRITICAL CARE CLINICS, 2015, 31 (03) : 435 - +
  • [45] Assessment of Effective Hepatic Blood Flow in Critically Ill Patients by Noninvasive Pulse Dye-Densitometry
    Yasuaki Mizushima
    Hideo Tohira
    Yasumitsu Mizobata
    Tetsuya Matsuoka
    Junichiro Yokota
    Surgery Today, 2003, 33 : 101 - 105
  • [46] Assessment of effective hepatic blood flow in critically ill patients by noninvasive pulse dye-densitometry
    Mizushima, Y
    Tohira, H
    Mizobata, Y
    Matsuoka, T
    Yokota, J
    SURGERY TODAY, 2003, 33 (02) : 101 - 105
  • [47] Noninvasive Cardiac Output Monitoring in Critically Ill Adults: A Comparison of Bioreactance (NICOM™) with Transthoracic Echocardiography
    Abdulmahdi, M.
    Amaral, V.
    Skaanland, A.
    Garcia, P.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2018, 197
  • [48] Comparison of noninvasive electrical cardiometry and transpulmonary thermodilution for cardiac output measurement in critically ill patients: a prospective observational study
    Song, Wenliang
    Guo, Jiayan
    Cao, Daiyin
    Jiang, Jinlong
    Yang, Tao
    Ma, Xiaoxun
    Yuan, Hao
    Wu, Jianfeng
    Guan, Xiangdong
    Si, Xiang
    BMC ANESTHESIOLOGY, 2025, 25 (01):
  • [49] EFFECTS OF NOREPINEPHRINE-DRIVEN CHANGE IN ARTERIAL BLOOD PRESSURE ON FOUR DIFFERENT CONTINUOUS CARDIAC OUTPUT SYSTEMS IN CRITICALLY ILL PATIENTS
    Smith, J.
    Scaramuzzi, M.
    Charlton, P.
    Brooks, J.
    Arces, D.
    Wong, G.
    Camporota, L.
    Beale, R.
    INTENSIVE CARE MEDICINE, 2011, 37 : S280 - S280
  • [50] Pilot study on the influence of liver blood flow and cardiac output on the clearance of propofol in critically ill patients
    Peeters, Mariska Y. M.
    Aarts, Leon P. H. J.
    Boom, Ferenc A.
    Bras, Leo J.
    Tibboel, Dick
    Danhof, Meindert
    Knibbe, Catherijne A. J.
    EUROPEAN JOURNAL OF CLINICAL PHARMACOLOGY, 2008, 64 (03) : 329 - 334