Clinical evaluation of the repeatability of ocular aberrometry obtained with a new pyramid wavefront sensor

被引:18
|
作者
Plaza-Puche, Ana B. [1 ]
Salerno, Liberdade C. [1 ]
Versaci, Francesco [2 ]
Romero, Daniel [3 ]
Alio, Jorge L. [1 ,3 ]
机构
[1] Vissum, Avda Denia S-N,Edificio Vissum, Alicante 03016, Spain
[2] CSO, R&D Dept, Florence, Italy
[3] Univ Miguel Hernandez, Div Ophthalmol, Alicante, Spain
关键词
Ocular aberrometry; pyramidal aberrometer; intrasubject repeatability; IN-SITU KERATOMILEUSIS; HARTMANN-SHACK ABERROMETER; ADAPTIVE OPTICS; ABERRATIONS; PRINCIPLES; PRECISION; QUALITY; EYE;
D O I
10.1177/1120672118816060
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate the intrasubject repeatability of the ocular aberrometry obtained with a new ocular pyramidal aberrometer technology in a sample of normal eyes. Methods: A total of 53 healthy eyes of 53 subjects with ages ranging from 18 to 45 years were included in this study. In all cases, three consecutive acquisitions were obtained. Intrasubject repeatability of the measurements with a pyramidal aberrometer was calculated. Intrasubject repeatability for 4.0- and 6.0-mm pupils was evaluated within the subject standard deviation (S-w) and intraclass correlation coefficient. Results: Low values of the S-w and intraclass correlation coefficient outcomes close to 1 were observed for the sphere and cylinder at 3.0-mm pupil size. Most low S-w and intraclass correlation coefficient values close to 1 were observed for total, low-order aberrations and higher-order aberrations root mean square and for each Zernike coefficient analysis (intraclass correlation coefficient > 0.798) at 4.0-mm pupil size, with more limited outcomes for the aberrometric coefficient of Z(4, 4) with an intraclass correlation coefficient of 0.683. For a 6.0 mm pupil diameter, low S-w and intraclass correlation coefficient values close to 1 were observed for all aberrometric parameters or Zernike coefficients analyzed (intraclass correlation coefficient > 0.850). Conclusion: The new pyramidal aberrometer Osiris provides repeatable and consistent measurements of ocular aberrometry measurements in normal eyes.
引用
收藏
页码:585 / 592
页数:8
相关论文
共 20 条
  • [1] In vivo aberrometry with a pyramid wavefront sensor
    Chamot, SR
    Sheehan, M
    Esposito, S
    Dainty, JC
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [2] Intrasubject Repeatability of Internal Aberrometry Obtained With a New Integrated Aberrometer
    Pinero, David P.
    Tomas Juan, Javier
    Alio, Jorge L.
    JOURNAL OF REFRACTIVE SURGERY, 2011, 27 (07) : 509 - 517
  • [3] Accuracy, repeatability, and clinical application of spherocylindrical automated refraction using time-based wavefront aberrometry measurements
    Nissman, SA
    Tractenberg, RE
    Saba, CM
    Douglas, JC
    Lustbader, JM
    OPHTHALMOLOGY, 2006, 113 (04) : 570 - 577
  • [4] Evaluation of the Agreement Between a New Pyramid Wavefront Sensor Aberrometer and Scheiner-Smirnov Aberrometers
    Yang, Yizhou
    Ning, Rui
    Xu, Shuoyu
    Xiahou, Jinxuan
    Li, Jiawei
    Savini, Giacomo
    Schiano-Lomoriello, Domenico
    Zhou, Xingtao
    Huang, Jinhai
    JOURNAL OF REFRACTIVE SURGERY, 2024, 40 (04) : e218 - e228
  • [5] Performance Evaluation of a SCAO System for a 42m Telescope using the Pyramid Wavefront Sensor
    Garcia-Rissmann, Aurea
    Le Louarn, Miska
    ADAPTIVE OPTICS SYSTEMS II, 2010, 7736
  • [6] Demographic aspects of allergic ocular diseases and evaluation of new criteria for clinical assessment of ocular allergy
    Uchio, Eiichi
    Kimura, Ryoji
    Migita, Hironori
    Kozawa, Masahiko
    Kadonosono, Kazuaki
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2008, 246 (02) : 291 - 296
  • [7] Demographic aspects of allergic ocular diseases and evaluation of new criteria for clinical assessment of ocular allergy
    Eiichi Uchio
    Ryoji Kimura
    Hironori Migita
    Masahiko Kozawa
    Kazuaki Kadonosono
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2008, 246 : 291 - 296
  • [8] Preliminary clinical evaluation of a new subcutaneous glucose sensor
    Mastrototaro, JJ
    VanAntwerp, WP
    Mestman, JH
    DIABETOLOGIA, 1996, 39 : 814 - 814
  • [9] Three-sided pyramid wavefront sensor, part II: preliminary demonstration on the new comprehensive adaptive optics and coronagraph test instrument testbed
    Schatz, Lauren
    Codona, Johanan
    Long, Joseph D.
    Males, Jared R.
    Pullen, Weslin
    Lumbres, Jennifer
    Van Gorkom, Kyle
    Chambouleyron, Vincent
    Close, Laird M.
    Correia, Carlos
    Fauvarque, Olivier
    Fusco, Thierry
    Guyon, Olivier
    Hart, Michael
    Janin-Potiron, Pierre
    Johnson, Robert
    Jovanovic, Nemanja
    Mateen, Mala
    Sauvage, Jean-Francois
    Neichel, Benoit
    JOURNAL OF ASTRONOMICAL TELESCOPES INSTRUMENTS AND SYSTEMS, 2022, 8 (04) : 49001
  • [10] Long Term Clinical Evaluation of a New Subconjunctival Glucose Sensor
    Hasslacher, Christoph
    Auffarth, Gerd
    Platten, Isabel
    Rabsilber, Tanja
    Smith, Beate
    Kulozik, Felix
    DIABETES, 2011, 60 : A65 - A65