A method to alleviate false-positive results of the Elecsys HIV combi PT assay

被引:2
|
作者
Lu, Xiaolan [1 ,2 ,3 ]
Zhang, Minghong [4 ]
Liu, Wen [1 ,2 ,3 ]
Sheng, Nan [4 ]
Du, Qin [1 ,2 ,3 ]
Zhang, Maoxin [1 ]
Guo, Xiaolan [1 ,2 ,3 ]
Wang, Guangrong [1 ,2 ,3 ]
Wang, Qiang [1 ,2 ,3 ]
机构
[1] North Sichuan Med Coll, Affiliated Hosp, Dept Lab Med, Nanchong, Sichuan, Peoples R China
[2] North Sichuan Med Coll, Coll Lab Med, Nanchong, Sichuan, Peoples R China
[3] North Sichuan Med Coll, Fac Lab Med, Ctr Translat Med, Nanchong, Sichuan, Peoples R China
[4] Nanchong Ctr Dis Control & Prevent, Nanchong, Sichuan, Peoples R China
关键词
P24; ANTIGEN; PREVALENCE; INFECTION;
D O I
10.1038/s41598-020-80047-0
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
To explore the effects of urea dissociation on reducing false-positive results of the Elecsys HIV combi PT assay. A retrospective analysis was used to evaluate the false-positive rate of the Elecsys HIV combi PT assay. Six false-positive sera, six positive sera and six sera from patients with early HIV infection were collected. Dissociation was performed using 1 mol/L, 2 mol/L, 4 mol/L, 6 mol/L, or 8 mol/L urea, and HIV screening assay were then detected to select the appropriate concentration of urea dissociation. Next, 55 false-positive sera and 15 sera from early HIV infection were used to verify the best concentration of urea to achieve dissociation. Retrospective analysis showed that the COI of the Elecsys HIV combi PT assay in false-positive sera ranged from 1.0 to 200.0, and approximately 97.01%(227/234) of false-positive sera were in the range of 1.0-15.0. The avidity index (AI) in positive and false-positive sera decreased as the urea dissociation concentration increased. When the dissociation concentration was 6 mol/L, the AI of false-positive serum was between 0.0234 and 0.2567, and the AI of early HIV infection sera was between 0.4325 and 0.5017. The difference in AI between false-positive and positive samples was significant. When negativity was defined as an AI of less than 0.3970, the sensitivity and specificity were 100.0% and 100.0%, respectively. Urea-mediated dissociation could significantly reduce the false-positive rate of the Elecsys HIV combi PT assay with a low COI. Our findings provided a reference for distinguishing positive and false-positive of the Elecsys HIV combi PT assay.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] FALSE-POSITIVE RESULTS WITH CLINISTIX
    SMITH, JL
    BRITISH MEDICAL JOURNAL, 1965, 1 (5427): : 130 - &
  • [22] False-positive results on videofluoroscopy
    Warms, TL
    DYSPHAGIA, 1998, 13 (03) : 191 - 192
  • [23] False-positive acetaminophen results in hyperbilirubinemic patients with an enzymatic colorimetric assay
    Stolk, LM
    Nelissen, B
    van der Heijden, J
    PHARMACY WORLD & SCIENCE, 2003, 25 (05): : 242 - 242
  • [24] FALSE FALSE-POSITIVE RESULTS IN DIAGNOSTIC CYTOLOGY
    TAO, LC
    WEISBROD, GL
    RITCEY, EL
    ILVES, R
    ACTA CYTOLOGICA, 1983, 27 (05) : 554 - 555
  • [25] Serum albumin leads to false-positive results in the XTT and the MTT assay
    Funk, Dorothee
    Schrenk, Hans-Hermann
    Frei, Eva
    BIOTECHNIQUES, 2007, 43 (02) : 178 - +
  • [26] Carryover can be a cause of false-positive results with the Beckman AccuTnI assay
    Dimeski, Goce
    Jones, Brock
    Brown, Nigel
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2012, 50 (06) : 1135 - 1136
  • [27] False-positive Results of Quantiferon-Tb-Gold Assay in Children
    Quintana-Ortega, Cristian
    Mendez-Echevarria, Ana
    del Rosal, Teresa
    Gonzalez-Munoz, Miguel
    Baquero-Artigao, Fernando
    PEDIATRIC INFECTIOUS DISEASE JOURNAL, 2020, 39 (07) : 620 - 623
  • [28] An Unusual Cause of False-Positive Results with the BD GeneOhm Cdiff Assay
    Louie, Lisa
    Wong, Henry
    Mubareka, Samira
    Simor, Andrew E.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2013, 51 (04) : 1348 - 1349
  • [29] False-positive acetaminophen results in hyperbilirubinemic patients with an enzymatic colorimetric assay
    L. M. Stolk
    B. Nelissen
    J. van der Heijden
    Pharmacy World and Science, 2003, 25 : 242 - 242
  • [30] FACTITIOUS FALSE-POSITIVE TEST FOR HIV
    WU, AW
    KENNEDY, CJ
    PARADISE, M
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1988, 259 (11): : 1647 - 1647