A method to alleviate false-positive results of the Elecsys HIV combi PT assay

被引:4
|
作者
Lu, Xiaolan [1 ,2 ,3 ]
Zhang, Minghong [4 ]
Liu, Wen [1 ,2 ,3 ]
Sheng, Nan [4 ]
Du, Qin [1 ,2 ,3 ]
Zhang, Maoxin [1 ]
Guo, Xiaolan [1 ,2 ,3 ]
Wang, Guangrong [1 ,2 ,3 ]
Wang, Qiang [1 ,2 ,3 ]
机构
[1] North Sichuan Med Coll, Affiliated Hosp, Dept Lab Med, Nanchong, Sichuan, Peoples R China
[2] North Sichuan Med Coll, Coll Lab Med, Nanchong, Sichuan, Peoples R China
[3] North Sichuan Med Coll, Fac Lab Med, Ctr Translat Med, Nanchong, Sichuan, Peoples R China
[4] Nanchong Ctr Dis Control & Prevent, Nanchong, Sichuan, Peoples R China
关键词
P24; ANTIGEN; PREVALENCE; INFECTION;
D O I
10.1038/s41598-020-80047-0
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
To explore the effects of urea dissociation on reducing false-positive results of the Elecsys HIV combi PT assay. A retrospective analysis was used to evaluate the false-positive rate of the Elecsys HIV combi PT assay. Six false-positive sera, six positive sera and six sera from patients with early HIV infection were collected. Dissociation was performed using 1 mol/L, 2 mol/L, 4 mol/L, 6 mol/L, or 8 mol/L urea, and HIV screening assay were then detected to select the appropriate concentration of urea dissociation. Next, 55 false-positive sera and 15 sera from early HIV infection were used to verify the best concentration of urea to achieve dissociation. Retrospective analysis showed that the COI of the Elecsys HIV combi PT assay in false-positive sera ranged from 1.0 to 200.0, and approximately 97.01%(227/234) of false-positive sera were in the range of 1.0-15.0. The avidity index (AI) in positive and false-positive sera decreased as the urea dissociation concentration increased. When the dissociation concentration was 6 mol/L, the AI of false-positive serum was between 0.0234 and 0.2567, and the AI of early HIV infection sera was between 0.4325 and 0.5017. The difference in AI between false-positive and positive samples was significant. When negativity was defined as an AI of less than 0.3970, the sensitivity and specificity were 100.0% and 100.0%, respectively. Urea-mediated dissociation could significantly reduce the false-positive rate of the Elecsys HIV combi PT assay with a low COI. Our findings provided a reference for distinguishing positive and false-positive of the Elecsys HIV combi PT assay.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] A method to alleviate false-positive results of the Elecsys HIV combi PT assay
    Xiaolan Lu
    Minghong Zhang
    Wen Liu
    Nan Sheng
    Qin Du
    Maoxin Zhang
    Xiaolan Guo
    Guangrong Wang
    Qiang Wang
    [J]. Scientific Reports, 11
  • [2] The evaluation of low cut-off index values of Elecsys® HIV combi PT assay in predicting false-positive results
    Tang, Zhuoyun
    Gou, Yu
    Zhang, Keyi
    Zhao, Zhongyi
    Wei, Yinhao
    Li, Dongdong
    Chen, Li
    Tao, Chuanmin
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2020, 34 (11)
  • [3] New Roche HIV combi PT Assay Reduces the Number of Old Roche HIV combi-Associated False-Positive Results
    Hawkins, Robert
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2012, 138 : A206 - A206
  • [4] FALSE-POSITIVE RESULTS WITH MTT ASSAY
    ROLLINO, C
    BORSA, S
    BELLONE, G
    PICCOLI, G
    EMANUELLI, G
    [J]. JOURNAL OF IMMUNOLOGICAL METHODS, 1995, 185 (01) : 141 - 143
  • [5] A retrospective analysis of the application of the Elecsys® HIV combi PT assay in southern China
    Liang, Shaocong
    Deng, Guihua
    Zhou, Shaosong
    Zeng, Jing
    Tan, Weiqing
    Yuan, Xiaopeng
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2018, 32 (03)
  • [6] Analysis of the characteristics of patients with false-positive HIV screening assay results
    Lin, Yi-Qiang
    Gao, Ya-Ling
    Wang, Mao
    Yan, Shui-Di
    Lin, Li-Rong
    [J]. INTERNATIONAL IMMUNOPHARMACOLOGY, 2022, 105
  • [7] HIV screening and false-positive results
    Guinn, Debra
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2007, 297 (09): : 947 - 947
  • [8] COMPARATIVE STUDY OF THE ELECSYS HIV COMBI PT IMMUNOASSAY (ROCHE DIAGNOSTICS) WITH THE ELECSYS HIV COMBI IMMUNOASSAY (ROCHE DIAGNOSTICS)
    Coorevits, Liselotte
    Vanpoucke, Hilde
    Debrabandere, Johan
    [J]. ACTA CLINICA BELGICA, 2012, 67 (01): : 63 - 63
  • [9] HIV screening and false-positive results - Reply
    Gostin, Lawrence O.
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2007, 297 (09): : 948 - 948
  • [10] FALSE-POSITIVE RESULTS AND HIV SEROPREVALENCE RATES
    TURNER, AR
    [J]. CANADIAN MEDICAL ASSOCIATION JOURNAL, 1991, 144 (06) : 627 - 627