A miniaturized trace-moisture sensor based on cavity ring-down spectroscopy

被引:6
|
作者
Abe, H. [1 ]
Hashiguchi, K. [1 ]
Lisak, D. [2 ]
Honda, S. [3 ]
Miyake, T. [3 ]
Shimizu, H. [3 ]
机构
[1] AIST, Natl Metrol Inst Japan NMIJ, Tsukuba Cent 3, Tsukuba, Ibaraki 3058563, Japan
[2] Nicolaus Copernicus Univ Torun, Fac Phys Astron & Informat, Inst Phys, Grudziadzka 5, PL-87100 Torun, Poland
[3] SHINYEI Technol Co LTD, Chuo Ku, 6-5-2 Minatojima Minamimachi, Kobe, Hyogo 6500047, Japan
关键词
Absorption spectroscopy; High sensitivity; Humidity; Laser spectroscopy; Metrological traceability; Parts-per-billion; ABSORPTION-SPECTROSCOPY; SPECTROMETER; STANDARD; H2O;
D O I
10.1016/j.sna.2021.112559
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We developed a miniaturized sensor based on cavity ring-down spectroscopy (the mini-CRDS sensor) for the real-time measurement of trace moisture. The dimensions and mass were 24 cm x 9 cm x 13 cm and 2.7 kg, respectively, exclusive of the laser controller and data processing unit. The distance between the two cavity mirrors was 5 cm, and the internal volume and surface area of the cavity between the mirrors were approximately 4 cm3 and 16 cm2, respectively. The Allan-Werle deviation of trace-moisture measurement was approximately 0.46 nmol/mol at an averaging time of 10 s. The minimum detectable absorption coefficient for an averaging time of 10 s was 6.8 x 10-10 cm-1. The limit of detection (3 times the standard deviation) of moisture in nitrogen was 3.1 nmol/mol (3.1 ppb) in the amount-of-substance fraction (mole fraction). The relative standard uncertainty was tentatively evaluated as 10 % to 0.67 % for the range of 12 nmol/mol (12 ppb) to 1.3 mu mol/mol (1.3 ppm). The measurement accuracy of the mini-CRDS in the above range was evaluated by comparison with a primary trace-moisture standard that is metrologically traceable to the International System of Units (SI), and the results demonstrated good agreement between the mini-CRDS and the primary trace-moisture standard. (c) 2021 Elsevier B.V. All rights reserved.
引用
收藏
页数:11
相关论文
共 50 条
  • [31] Cavity Ring-Down Spectroscopy for ultra sensitive trace gas analysis.
    Yan, WB
    Markowski, ML
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2001, 222 : U111 - U111
  • [32] Doppler broadening thermometry based on cavity ring-down spectroscopy
    Cheng, C-F
    Wang, J.
    Sun, Y. R.
    Tan, Y.
    Kang, P.
    Hu, S-M
    METROLOGIA, 2015, 52 (05) : S385 - S393
  • [33] High sensitivity detectors based on cavity ring-down spectroscopy
    Paldus, BA
    Fidric, BG
    Sanders, SS
    Tan, SM
    Pham, H
    Kachanov, AA
    Wahl, EH
    Crosson, ER
    OPTICALLY BASED BIOLOGICAL AND CHEMICAL SENSING FOR DEFENCE, 2004, 5617 : 312 - 322
  • [34] Cavity ring-down spectroscopy in the liquid phase
    Xu, SC
    Sha, GH
    Xie, JC
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (02): : 255 - 258
  • [35] Superposition principle and cavity ring-down spectroscopy
    Princeton Univ, Princeton, United States
    J Chem Phys, 23 (10263-10277):
  • [36] On electronic type cavity ring-down spectroscopy
    Hirabayashi, Yuki
    Tanimoto, Hirokazu
    Maeda, Yoshinobu
    IEEJ Transactions on Sensors and Micromachines, 2014, 134 (08) : 243 - 246
  • [37] Towards Supercontinuum Cavity Ring-Down Spectroscopy
    K. Stelmaszczyk
    M. Fechner
    P. Rohwetter
    M. Queißer
    A. Czyżewski
    T. Stacewicz
    L. Wöste
    Applied Physics B, 2009, 94 : 369 - 373
  • [38] Cavity ring-down spectroscopy of the benzyl radical
    Tonokura, K
    Koshi, M
    JOURNAL OF PHYSICAL CHEMISTRY A, 2003, 107 (22): : 4457 - 4461
  • [39] The superposition principle and cavity ring-down spectroscopy
    Lehmann, KK
    Romanini, D
    JOURNAL OF CHEMICAL PHYSICS, 1996, 105 (23): : 10263 - 10277
  • [40] Towards Supercontinuum Cavity Ring-Down Spectroscopy
    Stelmaszczyk, K.
    Fechner, M.
    Rohwetter, P.
    Queisser, M.
    Czyzewski, A.
    Stacewicz, T.
    Woeste, L.
    APPLIED PHYSICS B-LASERS AND OPTICS, 2009, 94 (03): : 369 - 373