Determination of lead isotope ratios in seawater by quadrupole inductively coupled plasma mass spectrometry after Mg(OH)2 co-precipitation

被引:41
|
作者
Weiss, D
Boyle, EA [1 ]
Chavagnac, V
Herwegh, M
Wu, JF
机构
[1] MIT, Dept Earth Atmospher & Planetary Sci, Cambridge, MA 02139 USA
[2] Univ Bern, Inst Geol, CH-3012 Bern, Switzerland
关键词
inductively coupled plasma mass spectrometry (ICP-MS); Pb; isotope ratio; seawater; pre-concentration;
D O I
10.1016/S0584-8547(00)00176-2
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A low blank Mg(OH)(2) pre-concentration method was evaluated for the determination of lead isotope ratios (Pb-208/Pb-206, Pb-207/Pb-206) in seawater using a quadrupole ICP-MS VG Plasma Quad II +. Possible matrix effects derived from the Mg(OH), co-precipitate were assessed by spiking lead-free seawater (PbFS) and 1% (v/v) HNO3 with the certified common lead standard NBS 981 to give solutions with concentrations in the lower picogram per millilitre range. The standard curves for all three masses were linear in both matrices with minor signal loss (similar to 18%) in the Mg matrix. Mass fractionation showed similar mass biases (< 2% frac./amu) for Pb-208/Pb-206 and Pb-207/Pb-206 in both seawater and 1% (v/v) HNO3, indicating that there is no significant matrix influence on the isotope ratio determination. Using the Mg(OH)(2) precipitation method, real seawater samples were pre-concentrated approximately 28-fold, and 1 ml of 5% (v/v) HNO3 end volume was used for the subsequent measurement. The data acquisition parameters dwell time, replicates per analysis, and acquisition time were first varied to optimize analytical precision and accuracy of the ICP-MS measurements. On the basis of these results, approximately 0.5 mi of pre-concentrated solution was finally used employing a low-flow, CETAC microconcentric nebulizer to minimize sample consumption and extend acquisition time. Analyzing 18 seawater samples from the North Atlantic, the average lo external precision of triplicate measurements was approximately 0.3% for both ratios, Pb-208/Pb-206 and Pb-207/Pb-206, at the level of approximately 20 pg ml(-1) Pb. Blanks accounted for less than 3% of total lead analyzed for each sample. Six samples were also measured with TIMS and agreed in average within 0.26% for Pb-207/Pb-206 and 0.37% for Pb-208/Pb-206. Three surface water samples from the Sargasso Sea, collected in 1989, showed ratios in line with previous published ratios from the western North Atlantic. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
下载
收藏
页码:363 / 374
页数:12
相关论文
共 50 条
  • [31] Determination of lead isotope ratios in port wine by inductively coupled plasma mass spectrometry after pre-treatment by UV-irradiation
    Almeida, CMR
    Vasconcelos, MTSD
    ANALYTICA CHIMICA ACTA, 1999, 396 (01) : 45 - 53
  • [32] VERIFICATION OF A CORRECTION PROCEDURE FOR MEASUREMENT OF LEAD ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY
    KETTERER, ME
    PETERS, MJ
    TISDALE, PJ
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (06) : 439 - 443
  • [33] Accurate Measurement of Isotope Amount Ratios of Lead in Bronze with Multicollector Inductively Coupled Plasma Mass Spectrometry
    Lee, Kyoung-Seok
    Kim, Jin-il
    Yim, Yong-Hyeon
    Hwang, Euijin
    Kim, Tae Kyu
    MASS SPECTROMETRY LETTERS, 2013, 4 (04) : 87 - 90
  • [34] INTERLABORATORY COMPARISON STUDY ON LEAD ISOTOPE RATIOS DETERMINED BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY
    FURUTA, N
    ANALYTICAL SCIENCES, 1991, 7 (05) : 823 - 826
  • [35] Plutonium determination in seawater by inductively coupled plasma mass spectrometry: A review
    Cao, Liguo
    Bu, Wenting
    Zheng, Jian
    Pan, Shaoming
    Wang, Zhongtang
    Uchida, Shigeo
    TALANTA, 2016, 151 : 30 - 41
  • [36] AN ONLINE METHOD FOR THE DETERMINATION OF LEAD AND LEAD-ISOTOPE RATIOS IN FRESH AND SALINE WATERS BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY
    HALICZ, L
    LAM, JWH
    MCLAREN, JW
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1994, 49 (07) : 637 - 647
  • [37] Capabilities of inductively coupled plasma mass spectrometry for the measurement of Fe isotope ratios
    Vanhaecke, F
    Balcaen, L
    De Wannemacker, G
    Moens, L
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2002, 17 (08) : 933 - 943
  • [38] Is measurement of Si isotope ratios by inductively coupled plasma mass spectrometry realistic?
    Linge, K. L.
    Jarvis, K. E.
    GEOCHIMICA ET COSMOCHIMICA ACTA, 2006, 70 (18) : A361 - A361
  • [39] Capabilities of inductively coupled plasma mass spectrometry for the measurement of Fe isotope ratios
    Vanhaecke, Frank
    Balcaen, Lieve
    De Wannemacker, Günther
    Moens, Luc
    Journal of Analytical Atomic Spectrometry, 2002, 17 (08): : 933 - 943
  • [40] SAMPLE INTRODUCTION TECHNIQUES FOR THE DETERMINATION OF OSMIUM ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY
    GREGOIRE, DC
    ANALYTICAL CHEMISTRY, 1990, 62 (02) : 141 - 146