Determination of glucose in whole blood samples by mid-infrared spectroscopy

被引:40
|
作者
Kim, YJ [1 ]
Hahn, S [1 ]
Yoon, G [1 ]
机构
[1] Samsung Adv Inst Technol, Med Applicat Team, Suwon 440600, South Korea
关键词
D O I
10.1364/AO.42.000745
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We have determined the glucose concentration of whole blood from mid-infrared spectra without sample preparation or use of chemical reagents. We selected 1119-1022 cm(-1) as the optimal wavelength range for our measurement by making a first-loading vector analysis based on partial least-squares regression. We examined the influence of hemoglobin on samples by using different calibration and prediction sets. The accuracy of glucose prediction depended on the hemoglobin level in the calibration model; the sample set should represent the entire range of hemoglobin concentration. We obtained an accuracy of 5.9% in glucose prediction, and this value is well within a clinically acceptable range. (C) 2003 Optical Society of America.
引用
收藏
页码:745 / 749
页数:5
相关论文
共 50 条
  • [31] Mid-infrared spectroscopy coupled with chemometrics for quantitative determination of biomaterial activity
    Cao, Hao
    Gu, Youlin
    Hu, Yihua
    Wang, Xinyu
    Ding, Wanying
    Chen, Guolong
    Chen, Xi
    He, Haihao
    OPTIK, 2023, 281
  • [32] Validation of chemometric models for the determination of deoxynivalenol on maize by mid-infrared spectroscopy
    G. Kos
    H. Lohninger
    R. Krska
    Mycotoxin Research, 2003, 19 (2) : 149 - 153
  • [33] Data for the Determination of Total Carbon in Biosolids using MID-Infrared Spectroscopy
    Albuquerque, N.
    Meehan, B.
    Hughes, J.
    Surapaneni, A.
    DATA IN BRIEF, 2020, 30
  • [34] Determination of Organic and Inorganic Carbon in Forest Soil Samples by Mid-Infrared Spectroscopy and Partial Least Squares Regression
    Tatzber, Michael
    Mutsch, Franz
    Mentler, Axel
    Leitgeb, Ernst
    Englisch, Michael
    Gerzabek, Martin H.
    APPLIED SPECTROSCOPY, 2010, 64 (10) : 1167 - 1175
  • [35] Recent Advances in the Determination of Milk Adulterants and Contaminants by Mid-Infrared Spectroscopy
    Ceniti, Carlotta
    Spina, Anna Antonella
    Piras, Cristian
    Oppedisano, Francesca
    Tilocca, Bruno
    Roncada, Paola
    Britti, Domenico
    Morittu, Valeria Maria
    FOODS, 2023, 12 (15)
  • [36] Research on Determination of Nitrogen Content in Petroleum Using Mid-infrared Spectroscopy
    Li Jingyan
    Chu Xiaoli
    Tian Songbai
    Lu Wanzhen (SINOPEC Research Institute of Petroleum Processing
    ChinaPetroleumProcessing&PetrochemicalTechnology, 2011, 13 (04) : 1 - 7
  • [37] Research on Determination of Nitrogen Content in Petroleum Using Mid-infrared Spectroscopy
    Li Jingyan
    Chu Xiaoli
    Tian Songbai
    Lu Wanzhen
    CHINA PETROLEUM PROCESSING & PETROCHEMICAL TECHNOLOGY, 2011, 13 (04) : 1 - 7
  • [38] Direct determination of nitrate concentration in environmental systems by mid-infrared spectroscopy
    Linker, R
    Shaviv, A
    Shmulevich, I
    LAND AND WATER MANAGEMENT: DECISION TOOLS AND PRACTICES, VOLS 1 AND 2, 2004, : 945 - 953
  • [39] Nitrate determination using anion exchange membrane and mid-infrared spectroscopy
    Linker, Raphael
    Shaviv, Avi
    APPLIED SPECTROSCOPY, 2006, 60 (09) : 1008 - 1012
  • [40] Determination of SARA fractions in asphalts by mid-infrared spectroscopy and multivariate calibration
    Li, Jin
    Xing, Xinyuan
    Hou, Xiangdao
    Wang, Tao
    Wang, Jiayu
    Xiao, Feipeng
    MEASUREMENT, 2022, 198