Determination of glucose in whole blood samples by mid-infrared spectroscopy

被引:40
|
作者
Kim, YJ [1 ]
Hahn, S [1 ]
Yoon, G [1 ]
机构
[1] Samsung Adv Inst Technol, Med Applicat Team, Suwon 440600, South Korea
关键词
D O I
10.1364/AO.42.000745
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We have determined the glucose concentration of whole blood from mid-infrared spectra without sample preparation or use of chemical reagents. We selected 1119-1022 cm(-1) as the optimal wavelength range for our measurement by making a first-loading vector analysis based on partial least-squares regression. We examined the influence of hemoglobin on samples by using different calibration and prediction sets. The accuracy of glucose prediction depended on the hemoglobin level in the calibration model; the sample set should represent the entire range of hemoglobin concentration. We obtained an accuracy of 5.9% in glucose prediction, and this value is well within a clinically acceptable range. (C) 2003 Optical Society of America.
引用
收藏
页码:745 / 749
页数:5
相关论文
共 50 条
  • [1] Application of molecular spectroscopy in the mid-infrared region to the determination of glucose and cholesterol in whole blood and in blood serum
    Budinova, G
    Salva, J
    Volka, K
    APPLIED SPECTROSCOPY, 1997, 51 (05) : 631 - 635
  • [2] Quantitation of glucose and urea in whole blood by mid-infrared spectroscopy of dry films
    Low-Ying, S
    Shaw, RA
    Leroux, M
    Mantsch, HH
    VIBRATIONAL SPECTROSCOPY, 2002, 28 (01) : 111 - 116
  • [3] Application of mid-infrared transmission spectrometry to the direct determination of glucose in whole blood
    Vonach, R
    Buschmann, J
    Falkowski, R
    Schindler, R
    Lendl, B
    Kellner, R
    APPLIED SPECTROSCOPY, 1998, 52 (06) : 820 - 822
  • [4] Determination of protein in ground wheat samples by mid-infrared diffuse reflectance spectroscopy
    Reeves, JB
    DelWiche, SR
    APPLIED SPECTROSCOPY, 1997, 51 (08) : 1200 - 1204
  • [5] Non-invasive Blood Glucose Measurement Based on mid-Infrared Spectroscopy
    Chen, Yang
    Liu, Jiang
    Pan, Zhenni
    Shimamoto, Shirgeru
    2020 IEEE 17TH ANNUAL CONSUMER COMMUNICATIONS & NETWORKING CONFERENCE (CCNC 2020), 2020,
  • [6] Determination of lignin in tobacco by mid-infrared spectroscopy
    Kong, Haohui, 1600, Editorial Office of Tobacco Science and Technology
  • [7] Determination of the Recovered-Fiber Content in Paperboard Samples by Applying Mid-Infrared Spectroscopy
    Riba, Jordi-Roger
    Canals, Trini
    Cantero, Rosa
    APPLIED SPECTROSCOPY, 2015, 69 (04) : 442 - 450
  • [8] Evaluation of Opportunities and Limitations of Mid-Infrared Skin Spectroscopy for Noninvasive Blood Glucose Monitoring
    Delbeck, Sven
    Heise, H. Michael
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2021, 15 (01): : 19 - 27
  • [9] Mid-infrared spectroscopy of alkali feldspar samples for space application
    Reitze, Maximilian P.
    Weber, Iris
    Kroll, Herbert
    Morlok, Andreas
    Hiesinger, Harald
    Helbert, Joern
    MINERALOGY AND PETROLOGY, 2020, 114 (05) : 453 - 463
  • [10] Mid-infrared spectroscopy of alkali feldspar samples for space application
    Maximilian P. Reitze
    Iris Weber
    Herbert Kroll
    Andreas Morlok
    Harald Hiesinger
    Jörn Helbert
    Mineralogy and Petrology, 2020, 114 : 453 - 463