Calibration of colloidal probes with atomic force microscopy for micromechanical assessment

被引:14
|
作者
Kain, Lukas [1 ]
Andriotis, Orestis G. [1 ,5 ]
Gruber, Peter [2 ,5 ]
Frank, Martin [1 ,5 ]
Markovic, Marica [2 ,5 ]
Grech, David [3 ]
Nedelkovski, Vedran [1 ,5 ]
Stolz, Martin [4 ]
Ovsianikov, Aleksandr [2 ,5 ]
Thurner, Philipp J. [1 ,5 ]
机构
[1] TU Wien, Inst Lightweight Design & Struct Biomech, Getreidemarkt 9, A-1060 Vienna, Austria
[2] TU Wien, Inst Mat Sci & Technol, A-1060 Vienna, Austria
[3] Univ Southampton, Dept Elect & Comp Sci, Nano Res Grp, Southampton SO17 1BJ, Hants, England
[4] Univ Southampton, Fac Engn & Environm, Natl Ctr Adv Tribol Southampton, Southampton, Hants, England
[5] Austrian Cluster Tissue Regenerat, A-1200 Vienna, Austria
关键词
Atomic force microscopy; Colloidal probes; Indentation; Microscale; Soft matter; Biological tissues; BIOLOGICAL-MATERIALS; ARTICULAR-CARTILAGE; ELASTIC-MODULUS; INDENTATION; MORPHOGENESIS;
D O I
10.1016/j.jmbbm.2018.05.026
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Mechanical assessment of biological materials and tissue-engineered scaffolds is increasingly focusing at lower length scale levels. Amongst other techniques, atomic force microscopy (AFM) has gained popularity as an instrument to interrogate material properties, such as the indentation modulus, at the microscale via cantilever-based indentation tests equipped with colloidal probes. Current analysis approaches of the indentation modulus from such tests require the size and shape of the colloidal probe as well as the spring constant of the cantilever. To make this technique reproducible, there still exist the challenge of proper calibration and validation of such mechanical assessment. Here, we present a method to (a) fabricate and characterize cantilevers with colloidal probes and (b) provide a guide for estimating the spring constant and the sphere diameter that should be used for a given sample to achieve the highest possible measurement sensitivity. We validated our method by testing agarose samples with indentation moduli ranging over three orders of magnitude via AFM and compared these results with bulk compression tests. Our results show that quantitative measurements of indentation modulus is achieved over three orders of magnitude ranging from 1 kPa to 1000 kPa via AFM cantilever-based microindentation experiments. Therefore, our approach could be used for quantitative micromechanical measurements without the need to perform further validation via bulk compression experiments.
引用
收藏
页码:225 / 236
页数:12
相关论文
共 50 条
  • [1] Large colloidal probes for atomic force microscopy: Fabrication and calibration issues
    Chighizola, Matteo
    Puricelli, Luca
    Bellon, Ludovic
    Podesta, Alessandro
    JOURNAL OF MOLECULAR RECOGNITION, 2021, 34 (01)
  • [2] Accurate noncontact calibration of colloidal probe sensitivities in atomic force microscopy
    Chung, Koo-Hyun
    Shaw, Gordon A.
    Pratt, Jon R.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2009, 80 (06):
  • [3] Nanomechanical and topographical imaging of living cells by atomic force microscopy with colloidal probes
    Puricelli, Luca
    Galluzzi, Massimiliano
    Schulte, Carsten
    Podesta, Alessandro
    Milani, Paolo
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2015, 86 (03):
  • [4] Improved in situ spring constant calibration for colloidal probe atomic force microscopy
    McBride, Sean P.
    Law, Bruce M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2010, 81 (11):
  • [5] Lateral Force Calibration: Accurate Procedures for Colloidal Probe Friction Measurements in Atomic Force Microscopy
    Chung, Koo-Hyun
    Pratt, Jon R.
    Reitsma, Mark G.
    LANGMUIR, 2010, 26 (02) : 1386 - 1394
  • [6] Note: Production of stable colloidal probes for high-temperature atomic force microscopy applications
    Ditscherlein, L.
    Peuker, U. A.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2017, 88 (04):
  • [7] Note: Mechanically and chemically stable colloidal probes from silica particles for atomic force microscopy
    Kuznetsov, V.
    Papastavrou, G.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2012, 83 (11):
  • [8] Lateral Force Calibration in Atomic Force Microscopy: Minireview
    Wang, Huabin
    SCIENCE OF ADVANCED MATERIALS, 2017, 9 (01) : 56 - 64
  • [9] Evaluating probes for "electrical" atomic force microscopy
    Trenkler, T
    Hantschel, T
    Stephenson, R
    De Wolf, P
    Vandervorst, W
    Hellemans, L
    Malavé, A
    Büchel, D
    Oesterschulze, E
    Kulisch, W
    Niedermann, P
    Sulzbach, T
    Ohlsson, O
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 2000, 18 (01): : 418 - 427
  • [10] Atomic force microscopy probes go electrochemical
    Gardner, CE
    Macpherson, JV
    ANALYTICAL CHEMISTRY, 2002, 74 (21) : 576A - 584A