An oesophageal pulse oximetry system utilising a fibre-optic probe

被引:3
|
作者
Phillips, J. P. [1 ]
Langford, R. M. [2 ]
Chang, S. H. [2 ]
Maney, K. [2 ]
Kyriacou, P. A. [1 ]
Jones, D. P. [1 ]
机构
[1] City Univ London, Sch Engn & Math Sci, London EC1V 0HB, England
[2] St Bartholomew s Hosp, Anaesthet Lab, London EC1A 7BE, England
来源
关键词
BLOOD-OXYGEN SATURATION; CARDIOTHORACIC SURGERY;
D O I
10.1088/1742-6596/178/1/012021
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A dual-wavelength fibre-optic pulse oximetry system is described for the purposes of estimating oxygen saturation (SPO2) from the oesophagus. A probe containing miniature right-angled glass prisms was used to record photoplethysmographic (PPG) signals from the oesophageal wall. Signals were recorded successfully in 19 of 20 patients, demonstrating that PPG signals could be reliably obtained from an internal vascularised tissue site such as the oesophageal epithelium. The value of the mean oxygen saturation recorded from the oesophagus was 94.0 +/- 4.0%. These results demonstrate that SPO2 may be estimated in the oesophagus using a fibre-optic probe.