Realization of total spectral radiant flux scale at NMIJ with a goniophotometer/spectroradiometer

被引:8
|
作者
Godo, Kenji [1 ]
Niwa, Kazuki [1 ]
Kinoshita, Kenichi [1 ]
Ichino, Yoshiro [1 ]
Zama, Tatsuya [1 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, Natl Metrol Inst Japan NMIJ, 1-1-1 Umezono, Tsukuba, Ibaraki, Japan
关键词
total spectral radiant flux; solid state lighting; total luminous flux; goniophotometer/spectroradiometer;
D O I
10.1088/0026-1394/53/2/853
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In response to the strong demand for a total spectral radiant flux (TSRF) standard from domestic lighting manufacturers, such a scale has been realized in the visible range by means of a relative gonio-spectroradiometric method at the National Metrology Institute of Japan (NMIJ). Our gonio-spectroradiometric method employs spectral irradiance as well as a luminous intensity standard as reference standards. We have investigated several models of quartz-halogen lamps from domestic manufacturers with respect to their stability and decided a set of reference standard lamps for TSRF. Our carefully selected quartz-halogen lamps have sufficient stability as the standard lamp for TSRF after a 100 h seasoning process. The relative expanded uncertainty (k = 2) for realization of the TSRF scale is between 3.1% (visible regions) and 4.1% (near ultraviolet region). We evaluated uncertainties related to the characteristics of the array spectroradiometer using experimental results and found some of those, such as effect of bandpass function, noticeably contributed to the total uncertainty.
引用
收藏
页码:853 / 859
页数:7
相关论文
共 50 条
  • [1] Methods of total spectral radiant flux realization at VNIIOFI
    Ivashin, Evgeniy
    Lalek, Jan
    Rybczynski, Andrzej
    Ogarev, Sergey
    Khlevnoy, Boris
    Dobroserdov, Dmitry
    Sapritsky, Victor
    [J]. 13TH INTERNATIONAL CONFERENCE ON NEW DEVELOPMENTS AND APPLICATIONS IN OPTICAL RADIOMETRY, 2018, 972
  • [2] Traceability of Total Spectral Radiant Flux (TSRF) Scale Using Spectral Irradiance and Total Luminous Flux Scale at CSIR-NPL, India
    Naval Vijeta
    Shibu Manchalwar
    V. K. Saha
    Parag Jaiswal
    [J]. MAPAN, 2023, 38 : 629 - 640
  • [3] Traceability of Total Spectral Radiant Flux (TSRF) Scale Using Spectral Irradiance and Total Luminous Flux Scale at CSIR-NPL, India
    Manchalwar, Naval
    Saha, Shibu
    Jaiswal, V. K.
    Sharma, Parag
    [J]. MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2023, 38 (03): : 629 - 640
  • [4] The realization of spectral total radiance flux scale and related calibration service at NIM
    Zhao Weiqiang
    Liu Hui
    Liu Jian
    [J]. PROCEEDINGS OF 2015 IEEE 12TH INTERNATIONAL CONFERENCE ON ELECTRONIC MEASUREMENT & INSTRUMENTS (ICEMI), VOL. 1, 2015, : 294 - 298
  • [5] REALIZATION OF THE NML SCALE OF TOTAL LUMINOUS FLUX
    BROWN, WJ
    [J]. METROLOGIA, 1979, 15 (04) : 167 - 171
  • [6] Realization of an infrared spectral radiant power response scale on a cryogenic bolometer
    Migdall, A
    Eppeldauer, G
    [J]. METROLOGIA, 1998, 35 (04) : 307 - 315
  • [7] COMPARISON ON TOTAL LUMINOUS FLUX MEASUREMENT OF SPECTROGONIOPHOTOMETER AND GONIOPHOTOMETER
    Zhao, W.
    Hui, L.
    Jian, L.
    Haisu, Z.
    Tong, C.
    [J]. PROCEEDINGS OF THE CIE CENTENARY CONFERENCE TOWARDS A NEW CENTURY OF LIGHT, 2013, : 634 - 636
  • [8] Establishment of 2π total spectral radiant flux scale with a broadband LED-based transfer standard source
    Nakazawa, Yuri
    Godo, Kenji
    Niwa, Kazuki
    Zama, Tatsuya
    Yamaji, Yoshiki
    Matsuoka, Shinya
    [J]. METROLOGIA, 2020, 57 (06)
  • [9] MEASUREMENT OF ANGULAR NONUNIFORMITY OF AN INTEGRATION SPHERE FOR TOTAL SPECTRAL RADIANT FLUX MEASUREMENT
    Niwa, K.
    Godo, K.
    Zama, T.
    Zong, Y.
    Ohno, Y.
    [J]. PROCEEDINGS OF CIE 2012 LIGHTING QUALITY AND ENERGY EFFICIENCY, 2012, : 659 - 662
  • [10] INTEGRATING SPHERE SIMULATION - APPLICATION TO TOTAL FLUX SCALE REALIZATION
    OHNO, Y