Requirements and conceptual superconducting magnet design fora 21 T Fourier Transform Ion Cyclotron Resonance Mass Spectrometer

被引:7
|
作者
Painter, Thomas A. [1 ]
Markiewicz, W. Denis
Miller, John R.
Bole, Scott T.
Dixon, Lain R.
Cantrell, Kurt R.
Kenney, Steve J.
Trowell, Amy J.
Kim, Dong Lak
Lee, Byoung Seob
Choi, Yeon Suk
Kim, Hyun Sik
Hendrickson, Christopher L.
Marshall, Alan G.
机构
[1] Natl High Magnet Field Lab, Tallahassee, FL 32310 USA
[2] Korea Basic Sci Inst, Taejon 305333, South Korea
关键词
FT-ICR; superconducting magnets; superfluid helium;
D O I
10.1109/TASC.2005.869678
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The National High Magnetic Field Laboratory (NHMFL) has identified as one of its next high-field persistent-magnet targets a 21 T superconducting magnet system to be used in a Fourier Transform Ion Cyclotron Resonance (FT-ICR) Mass Spectrometer. The required fundamental parameters of a 21 T central field and 110 nun warm bore diameter are similar to those already successfully demonstrated in the NHMFL's Ultra-Wide Bore 900 MHz Nuclear Magnetic Resonance magnet system. However, there are several ancillary requirements of the FT-ICR spectrometer that require technical attention such as fringe field restrictions requiring shielding, length of uniformity zone, magnitude of inhontogeneity, horizontal orientation of the magnet axis and a limitation on the length of the cryostat. This paper describes the overall system requirements, features and status of the superconducting magnet conceptual design proposed for meeting these requirements.
引用
收藏
页码:945 / 948
页数:4
相关论文
共 50 条
  • [31] Sidebands in Fourier transform ion cyclotron resonance mass spectra
    Miladinovic, Sasa M.
    Kozhinov, Anton N.
    Tsybin, Oleg Yu.
    Tsybin, Yury O.
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2012, 325 : 10 - 18
  • [33] Ion discrimination during ion accumulation in a quadrupole interface external to a Fourier transform ion cyclotron resonance mass spectrometer
    Belov, ME
    Nikolaev, EN
    Harkewicz, R
    Masselon, CD
    Alving, K
    Smith, RD
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2001, 208 (1-3) : 205 - 225
  • [34] A trapped ion mobility enabled Fourier transform ion cyclotron resonance mass spectrometer for infrared ion spectroscopy at FELIX
    Houthuijs, Kas J.
    van Tetering, Lara
    Schuurman, Jelle L.
    Wootton, Christopher A.
    Gebhardt, Christoph R.
    Ridgeway, Mark E.
    Berden, Giel
    Martens, Jonathan
    Oomens, Jos
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2024, 505
  • [35] Fourier transform ion cyclotron resonance mass spectrometry at the true cyclotron frequency
    Nagornov, Konstantin O.
    Tsybin, Oleg Y.
    Nicol, Edith
    Kozhinov, Anton N.
    Tsybin, Yury O.
    MASS SPECTROMETRY REVIEWS, 2022, 41 (02) : 314 - 337
  • [36] A low-noise, wideband preamplifier for a Fourier-transform ion cyclotron resonance mass spectrometer
    Mathur, Raman
    Knepper, Ronald W.
    O'Connor, Peter B.
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2007, 18 (12) : 2233 - 2241
  • [37] ULTRAHIGH-RESOLUTION FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETER
    ALBER, GM
    MARSHALL, AG
    HILL, NC
    SCHWEIKHARD, L
    RICCA, TL
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (07): : 1845 - 1852
  • [38] A Low-Noise, Wideband Preamplifier for a Fourier-Transform Ion Cyclotron Resonance Mass Spectrometer
    Mathur, Raman
    Knepper, Ronald W.
    O'Connor, Peter B.
    Journal of the American Society for Mass Spectrometry, 2007, 18 (12): : 2233 - 2241
  • [39] 5.6 Tesla Fourier transform ion cyclotron resonance mass spectrometer for analysis of volatile complex mixtuues
    Rodgers, RP
    White, FM
    McIntosh, DG
    Marshall, AG
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (06): : 2278 - 2284
  • [40] Sharing a pair of electrodes between excitation and detection in a Fourier transform ion cyclotron resonance mass spectrometer
    Laboratoire de Physico-Chimie des Rayonnements, Université de Paris-Sud, 91405 Orsay, France
    Rapid Commun. Mass Spectrom., 5 (591-593):