New 3D Silicon detectors for dosimetry in Microbeam Radiation Therapy

被引:8
|
作者
Lerch, M. L. F. [1 ]
Dipuglia, A. [1 ]
Cameron, M. [1 ]
Fournier, P. [1 ]
Davis, J. [1 ]
Petasecca, M. [1 ]
Cornelius, I. [1 ]
Perevertaylo, V. [2 ]
Rosenfeld, A. B. [1 ]
机构
[1] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
[2] SPA BIT, Kiev, Ukraine
基金
澳大利亚国家健康与医学研究理事会;
关键词
D O I
10.1088/1742-6596/777/1/012009
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Microbeam Radiation Therapy (MRT) involves the use of a spatially fractionated beam of synchrotron generated X-rays to treat tumours. MRT treatment is delivered via an array of high dose 'peaks' separated by low dose 'valleys'. A good Peak to Valley Dose Ratio (PVDR) is an important indicator of successful treatment outcomes. MRT dosimetry requires a radiation hard detector with high spatial resolution, large dynamic range, which is ideally real-time and tissue equivalent. We have developed a Silicon Strip Detector (SSD) and very recently, a new 3D MESA SSD to meet the very stringent requirements of MRT dosimetry. We have compared these detectors through the characterisation of the MRT radiation field at the Australian Synchrotron Imaging and Medical Beamline. The EPI SSD was able to measure the microbeam profiles and PVDRs, however the effective spatial resolution was limited by the detector alignment options available at the time. The geometry of the new 3D MESA SSD is less sensitive to this alignment restriction was able to measure the microbeam profiles within 2 mu m of that expected. The 3D MESA SSD measured PVDRs were possibly affected by undesired and slow charge collection outside the sensitive volume and additional scattering from the device substrate.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] High resolution 3D dosimetry for microbeam radiation therapy using optical CT
    McErlean, C.
    Braeuer-Krisch, E.
    Adamovics, J.
    Leach, M. O.
    Doran, S. J.
    [J]. 8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573
  • [2] Silicon planar structures as detectors for microbeam radiation therapy
    Anokhin, Igor E.
    Lerch, M.
    Petasecca, M.
    Zinets, O.
    Rosenfeld, A.
    [J]. 2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [3] Progress in 3D Silicon Radiation Detectors
    Dalla Betta, Gian-Franco
    Povoli, Marco
    [J]. Frontiers in Physics, 2022, 10
  • [4] Progress in 3D Silicon Radiation Detectors
    Dalla Betta, Gian-Franco
    Povoli, Marco
    [J]. FRONTIERS IN PHYSICS, 2022, 10
  • [5] Silicon semi 3D radiation detectors
    Eränen, S
    Virolainen, T
    Luusua, I
    Kalliopuska, J
    Kurvinen, K
    Eräluoto, M
    Härkönen, J
    Leinonen, K
    Palviainen, M
    Koski, M
    [J]. 2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 1231 - 1235
  • [6] Characterization of 3-D-Mesa Silicon Single Strip Detectors for Use in Synchrotron Microbeam Radiation Therapy
    Cameron, Matthew J.
    Davis, Jeremy A.
    Dipuglia, Andrew
    Chartier, Lachlan
    Tran, Linh T.
    Prokopovich, Dale A.
    Petasecca, Marco
    Perevertaylo, Vladimir L.
    Rosenfeld, Anatoly B.
    Lerch, Michael L. F.
    [J]. IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2020, 4 (04) : 470 - 478
  • [7] Simulation and test of 3D silicon radiation detectors
    Fleta, C.
    Pennicard, D.
    Bates, R.
    Parkes, C.
    Pellegrini, G.
    Lozano, M.
    Wright, V.
    Boscardin, M.
    Betta, G.-F. Dalla
    Piemonte, C.
    Pozza, A.
    Ronchin, S.
    Zorzi, N.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 579 (02): : 642 - 647
  • [8] Clinical 3D Dosimetry in Modern Radiation Therapy
    Mihailidis, Dimitris
    [J]. MEDICAL PHYSICS, 2020, 47 (01) : 290 - 291
  • [9] Preliminary Characterization of Microbeam Radiation Using Very High Resolution 3D Dosimetry
    Juang, T.
    Ger, R.
    Li, Q.
    Hadsell, M.
    Adamovics, J.
    Zhou, O.
    Oldham, M.
    Chang, S.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [10] Dosimetry of microbeam radiotherapy by flexible hydrogenated amorphous silicon detectors
    Large, Matthew James
    Kanxheri, Keida
    Posar, Jessie
    Aziz, Saba
    Bashiri, Aishah
    Calcagnile, Lucio
    Calvo, Daniela
    Caputo, Domenico
    Caricato, Anna Paola
    Catalano, Roberto
    Cirio, Roberto
    Cirrone, Giuseppe Antonio Pablo
    Croci, Tommaso
    Cuttone, Giacomo
    De Cesare, Gianpiero
    De Remigis, Paolo
    Dunand, Sylvain
    Fabi, Michele
    Frontini, Luca
    Grimani, Catia
    Guarrera, Mariacristina
    Ionica, Maria
    Lenta, Francesca
    Liberali, Valentino
    Lovecchio, Nicola
    Martino, Maurizio
    Maruccio, Giuseppe
    Mazza, Giovanni
    Menichelli, Mauro
    Monteduro, Anna Grazia
    Morozzi, Arianna
    Moscatelli, Francesco
    Nascetti, Augusto
    Pallotta, Stefania
    Passeri, Daniele
    Pedio, Maddalena
    Petringa, Giada
    Peverini, Francesca
    Placidi, Pisana
    Quarta, Gianluca
    Rizzato, Silvia
    Sabbatini, Federico
    Servoli, Leonello
    Stabile, Alberto
    Thomet, Jonathan Emanuel
    Tosti, Luca
    Villani, Mattia
    Wheadon, Richard James
    Wyrsch, Nicolas
    Zema, Nicola
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (15):