Ion Mobility-High-Resolution Mass Spectrometry (IM-HRMS) for the Analysis of Contaminants of Emerging Concern (CECs): Database Compilation and Application to Urine Samples

被引:33
|
作者
Belova, Lidia [1 ]
Caballero-Casero, Noelia [1 ]
van Nuijs, Alexander L. N. [1 ]
Covaci, Adrian [1 ]
机构
[1] Univ Antwerp, Toxicol Ctr, B-2610 Antwerp, Belgium
关键词
COLLISION CROSS-SECTION; FLAME RETARDANTS; STRUCTURAL-CHARACTERIZATION; IDENTIFICATION; SEPARATION; DUST;
D O I
10.1021/acs.analchem.1c00142
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Ion mobility mass spectrometry (IM-MS)-derived collision cross section (CCS) values can serve as a valuable additional identification parameter within the analysis of compounds of emerging concern (CEC) in human matrices. This study introduces the first comprehensive database of (CCSN2)-C-DT values of 148 CECs and their metabolites including bisphenols, alternative plasticizers (AP), organophosphate flame retardants (OP), perfluoroalkyl chemicals (PFAS), and others. A total of 311 ions were included in the database, whereby the (CCSN2)-C-DT values for 113 compounds are reported for the first time. For 105 compounds, more than one ion is reported. Moreover, the (CCSN2)-C-DT values of several isomeric CECs and their metabolites are reported to allow a distinction between isomers. Comprehensive quality assurance guidelines were implemented in the workflow of acquiring (CCSN2)-C-DT values to ensure reproducible experimental conditions. The reliability and reproducibility of the complied database were investigated by analyzing pooled human urine spiked with 30 AP and OP metabolites at two concentration levels. For all investigated metabolites, the (CCSN2)-C-DT values measured in urine showed a percent error of <1% in comparison to database values. (CCSN2)-C-DT values of OP metabolites showed an average percent error of 0.12% (50 ng/mL in urine) and 0.15% (20 ng/mL in urine). For AP metabolites, these values were 0.10 and 0.09%, respectively. These results show that the provided database can be of great value for enhanced identification of CECs in environmental and human matrices, which can advance future suspect screening studies on CECs.
引用
收藏
页码:6428 / 6436
页数:9
相关论文
共 6 条
  • [1] Analysis of cranberry proanthocyanidins using UPLC-ion mobility-high-resolution mass spectrometry
    Wang, Yifei
    Harrington, Peter de B.
    Chang, Tony
    Wu, Xianli
    Chen, Pei
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2020, 412 (15) : 3653 - 3662
  • [2] Comprehensive suspect screening for the identification of contaminants of emerging concern in urine of Flemish adolescents by liquid chromatography high-resolution mass spectrometry
    Roggeman, Maarten
    Belova, Lidia
    Fernandez, Sandra F.
    Kim, Da-Hye
    Jeong, Yunsun
    Poma, Giulia
    Remy, Sylvie
    Verheyen, Veerle J.
    Schoeters, Greet
    van Nuijs, Alexander L. N.
    Covaci, Adrian
    [J]. ENVIRONMENTAL RESEARCH, 2022, 214
  • [3] Direct Infusion Electrospray Ionization - Ion Mobility - High Resolution Mass Spectrometry (DIESI-IM-HRMS) for Rapid Characterization of Potential Bioprocess Streams
    Munisamy, Sharon M.
    Chambliss, C. Kevin
    Becker, Christopher
    [J]. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2012, 23 (07) : 1250 - 1259
  • [4] Rethinking Drug Analysis in Health Care: High-Throughput Analysis of 71 Drugs of Abuse in Oral Fluid Using Ion Mobility-High-Resolution Mass Spectrometry
    Bergstrom, Moa Andresen
    Lovgren, Hanna
    Abrahamsson, Anna
    Eriksson, Emma K.
    Andersson, Maria Lindbjer
    Komorowska, Marta
    Axelsson, Magnus A. B.
    [J]. JOURNAL OF ANALYTICAL TOXICOLOGY, 2022, 46 (07) : 765 - 775
  • [5] Analysis of emerging contaminants in water and solid samples using high resolution mass spectrometry with a Q Exactive orbital ion trap and estrogenic activity with YES-assay
    Comtois-Marotte, Simon
    Chappuis, Thomas
    Sung Vo Duy
    Gilbert, Nicolas
    Lajeunesse, Andre
    Taktek, Salma
    Desrosiers, Melanie
    Veilleux, Eloise
    Sauve, Sebastien
    [J]. CHEMOSPHERE, 2017, 166 : 400 - 411
  • [6] Resolving the microcosmos of complex samples: UPLC/travelling wave ion mobility separation high resolution mass spectrometry for the analysis of in vivo drug metabolism studies