A robotic protocol for high-throughput processing of samples for selected reaction monitoring assays

被引:10
|
作者
Zhu, Min [1 ]
Zhang, Pingbo [1 ,2 ]
Geng-Spyropoulos, Minghui [2 ]
Moaddel, Ruin [1 ]
Semba, Richard D. [2 ]
Ferrucci, Luigi [1 ]
机构
[1] Natl Inst Aging, Baltimore, MD 21224 USA
[2] Johns Hopkins Univ, Sch Med, Wilmer Eye Inst, Baltimore, MD USA
关键词
Absolute quantification; Automated sample preparation; Heavy isotope-labeled peptides; Reproducibility; SRM; Technology; PROTEOMICS; PLASMA;
D O I
10.1002/pmic.201600339
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Selected reaction monitoring mass spectrometry (SRM-MS) is a sensitive and accurate method for the quantification of targeted proteins in biological specimens. However, the sample throughput and reliability of this technique is limited by the complexity of sample preparation, as well as instrumentation and data processing. Modern robotic equipment allows for rapid and accurate processing of large number of samples and makes SRM-MS assay applicable in epidemiological studies. Herein, we describe an automated sample processing platform developed in the context of an SRM-MS protocol for the assay of complement factor H protein and its variants in human plasma. We report detailed performance data on plasma digestion, sample cleanup and optimized robotic handling implemented on a Biomek (R) NXp Workstation. Method validation was assessed with isotopically labeled peptide standards and had high reproducibility of intra-day assay (CVs from 2.7 to 17.5% with median CV of 5.3%) and inter-day assay (CVs from 4.8 to 17.6 with median CV of 7.2%) for all peptides.
引用
收藏
页数:5
相关论文
共 50 条
  • [31] Complex phenotypic assays in high-throughput screening
    Clemons, PA
    CURRENT OPINION IN CHEMICAL BIOLOGY, 2004, 8 (03) : 334 - 338
  • [32] High-Throughput Discovery of Aptamers for Sandwich Assays
    Csordas, Andrew T.
    Jorgensen, Anna
    Wang, Jinyeng
    Gruber, Emily
    Gong, Qiang
    Bagley, Elizabeth R.
    Nakamoto, Margaret A.
    Eisenstein, Michael
    Soh, H. Tom
    ANALYTICAL CHEMISTRY, 2016, 88 (22) : 10842 - 10847
  • [33] Importance of CAR-T cell therapy monitoring using high-throughput assays
    Pratt, Chelsea B.
    Alexander, Matt
    DRUG DISCOVERY TODAY, 2022, 27 (10)
  • [34] High-throughput immunoturbidimetric assays for inprocess determination of polyclonal antibody concentration and functionality in crude samples
    Bak, Hanne
    Kyhse-Andersen, Jan
    Thomas, Owen R. T.
    JOURNAL OF BIOTECHNOLOGY, 2007, 131 (02) : S70 - S70
  • [35] A Robotic Platform for Quantitative High-Throughput Screening
    Michael, Sam
    Auld, Douglas
    Klumpp, Carleen
    Jadhav, Ajit
    Zheng, Wei
    Thorne, Natasha
    Austin, Christopher P.
    Inglese, James
    Simeonov, Anton
    ASSAY AND DRUG DEVELOPMENT TECHNOLOGIES, 2008, 6 (05) : 637 - 657
  • [36] Development of a High-Throughput Kinetics Protocol and Application to an Aza-Michael Reaction
    Li, Xiao
    Dunn, Anna L.
    ORGANIC PROCESS RESEARCH & DEVELOPMENT, 2022, 26 (03) : 795 - 803
  • [37] HIGH-THROUGHPUT DIGITAL SAR PROCESSING
    JONES, M
    DEGAVRE, JC
    GUIGNARD, JP
    ESA JOURNAL-EUROPEAN SPACE AGENCY, 1983, 7 (02): : 145 - 162
  • [38] High-throughput screening for downstream processing
    Russell, Laura
    European Pharmaceutical Review, 2024, 29 (01): : 13 - 17
  • [39] High-Throughput Stream Processing with Actors
    Rinaldi, Luca
    Torquati, Massimo
    Mencagli, Gabriele
    Danelutto, Marco
    PROCEEDINGS OF THE 10TH ACM SIGPLAN INTERNATIONAL WORKSHOP ON PROGRAMMING BASED ON ACTORS, AGENTS, AND DECENTRALIZED CONTROL, AGERE 2020, 2020, : 1 - 10
  • [40] AHT-ChIP-seq: a completely automated robotic protocol for high-throughput chromatin immunoprecipitation
    Sarah Aldridge
    Stephen Watt
    Michael A Quail
    Tim Rayner
    Margus Lukk
    Michael F Bimson
    Daniel Gaffney
    Duncan T Odom
    Genome Biology, 14