Investigation of Depth of Interaction Encoding for a Pixelated LSO Array With a Single Multi-Channel PMT

被引:25
|
作者
Yang, Yongfeng [1 ]
Wu, Yibao [1 ]
Cherry, Simon R. [1 ]
机构
[1] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
关键词
Depth of interaction; positron emission tomography (PET); small-animal imaging; SMALL-ANIMAL PET; RESOLUTION RESEARCH TOMOGRAPH; POSITRON-EMISSION-TOMOGRAPHY; PERFORMANCE EVALUATION; DETECTOR MODULE; MICROPET SCANNER; IMAGING-SYSTEM; ECAT HRRT; DESIGN; ACQUISITION;
D O I
10.1109/TNS.2009.2016094
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A new approach to depth or interaction (DOI) encoding for a pixelated LSO array using a single multi-channel PMT was investigated. In this method the DOI information was estimated by taking advantage of optical crosstalk between LSO elements and examining the standard deviation (spread) of signals on all channels of the PMT. Unpolished and polished 6 x 6 LSO arrays with a crystal size of 1.3 x 1.3 x 20 mm(3) were evaluated on a Hamamatsu H7546 64-channel PMT. The arrays were placed on the center of the PMT and the central 16 channels of the PMT were individually read out and digitized. For the unpolished array, all crystals were resolved in the flood histogram. An average DOI resolution of 8 mm was obtained. The energy resolution was similar to 25% after the signal amplitude was corrected using the measured DOI information. For the polished array, the flood histo-ram was superior to the unpolished array, however no DOI information could be measured. Using unpolished crystals, this method could be a practical way to achieve limited DOI information in PET detectors. The standard deviation of all PMT channels can be readily obtained using a resistor network. Only five signals (four signals to determine the x-y position and one signal measuring the standard deviation) need to be digitized, and this method only requires a single photon detector to read out the array. Unlike phoswich detectors, the method does not require segmenting the scintillator array into layers. The measured DOI resolution was much worse than that obtained with the dual-ended readout method, however, it was similar to that obtained with a two-layer phoswich detector.
引用
收藏
页码:2594 / 2599
页数:6
相关论文
共 50 条
  • [1] Evaluation of optical fiber bundles for coupling a small LSO crystal array to a multi-channel PMT
    Chatziioannou, A
    Shao, YP
    Doshi, N
    Meadors, K
    Silverman, R
    Cherry, S
    1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 1483 - 1487
  • [2] PMT connectors for multi-channel film waveguides
    Hatakeyama, Yutaka
    Imamura, Saburo
    Kobayashi, Junya
    Takahara, Hideyuki
    ORGANIC PHOTONIC MATERIALS AND DEVICES XI, 2009, 7213
  • [3] A novel method to determine the depth of interaction position in the pixelated scintillation crystals with single-ended readout by multi-anode PMT
    Ito, Mikiko
    Hong, Seong Jong
    Sim, Kwang-Souk
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [4] Multi-channel optical mapping: Investigation of depth information
    Sase, I
    Eda, H
    Seiyama, A
    Tanabe, HC
    Takatsuki, A
    Yanagida, T
    OPTICAL TOMOGRAPHY AND SPECTROSCOPY OF TISSUE IV, 2001, 4250 : 29 - 36
  • [5] Segmented LSO crystals for depth-of-interaction encoding in PET
    Moisan, C
    Andreaco, MS
    Rogers, JG
    Paquet, S
    Vozza, D
    1997 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1 & 2, 1998, : 1112 - 1116
  • [6] Segmented LSO crystals for depth-of-interaction encoding in PET
    TRIUMF, Vancouver, BC, Canada
    IEEE Trans Nucl Sci, 6 pt 2 (3030-3035):
  • [7] Segmented LSO crystals for depth-of-interaction encoding in PET
    Moisan, C
    Andreaco, MS
    Rogers, JG
    Paquet, S
    Vozza, D
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (06) : 3030 - 3035
  • [8] Multi-channel array processing
    Peterson, Martin W.
    Rickerson, Ronald W.
    Sea Technology, 1991, 32 (03)
  • [9] Evaluation of multi-channel PMT's for readout of scintillator arrays
    Shao, YP
    Cherry, SR
    Siegel, S
    Silverman, RW
    Majewski, S
    1995 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD, VOLS 1-3, 1996, : 1055 - 1059
  • [10] An LSO scintillator array for a PET detector module with depth of interaction measurement
    Huber, JS
    Moses, WW
    Andreaco, MS
    Petterson, O
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (03) : 684 - 688