The long term risk of lead failure in patients with cardiovascular implantable electronic devices undergoing catheter ablation

被引:2
|
作者
Farkowski, Michal M. [1 ]
Maciag, Aleksander [1 ]
Ciszewski, Jan [1 ]
Kowalik, Ilona [1 ]
Syska, Pawel [1 ]
Sterlinski, Maciej [2 ]
Szwed, Hanna [1 ]
Pytkowski, Mariusz [1 ]
机构
[1] Inst Cardiol, Dept Coronary Artery Dis 2, Heart Arrhythmia Ward, Alpejska 42, PL-04628 Warsaw, Poland
[2] Inst Cardiol, Dept Heart Arrhythmia, Warsaw, Poland
关键词
Lead; damage; dysfunction; CIED; ablation; long-term; EXPERT CONSENSUS STATEMENT; RADIOFREQUENCY ABLATION; CARDIOVERTER-DEFIBRILLATOR; ATRIAL-FIBRILLATION; ESC GUIDELINES; COMPLICATIONS; MANAGEMENT; PACEMAKERS; EFFICACY; ARRHYTHMIAS;
D O I
10.1080/14017431.2019.1653489
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. The aim of the study was to assess the long-term influence of catheter ablation (CA) of different arrhythmias on cardiovascular implantable electronic devices (CIED) endocardial leads durability. Design. This was a retrospective case-control study. Ablation protocols and in- or outpatient medical records were reviewed to identify and extract data on adult patients with CIED undergoing a CA. A cohort of patients with hypertrophic cardiomyopathy and implantable cardioverter-defibrillators (ICD) served as a historical control group. The primary endpoint was the diagnosis of lead damage defined as permanent loss of proper function demanding replacement or removal. Results. Among 145 patients n = 177 catheter ablations were performed. Patients' mean age was 66.4 +/- 10.5, 66.1% had an ICD or ICD with cardiac resynchronization function (CRT-D), 18.1% had >1 CA. During median 812 days [IQR 381-1588] of follow-up, there were 11 (6.2%) cases of lead damage in the examined and 13 cases (13%) in the control group, p = 0.054. None of the technical aspects of the CA (indication, type of catheter, transseptal sheath) influenced the primary outcome. Both the number of leads and observation time after CA were significantly related to the risk of endocardial lead damage. Conclusion. This study did not find any significant influence of CA on the long-term durability of CIED endocardial leads. Reported risk factors were consistent with general population of CIED patients.
引用
收藏
页码:323 / 328
页数:6
相关论文
共 50 条
  • [41] Long Term Mortality Among Patients Undergoing Extraction of Infected Cardiac Implantable Electronic Devices is Significant and Could be Related to the Type of Infection
    Tarakji, Khaldoun G.
    Harb, Serge
    Wazni, Oussama
    Hsu, Amy
    Fraser, Thomas
    Gordon, Steven
    Wilkoff, Bruce L.
    CIRCULATION, 2011, 124 (21)
  • [42] Long-term mortality and cardiovascular outcomes in patients with atrial flutter after catheter ablation
    Yugo, Dony
    Chen, Yun-Yu
    Lin, Yenn-Jiang
    Chien, Kuo-Liong
    Chang, Shih-Lin
    Lo, Li-Wei
    Hu, Yu-Feng
    Chao, Tze-Fan
    Chung, Fa-Po
    Liao, Jo-Nan
    Chang, Ting-Yung
    Lin, Chin-Yu
    Tuan, Ta-Chuan
    Kuo, Ling
    Wu, Cheng-, I
    Liu, Chih-Min
    Liu, Shin-Huei
    Li, Cheng-Hung
    Hsieh, Yu-Cheng
    Chen, Shih-Ann
    EUROPACE, 2022, 24 (06): : 970 - 978
  • [43] Long-Term, Single-Centre Observation of Patients with Cardiac Implantable Electronic Devices
    Zaluska, Roman
    Milewska, Anna
    Moumtzoglou, Anastasius
    Grabowski, Marcin
    Drygas, Wojciech
    MEDICINA-LITHUANIA, 2021, 57 (12):
  • [44] Long-Term Complications Related to Cardiac Implantable Electronic Devices
    Simpson, Jamie
    Yoder, Mason
    Christian-Miller, Nathaniel
    Wheat, Heather
    Kovacs, Boldizsar
    Cunnane, Ryan
    Ghannam, Michael
    Liang, Jackson J.
    JOURNAL OF CLINICAL MEDICINE, 2025, 14 (06)
  • [45] Long-term Outcome of Implantable Cardioverter/Defibrillator Lead Failure
    Davidsson, Gustav Arnar
    Jonsdottir, Gudlaug M.
    Oddsson, Hjortur
    Lund, Sigrun H.
    Arnar, David O.
    JAMA INTERNAL MEDICINE, 2020, 180 (02) : 322 - 324
  • [46] Predictors of pacing-dependency in patients with cardiovascular implantable electronic devices
    Grimm, Wolfram
    Grimm, Kathrin
    Greene, Brandon
    Parahuleva, Mariana
    CARDIOLOGY JOURNAL, 2021, 28 (03) : 423 - 430
  • [47] MRI in Patients with Cardiovascular Implantable Electronic Devices and Fractured or Abandoned Leads
    Greenhill, Mark J.
    Rangan, Pooja
    Su, Wilber
    Weiss, J. Peter
    Zawaneh, Michael
    Unzek, Samuel
    Tamarappoo, Balaji
    Indik, Julia
    Tung, Roderick
    Morris, Michael F.
    RADIOLOGY-CARDIOTHORACIC IMAGING, 2024, 6 (03):
  • [48] Automated electronic alert for the detection of infected cardiovascular implantable electronic devices in patients with bacteremia
    Lin, Andrew Y.
    Torriani, Francesca
    Sung, Kevin
    Trefethen, Emily
    Near, Nicholas
    Ho, Gordon
    Pollema, Travis
    Birgersdotter-Green, Ulrika
    HEART RHYTHM, 2024, 21 (07) : 1178 - 1179
  • [49] Percutaneous Microwave Tumor Ablation Is Safe in Patients with Cardiovascular Implantable Electronic Devices: A Single-Institutional Retrospective Review
    Swietlik, John F.
    Longo, Katherine C.
    Knott, Emily A.
    Hinshaw, J. Louis
    Brace, Christopher L.
    Alexander, Marci
    Lubner, Meghan G.
    Wells, Shane A.
    Lee, Fred T., Jr.
    Ziemlewicz, Timothy J.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2019, 30 (03) : 396 - 400
  • [50] Long-Term Outcome of Infective Endocarditis Involving Cardiac Implantable Electronic Devices: Impact of Comorbidities and Lead Extraction
    Durante-Mangoni, Emanuele
    Ursi, Maria Paola
    Andini, Roberto
    Mattucci, Irene
    Della Ratta, Ester E.
    Iossa, Domenico
    Bertolino, Lorenzo
    De Vivo, Stefano
    Manduca, Sabrina
    Torella, Michele
    De Feo, Marisa
    Zampino, Rosa
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (24)