The long term risk of lead failure in patients with cardiovascular implantable electronic devices undergoing catheter ablation

被引:2
|
作者
Farkowski, Michal M. [1 ]
Maciag, Aleksander [1 ]
Ciszewski, Jan [1 ]
Kowalik, Ilona [1 ]
Syska, Pawel [1 ]
Sterlinski, Maciej [2 ]
Szwed, Hanna [1 ]
Pytkowski, Mariusz [1 ]
机构
[1] Inst Cardiol, Dept Coronary Artery Dis 2, Heart Arrhythmia Ward, Alpejska 42, PL-04628 Warsaw, Poland
[2] Inst Cardiol, Dept Heart Arrhythmia, Warsaw, Poland
关键词
Lead; damage; dysfunction; CIED; ablation; long-term; EXPERT CONSENSUS STATEMENT; RADIOFREQUENCY ABLATION; CARDIOVERTER-DEFIBRILLATOR; ATRIAL-FIBRILLATION; ESC GUIDELINES; COMPLICATIONS; MANAGEMENT; PACEMAKERS; EFFICACY; ARRHYTHMIAS;
D O I
10.1080/14017431.2019.1653489
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. The aim of the study was to assess the long-term influence of catheter ablation (CA) of different arrhythmias on cardiovascular implantable electronic devices (CIED) endocardial leads durability. Design. This was a retrospective case-control study. Ablation protocols and in- or outpatient medical records were reviewed to identify and extract data on adult patients with CIED undergoing a CA. A cohort of patients with hypertrophic cardiomyopathy and implantable cardioverter-defibrillators (ICD) served as a historical control group. The primary endpoint was the diagnosis of lead damage defined as permanent loss of proper function demanding replacement or removal. Results. Among 145 patients n = 177 catheter ablations were performed. Patients' mean age was 66.4 +/- 10.5, 66.1% had an ICD or ICD with cardiac resynchronization function (CRT-D), 18.1% had >1 CA. During median 812 days [IQR 381-1588] of follow-up, there were 11 (6.2%) cases of lead damage in the examined and 13 cases (13%) in the control group, p = 0.054. None of the technical aspects of the CA (indication, type of catheter, transseptal sheath) influenced the primary outcome. Both the number of leads and observation time after CA were significantly related to the risk of endocardial lead damage. Conclusion. This study did not find any significant influence of CA on the long-term durability of CIED endocardial leads. Reported risk factors were consistent with general population of CIED patients.
引用
收藏
页码:323 / 328
页数:6
相关论文
共 50 条
  • [1] Long-term risk of cardiovascular implantable electronic device-related infection after catheter ablation of arrhythmia
    Farkowski, Michal M.
    Maciag, Aleksander
    Pytkowski, Mariusz
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (01) : 370 - 370
  • [2] Long-term risk of cardiovascular implantable electronic device-related infection after catheter ablation of atrial fibrillation
    Dinshaw, Leon
    Meyer, Christian
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (01) : 371 - 372
  • [3] Long-term risk of replacement of cardiovascular implantable electronic devices following external cardioversion
    Elgaard, A. F.
    Dinesen, P. T.
    Riahi, S.
    Hansen, J.
    Lundbye-Christensen, S.
    Johansen, J. B.
    Nielsen, J. C.
    Larsen, J. M.
    EUROPEAN HEART JOURNAL, 2022, 43 : 569 - 569
  • [4] Long-term efficacy and safety of radiofrequency catheter ablation of atrial fibrillation in patients with cardiac implantable electronic devices and transvenous leads
    Dinshaw, Leon
    Schaeffer, Benjamin
    Akbulak, Oezgg
    Jularic, Mario
    Hartmann, Jens
    Klatt, Niklas
    Dickow, Jannis
    Gunawardene, Melanie
    Muenkler, Paula
    Hakmi, Samer
    Pecha, Simon
    Sultan, Arian
    Lueker, Jakob
    Pinnschmidt, Hans
    Hoffmann, Boris
    Gosau, Nils
    Eickholt, Christian
    Willems, Stephan
    Steven, Daniel
    Meyer, Christian
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2019, 30 (05) : 679 - 687
  • [5] Long-term monitoring of arrhythmias with cardiovascular implantable electronic devices in patients with cardiac sarcoidosis
    Bakker, Annelies
    Mathijssen, Harold
    Dorland, Galina
    Balt, Jippe C.
    van Dijk, Vincent F.
    Veltkamp, Marcel
    Akdim, Fatima C.
    Grutters, Jan C.
    Post, Martijn
    HEART RHYTHM, 2022, 19 (03) : 352 - 360
  • [6] Long-Term Outcomes of Cardiovascular Implantable Electronic Devices in Patients with Staphylococcus aureus Bacteremia
    Obeid, Karam M.
    Szpunar, Susan
    Khatib, Riad
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (08): : 961 - 965
  • [7] EFFICACY AND FEASIBILITY OF CARDIOVASCULAR IMPLANTABLE ELECTRONIC DEVICES LEAD EXTRACTION IN ELDERLY PATIENTS
    Nagashima, Michio
    An, Yoshimori
    Ando, Kenji
    Soga, Yoshimitsu
    Nobuyoshi, Masakiyo
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 59 (13) : E709 - E709
  • [8] Efficacy and feasibility of cardiovascular implantable electronic devices lead extraction in elderly patients
    Nagashima, M.
    An, Y.
    Fukunaga, M.
    Hiroshima, K.
    Yamashita, K.
    Goya, M.
    EUROPEAN HEART JOURNAL, 2014, 35 : 375 - 375
  • [9] EFFICACY AND FEASIBILITY OF CARDIOVASCULAR IMPLANTABLE ELECTRONIC DEVICES LEAD EXTRACTION IN ELDERLY PATIENTS
    Nagashima, M.
    Goya, M.
    Soga, Y.
    Hiroshima, K.
    Hayashi, K.
    Ohe, M.
    Makihara, Y.
    Niu, H.
    Ando, K.
    Iwabuchi, M.
    Nobuyoshi, M.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2011, 22 : S137 - S138
  • [10] Determining the Safety of Radiofrequency Ablation in Cardiovascular Implantable Electronic Devices
    Jabri, Ahmad
    Butt, Muhammad Umer
    Haddadin, Faris
    Alhuneafat, Laith
    Al Abdouh, Ahmad
    Mhanna, Mohammed
    Kumar, Ashish
    Nasser, Farhan
    Elhamdani, Adee
    Abu Jazar, Deaa
    Yousaf, Adnan
    Hamade, Hani
    Nabeel, Yassar
    Almahameed, Soufian
    CURRENT PROBLEMS IN CARDIOLOGY, 2022, 47 (10)