Distribution Templates of the Fiducial Points in Image-Guided Neurosurgery

被引:17
|
作者
Wang Manning [1 ]
Song Zhijian [1 ]
机构
[1] Fudan Univ, Shanghai Med Sch, Digital Med Res Ctr, Shanghai 200032, Peoples R China
关键词
Distribution template; Image-guided neurosurgery; Point-pair registration; Target registration error; PATIENT REGISTRATION; TARGET REGISTRATION; SURGERY; HEAD; ACCURACY; MR;
D O I
10.1227/01.NEU.0000365827.88888.80
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND: Point-pair registration is widely used in an image-guided neurosurgery system. Poor distribution of the fiducial points leads to an increase in the target registration error (TRE). OBJECTIVE: This study aimed to provide templates consisting of optimized positioning of the fiducial points to reduce the TRE in image-guided neurosurgery. METHODS: We divided the head into 6 regions and provided distribution templates of the fiducial points for each of them. A variable termed TREM(r) was used to express the approximate expected square of the TRE at the target point with a specified distribution of fiducial points. We randomly selected 85 patients from 5 hospitals who underwent image-guided neurosurgery and compared the TREM(r) of the real fiducial points with that of the templates. RESULTS: We grouped the patients by hospitals and regions. The mean TREM(r)s of the templates were much smaller than those of the real fiducial points. In each group, the range of the TREM(r) values of the templates was much smaller than that of the real fiducial points. CONCLUSION: This study provides an easy method to implement a good distribution of the fiducial points to help reduce TRE in image-guided neurosurgery. The templates are simple and exact and can be easily integrated into current workflow.
引用
收藏
页码:143 / 150
页数:8
相关论文
共 50 条
  • [41] Evaluation of errors influencing accuracy in image-guided neurosurgery
    Watanabe Y.
    Fujii M.
    Hayashi Y.
    Kimura M.
    Murai Y.
    Hata M.
    Sugiura A.
    Tsuzaka M.
    Wakabayashi T.
    Radiological Physics and Technology, 2009, 2 (2) : 120 - 125
  • [42] Superimposed holographic image-guided neurosurgery - Technical note
    Ko, K
    JOURNAL OF NEUROSURGERY, 1998, 88 (04) : 777 - 781
  • [43] Integration of functional brain mapping in image-guided neurosurgery
    Rezai, AR
    Mogilner, AY
    Cappell, J
    Hund, M
    Llinas, RR
    Kelly, PJ
    ADVANCES IN STEREOTACTIC AND FUNCTIONAL NEUROSURGERY 12: PROCEEDINGS OF THE 12TH MEETING OF THE EUROPEAN SOCIETY FOR STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, MILAN, 1996, 1997, 12 : 85 - 89
  • [44] Image-guided neurosurgery at Brigham and women's hospital
    Nakajima, S
    Kikinis', R
    Black, PML
    Atsumi, H
    Leventon, ME
    Hata, N
    Metcalf, DC
    Moriarty, TM
    Alexander, E
    Jolesz, FA
    COMPUTER-ASSISTED NEUROSURGERY, 1997, : 144 - 162
  • [45] Image-guided Neurosurgery/neuronavigation/the SurgiScope - Reflexions on a theme
    Haase, J
    MINIMALLY INVASIVE NEUROSURGERY, 1999, 42 (02) : 53 - 59
  • [46] Image-Guided Neurosurgery: History and Current Clinical Applications
    Thomas, Neil W. D.
    Sinclair, John
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2015, 46 (03) : 331 - 342
  • [47] Simulation of Brain Tumor Resection in Image-guided Neurosurgery
    Fan, Xiaoyao
    Ji, Songbai
    Fontaine, Kathryn
    Hartov, Alex
    Roberts, David
    Paulsen, Keith
    MEDICAL IMAGING 2011: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND MODELING, 2011, 7964
  • [48] Prediction of Fiducial Motion in Respiratory Tumors for Image-Guided Radiotherapy
    Riaz, N.
    Wiersma, R.
    Mao, W.
    Xing, L.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [49] Evaluation of a robust fiducial tracking algorithm for image-guided radiosurgery
    Hatipoglu, Serkan
    Mu, Zhiping
    Fu, Dongshan
    Kuduvalli, Gopinath
    MEDICAL IMAGING 2007: VISUALIZATION AND IMAGE-GUIDED PROCEDURES, PTS 1 AND 2, 2007, 6509
  • [50] Fusion of anatomic and electrophysiologic information for image-guided neurosurgery
    Gore, RK
    Galloway, RL
    Maciunas, RJ
    Dawant, BM
    IMAGE DISPLAY - MEDICAL IMAGING 1997, 1997, 3031 : 167 - 175