Standard MRI-based attenuation correction for PET/MRI phantoms: a novel concept using MRI-visible polymer

被引:7
|
作者
Rausch, Ivo [1 ]
Valladares, Alejandra [1 ]
Sundar, Lalith Kumar Shiyam [1 ]
Beyer, Thomas [1 ]
Hacker, Marcus [2 ]
Meyerspeer, Martin [3 ]
Unger, Ewald [1 ]
机构
[1] Med Univ Vienna, Ctr Med Phys & Biomed Engn, QIMO Team, Waehringer Guertel 18-20-4L, A-1090 Vienna, Austria
[2] Med Univ Vienna, Dept Biomed Imaging & Image Guided Therapy, Div Nucl Med, Vienna, Austria
[3] Med Univ Vienna, High Field MR Ctr, Ctr Med Phys & Biomed Engn, Vienna, Austria
基金
奥地利科学基金会;
关键词
PET; MRI; Phantom attenuation correction; MR visible polymer;
D O I
10.1186/s40658-021-00364-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background PET/MRI phantom studies are challenged by the need of phantom-specific attenuation templates to account for attenuation properties of the phantom material. We present a PET/MRI phantom built from MRI-visible material for which attenuation correction (AC) can be performed using the standard MRI-based AC. Methods A water-fillable phantom was 3D-printed with a commercially available MRI-visible polymer. The phantom had a cylindrical shape and the fillable compartment consisted of a homogeneous region and a region containing solid rods of different diameters. The phantom was filled with a solution of water and [18F]FDG. A 30 min PET/MRI acquisition including the standard Dixon-based MR-AC method was performed. In addition, a CT scan of the phantom was acquired on a PET/CT system. From the Dixon in-phase, opposed-phase and fat images, a phantom-specific AC map (Phantom MR-AC) was produced by separating the phantom material from the water compartment using a thresholding-based method and assigning fixed attenuation coefficients to the individual compartments. The PET data was reconstructed using the Phantom MR-AC, the original Dixon MR-AC, and an MR-AC just containing the water compartment (NoWall-AC) to estimate the error of ignoring the phantom walls. CT-based AC was employed as the reference standard. Average %-differences in measured activity between the CT corrected PET and the PET corrected with the other AC methods were calculated. Results The phantom housing and the liquid compartment were both visible and distinguishable from each other in the Dixon images and allowed the segmentation of a phantom-specific MR-based AC. Compared to the CT-AC PET, average differences in measured activity in the whole water compartment in the phantom of -0.3%, 9.4%, and -24.1% were found for Dixon phantom MR-AC, MR-AC, and NoWall-AC based PET, respectively. Average differences near the phantom wall in the homogeneous region were -0.3%, 6.6%, and -34.3%, respectively. Around the rods, activity differed from the CT-AC PET by 0.7%, 8.9%, and -45.5%, respectively. Conclusion The presented phantom material is visible using standard MR sequences, and thus, supports the use of standard, phantom-independent MR measurements for MR-AC in PET/MRI phantom studies.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] A PET/MRI phantom study using MRI based attenuation correction to account for attenuation of the phantom material
    Rausch, I.
    Valladares, A.
    Beyer, T.
    Unger, E.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2020, 47 (SUPPL 1) : S451 - S452
  • [22] Comparison between MRI-based attenuation correction methods for brain PET in dementia patients
    Cabello, Jorge
    Lukas, Mathias
    Kops, Elena Rota
    Ribeiro, Andre
    Shah, N. Jon
    Yakushev, Igor
    Pyka, Thomas
    Nekolla, Stephan G.
    Ziegler, Sibylle I.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 (12) : 2190 - 2200
  • [23] Comparison between MRI-based attenuation correction methods for brain PET in dementia patients
    Jorge Cabello
    Mathias Lukas
    Elena Rota Kops
    André Ribeiro
    N. Jon Shah
    Igor Yakushev
    Thomas Pyka
    Stephan G. Nekolla
    Sibylle I. Ziegler
    [J]. European Journal of Nuclear Medicine and Molecular Imaging, 2016, 43 : 2190 - 2200
  • [24] Generation of a Four-Class Attenuation Map for MRI-Based Attenuation Correction of PET Data in the Head Area Using a Novel Combination of STE/Dixon-MRI and FCM Clustering
    Parisa Khateri
    Hamidreza Saligheh Rad
    Amir Homayoun Jafari
    Anahita Fathi Kazerooni
    Afshin Akbarzadeh
    Mohsen Shojae Moghadam
    Arvin Aryan
    Pardis Ghafarian
    Mohammad Reza Ay
    [J]. Molecular Imaging and Biology, 2015, 17 : 884 - 892
  • [25] Generation of a Four-Class Attenuation Map for MRI-Based Attenuation Correction of PET Data in the Head Area Using a Novel Combination of STE/Dixon-MRI and FCM Clustering
    Khateri, Parisa
    Rad, Hamidreza Saligheh
    Jafari, Amir Homayoun
    Kazerooni, Anahita Fathi
    Akbarzadeh, Afshin
    Moghadam, Mohsen Shojae
    Aryan, Arvin
    Ghafarian, Pardis
    Ay, Mohammad Reza
    [J]. MOLECULAR IMAGING AND BIOLOGY, 2015, 17 (06) : 884 - 892
  • [26] A Path to Qualification of PET/MRI Scanners for Multicenter Brain Imaging Studies: Evaluation of MRI-Based Attenuation Correction Methods Using a Patient Phantom
    Catana, Ciprian
    Laforest, Richard
    An, Hongyu
    Boada, Fernando
    Cao, Tuoyu
    Faul, David
    Jakoby, Bjoern
    Jansen, Floris P.
    Kemp, Bradley J.
    Kinahan, Paul E.
    Larson, Peder
    Levine, Michael A.
    Maniawski, Piotr
    Mawlawi, Osama
    McConathy, Jonathan E.
    McMillan, Alan B.
    Price, Julie C.
    Rajagopal, Abhejit
    Sunderland, John
    Veit-Haibach, Patrick
    Wangerin, Kristen A.
    Ying, Chunwei
    Hope, Thomas A.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2022, 63 (04) : 615 - 621
  • [27] MRI based attenuation correction for brain PET images
    Kops, Elena Rota
    Qin, Peng
    Mueller-Veggian, Mattea
    Herzog, Hans
    [J]. ADVANCES IN MEDICAL ENGINEERING, 2007, 114 : 93 - +
  • [28] MRI-Based Attenuation Correction for Whole-Body PET/MRI: Quantitative Evaluation of Segmentation- and Atlas-Based Methods
    Hofmann, Matthias
    Bezrukov, Ilja
    Mantlik, Frederic
    Aschoff, Philip
    Steinke, Florian
    Beyer, Thomas
    Pichler, Bernd J.
    Schoelkopf, Bernhard
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2011, 52 (09) : 1392 - 1399
  • [29] Assessment of MRI-Based Attenuation Correction for MRI-Only Radiotherapy Treatment Planning of the Brain
    Ranta, Iiro
    Teuho, Jarmo
    Linden, Jani
    Klen, Riku
    Teras, Mika
    Kapanen, Mika
    Keyrilainen, Jani
    [J]. DIAGNOSTICS, 2020, 10 (05)
  • [30] MRI-visible polymer based on poly(methyl methacrylate) for imaging applications
    Younis, Mira
    Darcos, Vincent
    Paniagua, Cedric
    Ronjat, Pauline
    Lemaire, Laurent
    Nottelet, Benjamin
    Garric, Xavier
    Bakkour, Youssef
    El Nakat, John Hanna
    Coudane, Jean
    [J]. RSC ADVANCES, 2016, 6 (07): : 5754 - 5760