Standard MRI-based attenuation correction for PET/MRI phantoms: a novel concept using MRI-visible polymer

被引:7
|
作者
Rausch, Ivo [1 ]
Valladares, Alejandra [1 ]
Sundar, Lalith Kumar Shiyam [1 ]
Beyer, Thomas [1 ]
Hacker, Marcus [2 ]
Meyerspeer, Martin [3 ]
Unger, Ewald [1 ]
机构
[1] Med Univ Vienna, Ctr Med Phys & Biomed Engn, QIMO Team, Waehringer Guertel 18-20-4L, A-1090 Vienna, Austria
[2] Med Univ Vienna, Dept Biomed Imaging & Image Guided Therapy, Div Nucl Med, Vienna, Austria
[3] Med Univ Vienna, High Field MR Ctr, Ctr Med Phys & Biomed Engn, Vienna, Austria
基金
奥地利科学基金会;
关键词
PET; MRI; Phantom attenuation correction; MR visible polymer;
D O I
10.1186/s40658-021-00364-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background PET/MRI phantom studies are challenged by the need of phantom-specific attenuation templates to account for attenuation properties of the phantom material. We present a PET/MRI phantom built from MRI-visible material for which attenuation correction (AC) can be performed using the standard MRI-based AC. Methods A water-fillable phantom was 3D-printed with a commercially available MRI-visible polymer. The phantom had a cylindrical shape and the fillable compartment consisted of a homogeneous region and a region containing solid rods of different diameters. The phantom was filled with a solution of water and [18F]FDG. A 30 min PET/MRI acquisition including the standard Dixon-based MR-AC method was performed. In addition, a CT scan of the phantom was acquired on a PET/CT system. From the Dixon in-phase, opposed-phase and fat images, a phantom-specific AC map (Phantom MR-AC) was produced by separating the phantom material from the water compartment using a thresholding-based method and assigning fixed attenuation coefficients to the individual compartments. The PET data was reconstructed using the Phantom MR-AC, the original Dixon MR-AC, and an MR-AC just containing the water compartment (NoWall-AC) to estimate the error of ignoring the phantom walls. CT-based AC was employed as the reference standard. Average %-differences in measured activity between the CT corrected PET and the PET corrected with the other AC methods were calculated. Results The phantom housing and the liquid compartment were both visible and distinguishable from each other in the Dixon images and allowed the segmentation of a phantom-specific MR-based AC. Compared to the CT-AC PET, average differences in measured activity in the whole water compartment in the phantom of -0.3%, 9.4%, and -24.1% were found for Dixon phantom MR-AC, MR-AC, and NoWall-AC based PET, respectively. Average differences near the phantom wall in the homogeneous region were -0.3%, 6.6%, and -34.3%, respectively. Around the rods, activity differed from the CT-AC PET by 0.7%, 8.9%, and -45.5%, respectively. Conclusion The presented phantom material is visible using standard MR sequences, and thus, supports the use of standard, phantom-independent MR measurements for MR-AC in PET/MRI phantom studies.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Standard MRI-based attenuation correction for PET/MRI phantoms: a novel concept using MRI-visible polymer
    Ivo Rausch
    Alejandra Valladares
    Lalith Kumar Shiyam Sundar
    Thomas Beyer
    Marcus Hacker
    Martin Meyerspeer
    Ewald Unger
    [J]. EJNMMI Physics, 8
  • [2] Attenuation Correction for PET/MRI Using MRI-based Pseudo CT
    Wang, Tonghe
    Lei, Yang
    Dong, Xue
    Higgins, Kristin
    Liu, Tian
    Curran, Walter J.
    Mao, Hui
    Nye, Jonathon A.
    Yang, Xiaofeng
    [J]. MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312
  • [3] MRI-based attenuation correction and quantification tools for combined MRI/PET
    Fei, Baowei
    Yang, Xiaofeng
    Nye, Jonathon
    Jones, Margie
    Aarsvold, John
    Raghunath, Nivedita
    Meltzer, Carolyn
    Votaw, John
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [4] MRI-Based Attenuation Correction for PET/MRI Using Ultrashort Echo Time Sequences
    Keereman, Vincent
    Fierens, Yves
    Broux, Tom
    De Deene, Yves
    Lonneux, Max
    Vandenberghe, Stefaan
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (05) : 812 - 818
  • [5] Dynamic [18F]FET-PET/MRI using standard MRI-based attenuation correction methods
    Rausch, Ivo
    Zitterl, Andreas
    Berroteran-Infante, Neydher
    Rischka, Lucas
    Prayer, Daniela
    Fenchel, Matthias
    Sareshgi, Reza A.
    Haug, Alexander R.
    Hacker, Marcus
    Beyer, Thomas
    Traub-Weidinger, Tatjana
    [J]. EUROPEAN RADIOLOGY, 2019, 29 (08) : 4276 - 4285
  • [6] Dynamic [18F]FET-PET/MRI using standard MRI-based attenuation correction methods
    Ivo Rausch
    Andreas Zitterl
    Neydher Berroterán-Infante
    Lucas Rischka
    Daniela Prayer
    Matthias Fenchel
    Reza A. Sareshgi
    Alexander R. Haug
    Marcus Hacker
    Thomas Beyer
    Tatjana Traub-Weidinger
    [J]. European Radiology, 2019, 29 : 4276 - 4285
  • [7] MRI-Based Attenuation Correction for PET/MRI Using Multiphase Level-Set Method
    An, Hyun Joon
    Seo, Seongho
    Kang, Hyejin
    Choi, Hongyoon
    Cheon, Gi Jeong
    Kim, Han-Joon
    Lee, Dong Soo
    Song, In Chan
    Kim, Yu Kyeong
    Lee, Jae Sung
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2016, 57 (04) : 587 - 593
  • [8] Three MRI-based attenuation correction methods for PET
    Marshall, Harry
    Laidley, David
    Patrick, John
    Butler, John
    Thompson, Robert
    Prato, Frank
    Rachinsky, Irina
    Stodilka, Robert
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [9] MRI-Based Attenuation Correction for PET/MRI: A Novel Approach Combining Pattern Recognition and Atlas Registration
    Hofmann, Matthias
    Steinke, Florian
    Scheel, Verena
    Charpiat, Guillaume
    Farquhar, Jason
    Aschoff, Philip
    Brady, Michael
    Schoelkopf, Bernhard
    Pichler, Bernd J.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (11) : 1875 - 1883
  • [10] MRI-based Attenuation Correction for PET/MRI: A Novel Approach Combining Anatomic Signature and Machine Learning
    Yang, Xiaofeng
    Lei, Yang
    Higgins, Kristin
    Mao, Hui
    Liu, Tian
    Shim, Hyunsuk
    Curran, Walter
    Nye, Jonathon
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2018, 59