Reduction of femoral shaft fractures in vitro by a new developed reduction robot system 'RepoRobo'

被引:50
|
作者
Füchtmeier, B
Egersdoerfer, S
Mai, R
Hente, R
Dragoi, D
Monkman, G
Nerlich, M
机构
[1] Univ Clin Regensburg, Dept Trauma Surg, D-93042 Regensburg, Germany
[2] Univ Appl Sci, Mechantron Res Unit, D-93042 Regensburg, Germany
[3] Univ Appl Sci, Dept Mech Engn, D-93042 Regensburg, Germany
关键词
fracture reduction; reduction robot; RepoRobo;
D O I
10.1016/j.injury.2004.05.019
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Closed reduction of the tong bones is associated with the use of considerable force. This force must be maintained for the reduction maneuver and fixation process. At present, apart from the extension table or the large AO distractor, only rather inadequate reduction aids are available. A solution to this problem is being sought in the form of a robotic system with which precision can be improved and the holding effort reduced. In the research project presented here, a synthetic femur with integrated tensioned mainspring and a 32-A3 type fracture served as a bone reduction model. The fracture was stabilized with a standard AO fixator. A Staubli robot (model RX130) was converted by appropriate modification so that it could be used for the reduction of femoral shaft fractures in vitro. The robot was equipped with a pneumatic 2-fingered gripper, on which the fingers have been modified so that they can grip the AO fixator clamp. A Force-Feed back-Sensor was inserted between the gripper and the robot to obtain online recordings of the forces and moments in all three axes. With this setup it is possible to achieve precise reduction of the fracture in all planes under visual control.
引用
收藏
页码:113 / 119
页数:7
相关论文
共 50 条
  • [21] Displaced femoral shaft fractures treated by antegrade nailing with the assistance of an intramedullary reduction device
    Wei Chen
    Yongmin Jing
    Hongzhi Lv
    Juan Wang
    Zhiyong Hou
    Yingze Zhang
    International Orthopaedics, 2016, 40 : 1735 - 1739
  • [22] Manipulation with Schanz screws for closed reduction of femoral shaft fractures during intramedullary nailing
    Sadighi, Ali
    Navali, Amir M.
    Aslani, Hossein
    Arzromchilar, Ali
    SAUDI MEDICAL JOURNAL, 2009, 30 (05) : 662 - 666
  • [23] Plating of femoral shaft fractures: Open reduction and internal fixation versus submuscular fixation
    Zlowodzki, Michael
    Vogt, Dennis
    Cole, Peter A.
    Kregor, Philip J.
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2007, 63 (05): : 1061 - 1065
  • [24] The use of percutaneous joystick reduction and limited open reduction techniques in pediatric femoral shaft fractures: a study of 63 cases
    Liu, Fei
    Ju, Li
    Tang, Kai
    Lou, Yue
    JOURNAL OF PEDIATRIC ORTHOPAEDICS-PART B, 2016, 25 (04): : 375 - 380
  • [25] Design and evaluation of an intelligent reduction robot system for the minimally invasive reduction in pelvic fractures
    Zhao, Chunpeng
    Wang, Yu
    Wu, Xinbao
    Zhu, Gang
    Shi, Shuchang
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2022, 17 (01)
  • [26] Design and evaluation of an intelligent reduction robot system for the minimally invasive reduction in pelvic fractures
    Chunpeng Zhao
    Yu Wang
    Xinbao Wu
    Gang Zhu
    Shuchang Shi
    Journal of Orthopaedic Surgery and Research, 17
  • [27] ACCURACY OF REDUCTION OF FEMORAL SUBCAPITAL FRACTURES
    COMPTON, EH
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1977, 9 (01): : 71 - 73
  • [28] Surgical outcomes of minimally invasive cerclage clamping technique using a pointed reduction clamp for reduction of nonisthmal femoral shaft fractures
    Lim, Eic Ju
    Kim, Joon-Woo
    Vemulapalli, Krishna Chandra
    Yoon, Yong-Cheol
    Cho, Jae-Woo
    Oh, Jong-Keon
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2021, 52 (07): : 1897 - 1902
  • [29] Nonanatomical Reduction of Displaced Subcapital Femoral Fractures (Gotfried Reduction)
    Gotfried, Yechiel
    Kovalenko, Sergey
    Fuchs, Dror
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2013, 27 (11) : e254 - e259
  • [30] Accuracy of Reduction of Ipsilateral Femoral Neck and Shaft Fractures - An Analysis of Various Internal Fixation Strategies
    Bedi, Asheesh
    Karunakar, Madhav A.
    Caron, Troy
    Sanders, Roy W.
    Haidukewych, George J.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2009, 23 (04) : 249 - 253