Body composition measurements using bioimpedance analysis in peritoneal dialysis patients are affected by the presence of dialysate

被引:13
|
作者
Kang, Seok Hui [1 ]
Cho, Kyu Hyang [1 ]
Park, Jong Won [1 ]
Yoon, Kyung Woo [1 ]
Do, Jun Young [1 ]
机构
[1] Yeungnam Univ Hosp, Dept Internal Med, Div Nephrol, Taegu 705717, South Korea
关键词
bioimpedance analysis; body composition; dialysate; peritoneal dialysis; BIOELECTRICAL-IMPEDANCE ANALYSIS; MULTIFREQUENCY BIOIMPEDANCE; CARDIOVASCULAR RISK; KIDNEY-DISEASE; VOLUME; ASSESSMENTS; MORTALITY;
D O I
10.1111/nep.12322
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
The presence of peritoneal dialysate when performing bioimpedance analysis may affect body composition measurements. The aim of this study was to evaluate the impact of dialysate on body composition measurements in Asians. Forty-one patients undergoing maintenance peritoneal dialysis in our hospital peritoneal dialysis unit were included in this study. Dialysate was drained from the abdomen prior to measurement, and bioimpedance analysis was performed using multi-frequency bioimpedance analysis, with each subject in a standing position (D-). Dialysate was then administered and the measurement was repeated (D+). The presence of peritoneal dialysate led to an increase in intracellular water (ICW), extracellular water (ECW), and total body water (D-: 20.33 +/- 3.72L for ICW and 13.53 +/- 2.54L for ECW; D+: 20.96 +/- 3.78L for ICW and 14.10 +/- 2.59L for ECW; P<0.001 for both variables). Total and trunk oedema indices were higher in the presence of peritoneal dialysate. In addition, the presence of peritoneal dialysate led to an overestimation of mineral content and free fat mass (FFM) for the total body; but led to an underestimation of body fat (D-: 45.80 +/- 8.26kg for FFM and 19.30 +/- 6.27kg for body fat; D+: 47.51 +/- 8.38kg for FFM and 17.59 +/- 6.47kg for body fat; P<0.001 for both variables). Our results demonstrate that the presence of peritoneal dialysate leads to an overestimation of FFM and an underestimation of fat mass. An empty abdomen is recommended when evaluating body composition using bioimpedance analysis.
引用
收藏
页码:727 / 731
页数:5
相关论文
共 50 条
  • [11] Changes in body composition assessed by bioimpedance analysis in the first 6 months of chronic peritoneal dialysis
    Edefonti, A
    Carcano, A
    Damiani, B
    Ghio, L
    Consalvo, G
    Picca, M
    [J]. ADVANCES IN PERITONEAL DIALYSIS/1997, VOL 13: SELECTED PAPERS FROM THE SEVENTEENTH ANNUAL CONFERENCE ON PERITONEAL DIALYSIS, DENVER, COLORADO, FEBRUARY 1997, 1997, 13 : 267 - 270
  • [12] Thoracic versus whole body bioimpedance measurements:: the relation to hydration status and hypertension in peritoneal dialysis patients
    Nescolarde, L.
    Garcia-Gonzalez, M. A.
    Rosell-Ferrer, J.
    Donate, T.
    Querfeld, U.
    [J]. PHYSIOLOGICAL MEASUREMENT, 2006, 27 (10) : 961 - 971
  • [13] Body composition in patients treated with peritoneal dialysis
    Johansson, AC
    Samuelsson, O
    Haraldsson, B
    Bosaeus, I
    Attman, PO
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 1998, 13 (06) : 1511 - 1517
  • [14] Bioimpedance analysis and assessment of intracellular water in peritoneal dialysis patients
    de Kerkhof, JV
    Konings, CJ
    van Kreel, B
    van der Sande, FM
    Leunissen, KML
    Kooman, JP
    [J]. PERITONEAL DIALYSIS INTERNATIONAL, 2003, 23 (06): : 591 - 593
  • [15] EVALUATION OF NUTRITIONAL STATUS OF PERITONEAL DIALYSIS PATIENTS BY BIOIMPEDANCE ANALYSIS
    Akdam, Hakan
    Alp, Alper
    Ozbek, Ozgul
    Kahraman, Aysegul
    Ozisik, Melih
    Arda, Hayri Ustun
    Yenicerioglu, Yavuz
    [J]. ACTA MEDICA MEDITERRANEA, 2016, 32 (03): : 685 - 690
  • [16] Longitudinal body composition analysis (BCA) in peritoneal dialysis patients.
    Frumkin, D
    Stall, S
    DeVita, MV
    Michelis, MF
    [J]. AMERICAN JOURNAL OF KIDNEY DISEASES, 2000, 35 (04) : A11 - A11
  • [17] Body Composition in Dialysis Patients: A Functional Assessment of Bioimpedance Using Different Prediction Models
    Broers, Natascha J. H.
    Martens, Remy J. H.
    Cornelis, Tom
    Diederen, Nanda M. P.
    Wabel, Peter
    van der Sande, Frank M.
    Leunissen, Karel M. L.
    Kooman, Jeroen P.
    [J]. JOURNAL OF RENAL NUTRITION, 2015, 25 (02) : 121 - 128
  • [18] Assessment of Body Composition in Dialysis Patients by Arm Bioimpedance Compared to MRI and 40K Measurements
    Carter, M.
    Zhu, F.
    Kotanko, P.
    Kuhlmann, M.
    Ramirez, L.
    Heymsfield, S. B.
    Handelman, G.
    Levin, N. W.
    [J]. BLOOD PURIFICATION, 2009, 27 (04) : 330 - 337
  • [19] Estimation of body fluid changes during peritoneal dialysis by segmental bioimpedance analysis
    Zhu, FS
    Schneditz, D
    Kaufman, AM
    Levin, NW
    [J]. KIDNEY INTERNATIONAL, 2000, 57 (01) : 299 - 306
  • [20] Application of Body Composition Monitoring to Peritoneal Dialysis Patients
    Crepaldi, Carlo
    Soni, Sachin
    Chionh, Chang Yin
    Wabel, Peter
    Cruz, Dinna N.
    Ronco, Claudio
    [J]. PERITONEAL DIALYSIS - FROM BASIC CONCEPTS TO CLINICAL EXCELLENCE, 2009, 163 : 1 - 6