Clinical optimization of filters in direct a-se FFDM (full field digital mammography) system

被引:0
|
作者
Uchiyama, Nachiko
Moriyama, Noriyuki
Kitagawa, Mayumi
Gomi, Shiho
Nagai, Yuichi
机构
[1] Natl Canc Ctr, Dept Canc Screening, Res Ctr Canc Screening, Chuo Ku, Tokyo 1040045, Japan
[2] Natl Canc Ctr, Dept Radiol, Res Ctr Canc Screening, Chuo Ku, Tokyo 1040045, Japan
来源
关键词
D O I
暂无
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We evaluated three combinations of filters (Mo/Mo, Mo/ Rh, and W /Rh) in direct a-Se FFDM system to optimize radiation dose clinically. We measured CNR (Contrast to Noise Ratio) as physical characteristics changing radiation dose and phantom thickness in clinical range. In 20, 30, 40, and 50mm PMMA phantoms, Mo / Mo showed the best performances. On the other hand, in 60 and 70 mm, W/Rh 30kV showed best performance. In addition, in 40 and 50mm PMMA phantoms, W/Rh 30kV showed the second best performance. In direct a-Se FFDM system, W/Rh was valuable in minimizing radiation dose.
引用
收藏
页码:315 / 323
页数:9
相关论文
共 50 条
  • [1] Anatomically adaptable automatic exposure control (AEC) for amorphous selenium (a-Se) full field digital mammography (FFDM) system
    Varjonen, Mari
    Strommer, Pekka
    Medical Imaging 2006: Physics of Medical Imaging, Pts 1-3, 2006, 6142 : 14207 - 14207
  • [2] Direct flat-panel detectors: A-Se TFT detector for full field digital mammography
    Gkanatsios, N
    MEDICAL PHYSICS, 2003, 30 (06) : 1434 - 1434
  • [3] Physical characteristics of a clinical prototype for full-field digital mammography with an a-Se flat panel detector
    Lazzari, B
    Belli, G
    Gori, C
    Nykdnen, K
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 656 - 666
  • [4] Implementation of telemanagement for full-field digital mammography (FFDM)
    Fan, Y
    Lou, SL
    Wong, A
    Zhang, H
    Sickles, EA
    MEDICAL IMAGING 2000: PACS DESIGN AND EVALUATION - ENGINEERING AND CLINICAL ISSUES, 2000, 3980 : 400 - 407
  • [5] Value and importance of full-field digital mammography (FFDM) for screening
    Grabbe, E
    Fischer, U
    Funke, M
    Hermann, KP
    Obenauer, S
    Baum, F
    RADIOLOGE, 2001, 41 (04): : 359 - 365
  • [6] Characterization of an FFDM unit based on a-Se direct conversion detector
    Albanese, A
    Bevilacqua, A
    Campanini, R
    Iampieri, E
    Lanconelli, N
    Rossi, M
    Romani, D
    Salomoni, V
    Vignoli, P
    DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2003, : 69 - 71
  • [7] Surgical management of mammographic microcalcification is improved by full field digital mammography (FFDM)
    Bundred, S. M.
    Zhou, J.
    Hurley, E.
    Wilson, M.
    Morris, J.
    Bundred, N. J.
    CANCER RESEARCH, 2012, 72
  • [8] The medical physicist's role in a Full Field Digital Mammography (FFDM) facility
    Chakrabarti, K
    MEDICAL PHYSICS, 2002, 29 (06) : 1355 - 1355
  • [9] Computer-aided detection (CAD) in direct full-field digital mammography (FFDM): Preliminary results
    Baum, FT
    Fischer, U
    Obenauer, S
    Muller, SL
    Benali, K
    Grabbe, EH
    RADIOLOGY, 2001, 221 : 471 - 471
  • [10] Hardcopy requirements and imager characteristics for full field digital mammography (FFDM) applications
    Mohapatra, SK
    Anderson, WF
    Keyes, GS
    Lindquist, TR
    Pearson, VE
    MEDICAL IMAGING 1999: IMAGE DISPLAY, 1999, 3658 : 269 - 279