Accuracy of the dose delivery in prostate cancer patients-using an electronic portal imaging device (EPID)

被引:6
|
作者
Mahdavi, S. R. [1 ]
Gharehbagh, E. Jazayeri [2 ]
Mofid, B. [3 ]
Jafari, A. H. [4 ]
Nikoofar, A. R. [5 ]
机构
[1] Iran Univ Med Sci, Dept Med Phys, Tehran, Iran
[2] Univ Tehran Med Sci, Sch Allied Hlth Sci, Dept Radiol & Radiotherapy Technol, Tehran, Iran
[3] Shahid Beheshti Univ Med Sci, Dept Radiotherapy, Tehran, Iran
[4] Tehran Univ Medial Sci, Res Ctr Biomed Technol & Robot, Tehran, Iran
[5] Iran Univ Med Sci, Dept Radiotherapy, Tehran, Iran
来源
关键词
Prostate cancer; EPID; systematic errors; random errors; RADIOPAQUE MARKERS; ORGAN MOTION; TREATMENT VERIFICATION; STANDARD RADIOTHERAPY; ROUTINE RADIOTHERAPY; GUIDED RADIOTHERAPY; PELVIC RADIOTHERAPY; IRRADIATION; POSITION; LOCALIZATION;
D O I
10.18869/acadpub.ijrr.15.1.39
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: To correct patient positioning errors (setup errors) during prostate cancer treatment using EPID and fiducial gold markers, to improve the accuracy of the dose delivery in these patients. Materials and Methods: Fifteen patients with localized prostate carcinoma after implantation of fiducial gold markers in their prostate gland underwent the five-field IMRT planning technique. The plan was prepared in accordance with ICRU 50 guidance (PTV to receive 95-107% dose). The software program reconstructed the three-dimensional position of the markers from the different Beams Eye Views (BEV). The discrepancies of the seeds' positions (prostate surrogate) between plan and daily images were calculated three dimensionally. Then, necessary corrections were applied to match the prostate fiducial markers in the portal image with the BEV image in the planned one by moving the couch in the X, Y and Z directions. Results: Data from 15 patients and 469 fractions of radiotherapy were analyzed in this study. Two sets of data were available from EPID software before and after 3D set-up corrections. The mean of the population displacement in Left /Right (L/R), Anterior/Posterior (A/P) and Crania/Caudal (C/C) directions were 0.5,-1.0 and 2.4mm before, and-0.1,-0.5 and 0.9mm after corrections, respectively. The systematic and random errors for the measured populations in the three mentioned directions were 2.4, 2.7 and 2mm and 6.4, 5.9 and 6.1mm before corrections, and 1.1, 2.4 and 1.4mm and 3.8, 3.9 and 3.6mm after corrections, correspondingly. Conclusion: This study provides further evidence that using gold markers in the prostate improves dose delivery to the prostate. Also, it has been demonstrated that the EPID can be a powerful tool in the reduction of treatment setup errors and the quality assurance and verification of complex treatments.
引用
收藏
页码:39 / 47
页数:9
相关论文
共 50 条
  • [21] Concept for On Line - Verification of IMRT - Radiation treatment using electronic portal imaging device (EPID).
    Muehle, S.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2010, 186 : 18 - 18
  • [22] Verification of radiosurgery target point alignment with an electronic portal imaging device (EPID)
    Dong, L
    Shiu, A
    Tung, S
    Boyer, A
    [J]. MEDICAL PHYSICS, 1997, 24 (02) : 263 - 267
  • [23] SIMPLE PROPOSAL FOR DOSIMETRY WITH AN ELEKTA IVIEWGTTM ELECTRONIC PORTAL IMAGING DEVICE (EPID)
    Liebich, J.
    Licher, J.
    Scherf, C.
    Kara, E.
    Koch, N.
    Roedel, C.
    Ramm, U.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S419 - S419
  • [24] Electronic portal imaging device (EPID) based on a novel camera with avalanche multiplication
    Pang, G
    Zhao, W
    Luhta, R
    Rowlands, JA
    [J]. PHYSICS OF MEDICAL IMAGING, 1998, 3336 : 195 - 201
  • [25] A Novel Method for Fast Image Simulation of Electronic Portal Imaging Device (EPID)
    Shi, M.
    Myronakis, M.
    Hu, Y.
    Berbeco, R.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E478 - E478
  • [26] Prospective study on set-up random errors in 3D-conformal prostate cancer radiotherapy by using an electronic portal imaging device (EPID)
    Zerini, D
    Antognoni, P
    Gandini, S
    Monti, L
    Visconti, M
    Orecchia, R
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S288 - S288
  • [27] Geometrical Verification of Real-Time Tumor Tracking Using Electronic Portal Imaging Device (EPID)
    Han-Oh, Y.
    Yi, B.
    Berman, B.
    Yu, C.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [28] Assessing the accuracy of electronic portal imaging device (EPID)-based dosimetry: II. Evaluation of a dosimetric uncertainty budget and development of a new film-in-EPID absorbed dose calibration methodology
    Renaud, James
    Muir, Bryan
    [J]. MEDICAL PHYSICS, 2022, 49 (02) : 1238 - 1247
  • [29] Movement of an Elekta Electronic Portal Imaging Device (EPID) during Gantry Rotation: Quantification and Correction for the Verification of Dose Distributions
    Koehn, J.
    Licher, J.
    Mielke, M.
    Loutfi-Krauss, B.
    Bluemer, N.
    Roedel, C.
    Scherf, C.
    Ramm, U.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2017, 193 : S36 - S36
  • [30] Verification of segmented beam delivery using a commercial electronic portal imaging device
    Curtin-Savard, AJ
    Podgorsak, EB
    [J]. MEDICAL PHYSICS, 1999, 26 (05) : 737 - 742