Comparison of Central Corneal Thickness Measurements between Noncontact Specular Microscopy and Ultrasound Pachymetry

被引:2
|
作者
Ohn, Kyoung [1 ]
Lee, Mee Yon [1 ]
Lee, Young Chun [1 ]
Shin, Hye Young [1 ]
机构
[1] Catholic Univ Korea, Uijeongbu St Marys Hosp, Coll Med, Dept Ophthalmol, Uijongbu, South Korea
来源
关键词
Central corneal thickness; Endothelial cell density; Noncontact specular microscopy; Ultrasound pachymetry; INTRAOCULAR-PRESSURE; ENDOTHELIAL DAMAGE; TOPCON SP-3000P; SCANNING-SLIT; AGE; POPULATION; TONOMETRY; GLAUCOMA; DENSITY; DEVICES;
D O I
10.3341/jkos.2019.60.7.635
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: We compared and analyzed central corneal thickness (CCT) measurements according to the corneal thickness obtained with noncontact specular microscopy (NCSM) and ultrasound pachymetry (USP). Methods: CCT was measured in the order of NCSM and USP by a single optometrist in 120 eyes of 120 healthy subjects. The measurements were compared between the devices and the measurement agreements and correlations between the devices were analyzed. To determine if the measurements differed depending on the thickness of the cornea, the patients were divided into three tertile groups from the thinnest patient by CCT measurement using USP, and then the differences in CCT measured by the two devices were analyzed. Results: The CCT measurements using NCSM and USP were 548.25 +/- 4.64 mu m and 533.09 +/- 35.96 mu m, respectively. NCSM measurements were found to be thicker, showing statistically significant differences between the measurements (p < 0.001). The two examinations showed a high degree of correlation (r = 0.878; p < 0.01). In the three groups, the differences in CCT measurements between NCSM and USP were 12.93 +/- 21.88 mu m, 16.85 +/- 15.89 mu m, and 15.70 +/- 20.46 mu m, respectively, but the differences between the three groups were not statistically significant (p = 0.655). Conclusions: Our results show that although the differences in CCT measurements using NCSM and USP were consistent regardless of the corneal thickness, the CCT measurements by NCSM and USP were highly correlated.
引用
收藏
页码:635 / 642
页数:8
相关论文
共 50 条
  • [1] Comparison of central corneal thickness measurements by specular microscopy, ultrasound pachymetry, and ultrasound biomicroscopy
    Tam, ES
    Rootman, DS
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (06): : 1179 - 1184
  • [2] Comparison of Pentacam Scheimpflug camera with ultrasound pachymetry and 0 noncontact specular Microscopy in measuring central corneal thickness
    Fujioka, Miyuki
    Nakamura, Makoto
    Tatsumi, Yasuko
    Kusuhara, Azusa
    Maeda, Hidetaka
    Negi, Akira
    [J]. CURRENT EYE RESEARCH, 2007, 32 (02) : 89 - 94
  • [3] Comparison of Central Corneal Thickness with Ultrasound Pachymetry, Noncontact Specular Microscopy and Spectral Domain Optical Coherence Tomography
    Erdur, Sevil Karaman
    Demirci, Goktug
    Dikkaya, Funda
    Kocabora, Mehmet Selim
    Ozsutcu, Mustafa
    [J]. SEMINARS IN OPHTHALMOLOGY, 2018, 33 (06) : 782 - 787
  • [4] Comparison of Central Corneal Thickness Measurements Using Ultrasonic Pachymetry, Anterior Segment OCT and Noncontact Specular Microscopy
    Scotto, Riccardo
    Bagnis, Alessandro
    Papadia, Marina
    Cutolo, Carlo Alberto
    Risso, Domenico
    Traverso, Carlo Enrico
    [J]. JOURNAL OF GLAUCOMA, 2017, 26 (10) : 860 - 865
  • [5] Comparison of central corneal thickness measurements by ultrasound pachymetry, Konan noncontact optical pachymetry, and Orbscan pachymetry
    Christensen, Amarilda
    Narvaez, Julio
    Zimmerman, Grenith
    [J]. CORNEA, 2008, 27 (08) : 862 - 865
  • [6] Discrepancy between central and midperipheral corneal thickness measurements obtained with slit-scanning pachymetry and noncontact specular microscopy
    Jonuscheit, Sven
    Doughty, Michael J.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (12): : 2127 - 2135
  • [7] Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes
    Al-Ageel, Saleh
    Al-Muammar, Abdulrahman M.
    [J]. SAUDI JOURNAL OF OPHTHALMOLOGY, 2009, 23 (3-4) : 181 - 187
  • [8] Comparison of central corneal thickness measured by noncontact tonometry and ultrasound pachymetry
    Park, Hyeyeon
    Shin, Hye Young
    [J]. ACTA OPHTHALMOLOGICA, 2021, 99
  • [9] Corneal thickness measurements with contact and noncontact specular microscopic and ultrasonic pachymetry
    Modis, L
    Langenbucher, A
    Seitz, B
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2001, 132 (04) : 517 - 521
  • [10] Corneal thickness measurements: Scanning-slit corneal topography and noncontact specular microscopy versus ultrasonic pachymetry
    Suzuki, S
    Oshika, T
    Oki, K
    Sakabe, I
    Iwase, A
    Amano, S
    Araie, M
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (07): : 1313 - 1318