Development of a new dynamometer for measuring grip strength applied on a cylindrical handle

被引:45
|
作者
Wimer, Bryan [1 ]
Dong, Ren G. [1 ]
Welcome, Daniel E. [1 ]
Warren, Christopher [1 ]
McDowell, Thomas W. [1 ]
机构
[1] NIOSH, Engn & Control Technol Branch, Morgantown, WV 26505 USA
关键词
Grip force; Grip strength; Hand force; Instrumented handle; Dynamometer; HANDGRIP FORCE; INTERFACE; DIAMETERS;
D O I
10.1016/j.medengphy.2009.01.009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective of this study is to enhance the understanding of the hand grip force applied to a cylindrical handle and to develop a new dynamometer for measuring maximum grip force or grip strength. Specifically, a 40 mm instrumented cylindrical handle with six measuring arms was developed. A theoretical model was proposed and used to analyze the principle of the measurement. Human test subjects were used in conclucting two sets of experiments to evaluate the handle and to assess the measurement method. This study confirmed that some friction force exists in the grip-only action, but its level is not comparable with the normal force. This study also found that the friction force can stabilize the grip action and marginally increase the grip strength. No reliable correlation between the grip strengths measured with the 40 mm cylindrical handle and Jamar handle with a 47.6 mm span was observed. This suggests that grip strength measured with Jamar handle may not be reliably applicable to the design and risk assessment of some tools or machines with cylindrical handles. In contrast, the cylindrical handle proved to be able to determine the overall grip strength for a subject, as well as show the grip force distribution around the circumference of the handle. The handle is accurate with less than 4% error, and it demonstrates that the measurement is independent of the loading position along the handle. Therefore, this study concluded that this new dynamometer is suitable for measuring grip strength with sufficient precision. Published by Elsevier Ltd on behalf of IPEM
引用
收藏
页码:695 / 704
页数:10
相关论文
共 50 条
  • [41] DYNAMOMETER MEASUREMENTS OF GRIP AND KNEE EXTENSION STRENGTH: ARE THEY INDICATIVE OF OVERALL LIMB AND TRUNK MUSCLE STRENGTH?
    Bohannon, Richard W.
    PERCEPTUAL AND MOTOR SKILLS, 2009, 108 (02) : 339 - 342
  • [42] Validity of the Handheld Dynamometer Compared with an Isokinetic Dynamometer in Measuring Peak Hip Extension Strength
    Keep, Heather
    Luu, Levana
    Berson, Ayli
    Garland, S. Jayne
    PHYSIOTHERAPY CANADA, 2016, 68 (01) : 15 - 22
  • [43] Reliability of a new hand-held dynamometer in measuring shoulder range of motion and strength
    Cadogan, Angela
    Laslett, Mark
    Hing, Wayne
    McNair, Peter
    Williams, Maynard
    MANUAL THERAPY, 2011, 16 (01) : 97 - 101
  • [44] Do Grip Strength Dynamometer Readings Improve After Cervical Spine Surgery?
    Zuckerman, Scott L.
    Goldberg, Jacob L.
    Cerpa, Meghan
    Vulapalli, Meghana
    Delgardo, Mychael W.
    Flowers, Xena E.
    Leskinen, Sandra
    Kerolus, Mena G.
    Buchanan, Ian A.
    Ha, Alex S.
    Riew, K. Daniel
    GLOBAL SPINE JOURNAL, 2025, 15 (01) : 76 - 83
  • [45] Longitudinal assessment of grip strength using bulb dynamometer in Duchenne Muscular Dystrophy
    Pizzato, Tatiana M.
    Baptista, Cyntia R. J. A.
    Souza, Mariana A.
    Benedicto, Michelle M. B.
    Martinez, Edson Z.
    Mattiello-Sverzut, Ana C.
    BRAZILIAN JOURNAL OF PHYSICAL THERAPY, 2014, 18 (03) : 245 - 251
  • [46] Do Grip Strength Dynamometer Readings Improve After Cervical Spine Surgery?
    Zuckerman, Scott L.
    Goldberg, Jacob L.
    Cerpa, Meghan
    Vulapalli, Meghana
    Delgardo, Mychael W.
    Flowers, Xena E.
    Leskinen, Sandra
    Kerolus, Mena G.
    Buchanan, Ian A.
    Ha, Alex S.
    Riew, K. Daniel
    GLOBAL SPINE JOURNAL, 2023,
  • [47] A Web Service and Embedded System Based hand dynamometer for Grip strength Assessment
    Luo, Kan
    Huang, Bingfa
    Ma, Ying
    Chen, Wei
    Lai, Zhichen
    Li, Jianxing
    2019 CHINESE AUTOMATION CONGRESS (CAC2019), 2019, : 4062 - 4066
  • [48] Parallel comparison of grip strength measures obtained with a MicroFET 4 and a jamar dynamometer
    Bohannon, RW
    PERCEPTUAL AND MOTOR SKILLS, 2005, 100 (03) : 795 - 798
  • [49] Measuring hand grip strength in rheumatoid arthritis
    Susie C. Higgins
    Jo Adams
    Rodney Hughes
    Rheumatology International, 2018, 38 : 707 - 714
  • [50] Measuring hand grip strength in rheumatoid arthritis
    Higgins, Susie C.
    Adams, Jo
    Hughes, Rodney
    RHEUMATOLOGY INTERNATIONAL, 2018, 38 (05) : 707 - 714