Accuracy of a cryocooler-based programmable Josephson voltage standard

被引:10
|
作者
Behr, R [1 ]
Schubert, M
May, T
机构
[1] PTB, D-38116 Braunschweig, Germany
[2] IPHT, D-07702 Jena, Germany
关键词
compact microwave source; comparison; cryocooler; superconductor-insulator-normal metal-insulator-superconductor (SINIS) Josephson arrays; voltage standard;
D O I
10.1109/TIM.2004.827068
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A cryocooler-based programmable Josephson array voltage standard (JAVS) for 4-K operation and a very compact microwave source are described. Both components are crucial for future mobile voltage standards. A direct comparison between two programmable JAVS using these components was performed at 1V in order to test whether these systems are suitable for metrological application at the highest level of precision. The results of the comparison show that the new components are suitable for mobile JAVS with an uncertainty of less than 1 x 10(-9).
引用
收藏
页码:822 / 825
页数:4
相关论文
共 50 条
  • [41] A Josephson Impedance Bridge Based on Programmable Josephson Voltage Standards
    Hagen, Thomas
    Palafox, Luis
    Behr, Ralf
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2017, 66 (06) : 1539 - 1545
  • [42] Effect of Input Bandwidth of a Sampler on Differential Sampling Based on a Programmable Josephson Voltage Standard
    Kim, Mun-Seog
    Cho, Hehree
    Chayramy, R.
    Solve, Stephane
    2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [43] Direct Comparison between the NIST 10 V Conventional Josephson Voltage Standard and 2.5 V Programmable Josephson Voltage Standard
    Jaiswal, Shiv Kumar
    MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2011, 26 (04): : 339 - 348
  • [44] Direct comparison between the NIST 10 V Compact Josephson Voltage Standard and the 2.5 V Programmable Josephson Voltage Standard
    Jaiswal, Shiv Kumar
    Tang, Y.
    MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2009, 24 (04): : 207 - 214
  • [45] Direct comparison between the NIST 10 V Compact Josephson Voltage Standard and the 2.5 V Programmable Josephson Voltage Standard
    Shiv Kumar Jaiswal
    Y. Tang
    MAPAN, 2009, 24 : 207 - 214
  • [46] Direct comparison between the NIST 10 V Conventional Josephson Voltage Standard and 2.5 V Programmable Josephson Voltage Standard
    Shiv Kumar Jaiswal
    MAPAN, 2011, 26 : 339 - 348
  • [47] Automated 10 Volt Josephson voltage standard system operating with liquid helium and with a cryocooler
    Schubert, M.
    Starkloff, M.
    Meyer, M.
    Wende, G.
    Anders, S.
    Steinbach, B.
    May, T.
    Meyer, H. -G.
    2008 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2008, : 370 - +
  • [48] Calibration of Gigahertz Microwave Generator for the Programmable Josephson Voltage Standard at NRC
    Gertsvolf, M.
    Granger, G.
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2019, 68 (06) : 1901 - 1906
  • [49] DESIGN OF THE NIST 10 V PROGRAMMABLE JOSEPHSON VOLTAGE STANDARD SYSTEM
    Burroughs, C. J.
    Dresselhaus, P. D.
    Ruefenacht, A.
    Elsbury, M. M.
    Benz, P.
    2010 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS CPEM, 2010, : 24 - +
  • [50] Maximizing the Resolution of a Dual RF Drive Programmable Josephson Voltage Standard
    Georgakopoulos, Dimitrios
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2015, 64 (06) : 1503 - 1508