Feasibility study of low tube voltage (80 kVp) coronary CT angiography combined with contrast medium reduction using iterative model reconstruction (IMR) on standard BMI patients

被引:28
|
作者
Zhang, Fan [1 ]
Yang, Li [2 ]
Song, Xiang [2 ]
Li, Ying-Na [2 ]
Jiang, Yan [3 ]
Zhang, Xing-Hua [2 ]
Ju, Hai-Yue [2 ]
Wu, Jian [2 ]
Chang, Rui-Ping [2 ]
机构
[1] Chinese Peoples Liberat Army Gen Hosp, Dept Radiol, Hainan Branch, Beijing, Peoples R China
[2] Chinese Peoples Liberat Army Gen Hosp, Dept Radiol, Beijing, Peoples R China
[3] Philips Healthcare, Clin Sci Imaging Syst, Shanghai, Peoples R China
来源
BRITISH JOURNAL OF RADIOLOGY | 2016年 / 89卷 / 1058期
基金
中国博士后科学基金;
关键词
CARDIOVASCULAR COMPUTED-TOMOGRAPHY; RADIATION-DOSE REDUCTION; IMAGE QUALITY; CARDIAC-CT; HYBRID; ALGORITHM; COMBINATION; PERFORMANCE; PROTOCOLS; SOCIETY;
D O I
10.1259/bjr.20150766
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To investigate the feasibility of low-tube-voltage (80kVp) coronary CT angiography (CCTA) combined with contrast medium (CM) reduction and iterative model reconstruction (IMR) on patients with standard body mass index compared with clinical routine protocol. Methods: Retrospectively gated helical CCTA scans were acquired using a 256-slice multi-slice CT (Brilliance iCT; Philips Healthcare, Cleveland, OH) on 94 patients with standard body mass index (20-25 kgm(-2)) who were randomly assigned into 2 groups. The scan protocol for Group 1 was 100kVp and 600mAs with 70 ml CM at an injection rate of 4.5-5.5 mls(-1); images were reconstructed by a hybrid iterative reconstruction technique (iDose(4); Philips Healthcare). Group 2 was scanned at 80kVp and 600mAs with 35 ml CM at an injection rate of 3.5-4.5 mls(-1); images were reconstructed with IMR. Objective measurements such as the mean image noise and contrast-to-noise ratio of the two groups were measured on CT images and compared using the paired t-test. In addition, a subjective image quality evaluation was performed by two radiologists who were blinded to the scan protocol, using a 5-point scale [1 (poor) to 5 (excellent)]. The results of the two groups were compared using Mann-Whitney U test. Results: The iodine delivery rate of Group 2 was 1.0 +/- 0.5 gls(-1) compared with 2.1 +/- 0.5 gls(-1) in Group 1 resulting in a reduction of 52.4%. In addition, an effective radiation dose reduction of 56.4% was achieved in Group 2 (2.4 +/- 1.2mSv) compared with Group 1 (5.5 +/- 1.4mSv). The mean CT attenuation, contrast-to-noise ratio and image quality of all segments in Group 2 were significantly improved compared with those in Group 1 (all, p<0.01). Conclusion: The use of IMR along with a low tube voltage (80kVp) combined with a low CM protocol for CCTA can reduce both radiation and CM dose with improved image quality. Advances in knowledge: In this study, we used a novel knowledge-based IMR which remarkably reduced the image noise. We compared the quality of the images obtained when the tube voltage was reduced to 80 kVp and that of those obtained according to the clinical routine protocols to determine whether ultra-low-dose imaging plus IMR is feasible in CCTA scans. We found that a low dose protocol combined with 80 kVp and reduced CM for CCTA can reduce both radiation dose and CM dosewith improved image quality by the use of IMR in non-obese patients.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Image quality evaluation of iterative model reconstruction on low tube voltage (80 kVp) coronary CT angiography in an animal study
    Hu, Mao-Qing
    Li, Meng
    Liu, Zai-Yi
    Huang, Mei-Ping
    Liu, Hui
    Liang, Chang-Hong
    ACTA RADIOLOGICA, 2016, 57 (02) : 170 - 177
  • [2] Application of 80-kVp tube voltage, low-concentration contrast agent and iterative reconstruction in coronary CT angiography: evaluation of image quality and radiation dose
    Wu, Qiyong
    Wang, Yong
    Kai, Huihua
    Wang, Tao
    Tang, Xiaoqiang
    Wang, Xiaoqin
    Pan, Changjie
    INTERNATIONAL JOURNAL OF CLINICAL PRACTICE, 2016, 70 (9B) : B50 - B55
  • [3] Lower extremity CT angiography at 80 kVp using iterative model reconstruction
    Liu, B.
    Gao, S.
    Chang, Z.
    Wang, C.
    Liu, Z.
    Zheng, J.
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2018, 99 (09) : 561 - 568
  • [4] Low-tube-voltage (80 kVp) CT aortography using 320-row volume CT with adaptive iterative reconstruction: lower contrast medium and radiation dose
    Chen, Chien-Ming
    Chu, Sung-Yu
    Hsu, Ming-Yi
    Liao, Ying-Lan
    Tsai, Hui-Yu
    EUROPEAN RADIOLOGY, 2014, 24 (02) : 460 - 468
  • [5] Low-tube-voltage (80 kVp) CT aortography using 320-row volume CT with adaptive iterative reconstruction: lower contrast medium and radiation dose
    Chien-Ming Chen
    Sung-Yu Chu
    Ming-Yi Hsu
    Ying-Lan Liao
    Hui-Yu Tsai
    European Radiology, 2014, 24 : 460 - 468
  • [6] Impact of 80 kVp with iterative reconstruction algorithm and low-dose contrast medium on the image quality of craniocervical CT angiography
    Chen, Po-An
    Chen, Chih-Wei
    Chou, Chiung-Chen
    Fu, Jui-Hsun
    Wang, Po-Chin
    Hsu, Shuo-Hsiu
    Lai, Ping-Hong
    CLINICAL IMAGING, 2020, 68 (68) : 124 - 130
  • [7] Feasibility study of iterative model reconstruction combined with low tube voltage, low iodine load, and low iodine delivery rate in craniocervical CT angiography
    Cai, W.
    Hu, C.
    Hu, S.
    Wang, X.
    Gong, J.
    Zhang, W.
    Shi, D.
    Cheng, B.
    CLINICAL RADIOLOGY, 2018, 73 (02) : 217.e1 - 217.e6
  • [8] Evaluation of an iterative model-based reconstruction algorithm for low-tube-voltage (80 kVp) computed tomography angiography
    Noel, Peter B.
    Koehler, Thomas
    Fingerle, Alexander A.
    Brown, Kevin M.
    Zabic, Stanislav
    Muenzel, Daniela
    Haller, Bernhard
    Baum, Thomas
    Henninger, Martin
    Meier, Reinhard
    Rummeny, Ernst J.
    Dobritz, Martin
    JOURNAL OF MEDICAL IMAGING, 2014, 1 (03)
  • [9] Low concentration contrast medium for dual-source computed tomography coronary angiography by a combination of iterative reconstruction and low-tube-voltage technique: Feasibility study
    Zheng, Minwen
    Liu, Ying
    Wei, Mengqi
    Wu, Yongjie
    Zhao, Hongliang
    Li, Jian
    EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (02) : E92 - E99
  • [10] Imaging Quality Evaluation of Low Tube Voltage Coronary CT Angiography Using Low Concentration Contrast Medium
    Zhang, Chengzhong
    Yu, Yuejun
    Zhang, Zaixian
    Wang, Qingguo
    Zheng, Linfeng
    Feng, Yan
    Zhou, Zhiguo
    Zhang, Guixiang
    Li, Kangan
    PLOS ONE, 2015, 10 (03):