Failure of the uncoated titanium ProxiLock femoral hip prosthesis

被引:16
|
作者
Luites, J. W. H.
Spruit, M.
van Hellemondt, G. G.
Horstmann, W. G.
Valstar, E. R.
机构
[1] St Maartensklin Res Dev & Educ, OrthoRes Unit, NL-9500 GM Nijmegen, Netherlands
[2] St Maartensklin Res Dev & Educ, Dept Orthopaed, NL-9500 GM Nijmegen, Netherlands
[3] Leiden Univ, Med Ctr, Dept Orthopaed, Leiden, Netherlands
[4] Delft Univ Technol, Fac Mech Maritime & Mat Engn, Dept Biomech, Delft, Netherlands
关键词
D O I
10.1097/01.blo.0000224011.12175.83
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
New prostheses should be evaluated for stability and clinical performance. In a prospective randomized clinical trial, we implanted 22 titanium (Ti) and 20 hydroxyapatite-coated (HA) ProxiLock femoral hip prostheses during total hip arthroplasty in 42 patients. The patients were followed for 24 months with clinical, radiographic and radiostereometric analysis. Full weightbearing was allowed immediately postoperatively. One patient with a titanium stem was lost to followup. During the first two months, 34 of the 41 stems subsided and/or rotated towards retroversion, regardless of stem type. At the 24-month followup 35 of the 41 prostheses were either fully stabilized (16 HA and 11 Ti stems) or had clinical irrelevant migration (four HA and four Ti stems). Six Ti prostheses showed continuous migrations with maximums of 4.7 mm translation and 12.2 degrees retroversion; four of these were revised, the other two had no clinical complaints. Clinical relevance: The migration pattern we found indicates insufficient primary fixation of the ProxiLock stem in an immediate full weightbearing protocol. The HA coating improves the secondary stability of the prosthesis compared to the uncoated stem. Early migration is associated with an increased risk of possible future loosening and revision, and therefore we discontinued the use of this prosthesis.
引用
收藏
页码:79 / 86
页数:8
相关论文
共 50 条
  • [31] RADIOLOGIC ASPECTS OF NEW UNIVERSAL PROXIMAL FEMORAL HIP PROSTHESIS
    DRINKER, H
    MALL, JC
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1977, 129 (03) : 531 - 533
  • [32] DISTRACTION ARTHROPLASTY OF THE HIP BY BICENTRIC FEMORAL-HEAD PROSTHESIS
    ITO, T
    NAKAYAMA, Y
    TANAKA, H
    ISHIDA, K
    MASUDA, K
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1990, (255) : 186 - 193
  • [33] ANALYTICAL STUDY OF STRESSES IN A FEMORAL STEM OF A TOTAL HIP PROSTHESIS
    ANDRIACCHI, TP
    GALANTE, JO
    ROSTOKER, W
    HAMPTON, S
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1975, A 57 (04): : 565 - 565
  • [34] FATIGUE FRACTURES OF FEMORAL PROSTHESIS IN TOTAL HIP-REPLACEMENT
    CHARNLEY, J
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1976, 58 (03): : 390 - 390
  • [35] Analysis of a femoral hip prosthesis designed to reduce stress shielding
    Joshi, MG
    Advani, SG
    Miller, F
    Santare, MH
    JOURNAL OF BIOMECHANICS, 2000, 33 (12) : 1655 - 1662
  • [36] Total hip arthroplasty with the cementless spiron femoral neck prosthesis
    Lugeder, A.
    Haering, E.
    Mueller, A.
    Droste, P.
    Zeichen, J.
    OPERATIVE ORTHOPADIE UND TRAUMATOLOGIE, 2013, 25 (04): : 388 - 397
  • [37] MECHANISM OF FRACTURE OF THE FEMORAL PROSTHESIS IN TOTAL HIP-REPLACEMENT
    WROBLEWSKI, BM
    INTERNATIONAL ORTHOPAEDICS, 1979, 3 (02) : 137 - 139
  • [38] FRACTURE OF THE FEMORAL PROSTHESIS AFTER TOTAL HIP-REPLACEMENT
    MCNEUR, JC
    MEDICAL JOURNAL OF AUSTRALIA, 1984, 141 (06) : 341 - 344
  • [39] Osteolysis of the femoral shaft after implantation of a cemented hip prosthesis
    Katzer, A.
    Behnke, B.
    von, Foerster, G.
    Journal of Biomechanics, 2001, 34 (SUPPL. 1)
  • [40] FRACTURE OF THE FEMORAL STEM OF THE RING TCH HIP-PROSTHESIS
    WILSON, LF
    NOLAN, JF
    HEYWOODWADDINGTON, MB
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (05): : 725 - 728