Dislodgement of the locking cap from the AO unreamed intramedullary nail: A case report
被引:2
|作者:
Clough, TM
[1
]
Sampath, J
[1
]
Shah, V
[1
]
机构:
[1] Royal Preston Hosp, Dept Orthopaed Surg, Preston, Lancs, England
关键词:
intramedullary nail;
locking cap;
D O I:
10.1097/00005131-200002000-00006
中图分类号:
R826.8 [整形外科学];
R782.2 [口腔颌面部整形外科学];
R726.2 [小儿整形外科学];
R62 [整形外科学(修复外科学)];
学科分类号:
摘要:
We report on the dislodgement of the proximal locking cap from an AO solid femoral intramedullary nail with spiral blade. The pathologic femoral fracture subsequently went on to heal, and the spiral blade did not back out of the femoral neck. Meticulous intraoperative attention to detail is required to select the correct offset for the proximal cap, because the turning moments of the hip abductor muscles are sufficient to unscrew a prominent cap.