Optimization of Spatial Resolution and Image Reconstruction Parameters for the Small-Animal Metis™ PET/CT System

被引:0
|
作者
Zhao, Jie [1 ]
Liu, Qiong [1 ]
Li, Chaofan [2 ]
Song, Yunfeng [1 ]
Zhang, Ying [1 ]
Chen, Jyh-Cheng [1 ,3 ,4 ]
机构
[1] Xuzhou Med Univ, Sch Med Imaging, Xuzhou 221004, Jiangsu, Peoples R China
[2] Xuzhou Med Univ, Sch Med Informat & Engn, Xuzhou 221004, Jiangsu, Peoples R China
[3] Natl Yang Ming Chiao Tung Univ, Dept Biomed Imaging & Radiol Sci, Taipei 11221, Taiwan
[4] Yuanpei Univ Med Technol, Dept Med Imaging & Radiol Technol, Hsinchu 30015, Taiwan
基金
中国博士后科学基金;
关键词
PET imaging; reconstruction parameters; spatial resolution; Derenzo phantom; image quality; PERFORMANCE; SCANNER; FBP; EM;
D O I
10.3390/electronics11101542
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Purpose: The aim of this study was to investigate the optimization of the spatial resolution and image reconstruction parameters related to image quality in an iterative reconstruction algorithm for the small-animal Metis (TM) PET/CT system. Methods: We used a homemade Derenzo phantom to evaluate the image quality using visual assessment, the signal-to-noise ratio, the contrast, the coefficient of variation, and the contrast-to-noise ratio of the 0.8 mm hot rods of eight slices in the center of the phantom PET images. A healthy mouse study was performed to analyze the influence of the optimal reconstruction parameters and the Gaussian post-filter FWHM. Results: In the phantom study, the image quality was the best when the phantom was placed at the end, keeping the central axis parallel to the X-axis of the system, and selecting between 30 and 40 iterations, a 0.314 mm reconstructed voxel size, and a 1.57 mm Gaussian post-filter FWHM. The optimization of the spatial resolution could reach 0.6 mm. In the animal study, it was suitable to choose a voxel size of 0.472 mm, between 30 and 40 iterations, and a 2.36 mm Gaussian post-filter FWHM. Conclusions: Our results indicate that the optimal imaging conditions and reconstruction parameters are very necessary to obtain high-resolution images and quantitative accuracy, especially for the high-precision recognition of tiny lesions.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Image quality phantom and parameters for high spatial resolution small-animal SPECT
    Visser, Eric P.
    Harteveld, Anita A.
    Meeuwis, Antoi P. W.
    Disselhorst, Jonathan A.
    Beekman, Freek J.
    Oyen, Wim J. G.
    Boerman, Otto C.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 654 (01): : 539 - 545
  • [2] Image quality phantom and parameters for high spatial resolution small-animal SPECT
    Visser, E. P.
    Harteveld, A. A.
    Meeuwis, A. P. W.
    Disselhorst, J. A.
    Oyen, W. J. G.
    Boerman, O. C.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 : S173 - S173
  • [3] Reconstruction algorithm with resolution deconvolution in a small-animal PET imager
    Tsyganov, EN
    Zinchenko, AI
    Slavine, NV
    Antich, PP
    Seliounine, SY
    Oz, OK
    Kulkami, PV
    Lewis, MA
    Mason, RP
    Parkey, RW
    [J]. SMALL ANIMAL SPECT IMAGING, 2005, : 163 - 175
  • [4] Spatial Resolution and Sensitivity of the Inveon Small-Animal PET Scanner
    Visser, Eric P.
    Disselhorst, Jonathan A.
    Brom, Maarten
    Laverman, Peter
    Gotthardt, Martin
    Oyen, Wim J. G.
    Boerman, Otto C.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (01) : 139 - 147
  • [5] Development of a small-animal PET system
    Ziegler, SI
    [J]. PROCEEDINGS OF THE IEEE, 2005, 93 (04) : 763 - 770
  • [6] Spatial resolution of the Inveon small-animal PET scanner for the entire field of view
    Disselhorst, J. A.
    Boerman, O. C.
    Oyen, W. J. G.
    Slump, C. H.
    Visser, E. P.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 615 (02): : 245 - 248
  • [7] Assessment of iterative image reconstruction techniques for small-animal PET imaging applications
    Karali, E.
    Pavlopoulos, S.
    Koutsouris, D.
    [J]. 8TH IEEE INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOENGINEERING, VOLS 1 AND 2, 2008, : 948 - 953
  • [8] A Prototype Ultra-High-Resolution Small-Animal PET System
    Liu, Jiguo
    Liu, Qiong
    Chen, Jyh-Cheng
    Wang, Deguang
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2021, 62
  • [9] Prototype Small-Animal PET-CT Imaging System for Image-Guided Radiation Therapy
    Mikhaylova, Ekaterina
    Brooks, Jamison
    Zuro, Darren M.
    Nouizi, Farouk
    Kujawski, Maciej
    Madabushi, Srideshikan Sargur
    Qi, Jinyi
    Zhang, Mengxi
    Chea, Junie
    Poku, Erasmus K.
    Bowles, Nicole
    Wong, Jeffrey Y. C.
    Shively, John E.
    Yazaki, Paul J.
    Gulsen, Gultekin
    Cherry, Simon R.
    Hui, Susanta K.
    [J]. IEEE ACCESS, 2019, 7 : 143207 - 143216
  • [10] The nanospect/ct: A high-sensitivity small-animal spect/ct with submillimeter spatial resolution
    Schramm, N. U.
    Lackas, C.
    Hoppin, J. W.
    Forrer, F.
    de Jong, M.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2006, 33 : S117 - S117