Effect of patient positioning on carbon-ion therapy planned dose distribution to pancreatic tumors and organs at risk

被引:8
|
作者
Miki, Kentaro [1 ]
Fukahori, Mai [1 ]
Kumagai, Motoki [2 ]
Yamada, Shigeru [1 ]
Mori, Shinichiro [1 ]
机构
[1] Natl Inst Radiol Sci, Res Ctr Charged Particle Therapy, Inage Ku, 4-9-1 Anagawa, Chiba 2638555, Japan
[2] Natl Canc Ctr East, Res Ctr Innovat Oncol, Particle Therapy Div, 6-5-1 Kashiwa, Chiba 2778577, Japan
关键词
Carbon ion radiotherapy; Dose distribution; Patient positioning; Pancreatic cancer; DOSIMETRIC FEASIBILITY; NORMAL TISSUE; RADIOTHERAPY; IRRADIATION; TRIAL; PRONE;
D O I
10.1016/j.ejmp.2016.12.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Pancreatic tumor treatment dose distribution variations associated with supine and prone patient positioning were evaluated. Methods: A total of 33 patients with pancreatic tumors who underwent CT in the supine and prone positions were analyzed retrospectively. Gross tumor volume (GTV), planning target volume (PTV), and organs at risk (OARs) (duodenum and stomach) were contoured. The prescribed dose of 55.2 Gy (RBE) was planned from four beam angles (0 degrees, 90 degrees, 180 degrees, and 270 degrees). Patient collimator and compensating boli were designed for each field. Dose distributions were calculated for each field in the supine and prone positions. To improve dose distribution, patient positioning was selected from supine or prone for each beam field. Results: Compared with conventional beam angle and patient positioning, D2cc of 1st- 2nd portion of duodenum (D1-D2), 3rd-4th portion of duodenum (D3-D4), and stomach could be reduced to a maximum of 6.4 Gy (RBE), 3.5 Gy (RBE), and 4.5 Gy (RBE) by selection of patient positioning. V10 of D1-D2, D3-D4, and stomach could be reduced to a maximum of 7.2 cc, 11.3 cc, and 11.5 cc, respectively. D95 of GTV and PTV were improved to a maximum of 6.9% and 3.7% of the prescribed dose, respectively. Conclusions: Optimization of patient positioning for each beam angle in treatment planning has the potential to reduce OARs dose maintaining tumor dose in pancreatic treatment. (C) 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:38 / 46
页数:9
相关论文
共 50 条
  • [1] Scintillator screen for measuring dose distribution in scanned carbon-ion therapy
    Yogo, Katsunori
    Tatsuno, Yuya
    Souda, Hikaru
    Matsumura, Akihiko
    Tsuneda, Masato
    Hirano, Yoshiyuki
    Ishiyama, Hiromichi
    Saito, Akito
    Ozawa, Shuichi
    Nagata, Yasushi
    Nakano, Takashi
    Hayakawa, Kazushige
    Kanai, Tatsuaki
    RADIATION MEASUREMENTS, 2019, 129
  • [2] The Immunomodulatory Effects of Carbon-Ion Therapy on Solid Tumors
    Spina, C. S.
    Tsuruoka, C.
    Mao, W.
    Sunaoshi, M.
    Chaimowitz, M.
    Shang, Y.
    Kakinuma, S.
    Drake, C. G.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : S124 - S124
  • [3] Gated carbon-ion scanning treatment for pancreatic tumour with field specific target volume and organs at risk
    Miki, Kentaro
    Mori, Shinichiro
    Shiomi, Miho
    Yamada, Shigeru
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (12): : 1521 - 1528
  • [4] Carbon-Ion Therapy: One More Step in the Endless Quest for the Ideal Dose Distribution
    Schulz, Robert J.
    Kagan, A. Robert
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 95 (01): : 561 - 561
  • [5] A randomized study of the effect of patient positioning on setup reproducibility and dose distribution to organs at risk in radiotherapy of rectal cancer patients
    Trude C. Frøseth
    Trond Strickert
    Kjersti S. Solli
    Øyvind Salvesen
    Gunilla Frykholm
    Randi J. Reidunsdatter
    Radiation Oncology, 10
  • [6] A randomized study of the effect of patient positioning on setup reproducibility and dose distribution to organs at risk in radiotherapy of rectal cancer patients
    Froseth, Trude C.
    Strickert, Trond
    Solli, Kjersti S.
    Salvesen, Oyvind
    Frykholm, Gunilla
    Reidunsdatter, Randi J.
    RADIATION ONCOLOGY, 2015, 10
  • [7] Immunomodulation of solid tumors by carbon-ion therapy versus conventional photon therapy
    Spina, Catherine
    Tsuruoka, Chizuru
    Mao, Wendy
    Sunaoski, Masaaki
    Chaimowitz, Mathew
    Shang, Yi
    Kakinuma, Shizuko
    Drake, Charles
    Spina, Catherine
    JOURNAL FOR IMMUNOTHERAPY OF CANCER, 2019, 7
  • [8] dosimetric evaluation of carbon-ion beam grid therapy of brain tumors
    Tsubouchi, T.
    Valdman, A.
    Siegbahn, A.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1202 - S1203
  • [9] A phantom to simulate organ motion and its effect on dose distribution in carbon ion therapy for pancreatic cancer
    Stengl, Christina
    Panow, Kathrin
    Arbes, Eric
    Munoz, Ivan D.
    Christensen, Jeppe B.
    Neelsen, Christian
    Dinkel, Fabian
    Weidner, Artur
    Runz, Armin
    Johnen, Wibke
    Liermann, Jakob
    Echner, Gernot
    Vedelago, Jose
    Jaekel, Oliver
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (24):
  • [10] Carbon Ion dose constraints for organs at risk at MedAustron
    Fossati, P.
    Perpar, A.
    Georg, P.
    Hug, E.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2020, 196 (03) : 307 - 307