CLS HXMA 2.0 T Superconducting Wiggler Upgrades

被引:2
|
作者
Lin, Linda [1 ]
Chen, Ning [1 ]
Khrushchev, Sergey [2 ]
机构
[1] CLSI, Saskatoon, SK, Canada
[2] BINP, Novosibirsk, Russia
基金
加拿大自然科学与工程研究理事会;
关键词
D O I
10.1063/1.5084569
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The 2.0 T superconducting (SC) Wiggler on the CLSI Hard X-ray Micro Analysis (HXMA) beamline [1-2] ran at a field of 1.9 Tesla during normal operation, lower than the design specification. After ten years of operation, CLSI experienced significant downtime and wiggler magnetic field decreasing due to frequent breakdowns at the cryogenic system [3]. There were total of three emergency repairs at the cryo-cooler and cryo-compressor in 2015. The magnetic field at the Wiggler had to reduce from its normal operation field of 1.9 Tesla to 1.7 Tesla to maintain the operation of the storage ring and HXMA beamline during most time in 2016 due to one of the 4.2K cryo-cooler reaching its limit temperature. The cryogenic system was not reliable and became a risk of extended beamline shutdown. The operating cost of the HXMA wiggler had been very high due to frequent cryo-cooler failures and high liquid helium consumption level. Wiggler upgrades were installed in spring, 2017 including a replacement of the entire cryogenic system from Leybold to Sumitomo [3]. The associated components were also replaced by Budker Institute of Nuclear Physics (BINP) with a new design at cold fingers (gilded pins) and design modifications at the current lead assemblies, the shield cryo-cooler housings and the central throat assembly. The operation temperatures and pressure inside the wiggler cryostat have reduced dramatically after this replacement and modifications. There is no liquid helium consumption and the wiggler has been running at a higher magnetic field of 2.2 Tesla compared to the initial design specification of 2.0 Tesla since this upgrade. The HXMA beamline performance has improved in overall at operation reliability, stability, flex and beam quality.
引用
收藏
页数:6
相关论文
共 50 条
  • [41] 5 NEW EXPERIMENTAL STATIONS AT THE SRS DARESBURY FROM A 6-T SUPERCONDUCTING WIGGLER MAGNET
    BILSBORROW, RL
    BLISS, N
    BORDAS, J
    CERNIK, RJ
    CLARK, GF
    CLARK, SM
    COLLINS, SP
    DOBSON, BR
    FELL, BD
    GRANT, AF
    HARRIS, NW
    SMITH, W
    TOWNSANDREWS, E
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02): : 1633 - 1635
  • [42] Construction and commissioning of a superconducting wiggler for the UVSOR storage ring
    Nakamura, E
    Hama, H
    Yamazaki, J
    Kinoshita, T
    Matsudo, O
    Yonehara, H
    Kasuga, T
    Isoyama, G
    Watanabe, M
    JOURNAL OF ELECTRON SPECTROSCOPY AND RELATED PHENOMENA, 1996, 80 : 421 - 424
  • [43] Effects of a 4.3T superconducting wiggler on the electron beam optics and dynamics at NSLS-II
    Smaluk, V.
    Li, Y.
    Hidaka, Y.
    Wang, G.
    Tanabe, T.
    Hidas, D.
    Yang, X.
    Bassi, G.
    Khan, A.
    Bacha, B.
    Todd, R.
    Ha, K.
    Tian, Y.
    Shaftan, T.
    Hetzel, C.
    Blednykh, A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2023, 1057
  • [44] X-ray beamlines at the superconducting wiggler of NSRL
    Xu, CY
    Pan, GQ
    Xu, FQ
    Wei, SQ
    Tian, TC
    Zhao, DM
    Wang, F
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 467 : 639 - 642
  • [45] Beam dynamics of the superconducting wiggler on the SSRF storage ring
    张庆磊
    田顺强
    姜伯承
    许皆平
    赵振堂
    Chinese Physics C, 2016, (03) : 130 - 133
  • [46] THE SUPERCONDUCTING WIGGLER AT THE NSLS - SYSTEMS DESCRIPTION AND INITIAL EXPERIENCE
    DECKER, G
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 291 (1-2): : 357 - 362
  • [47] Design of a. superconducting multipole wiggler for synchrotron radiation
    Hwang, CS
    Wang, B
    Chen, JY
    Wahrer, R
    Chang, CH
    Fan, TC
    Lin, FY
    Chen, HH
    Huang, MH
    Chen, CT
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2003, 13 (02) : 1209 - 1212
  • [48] Beam dynamics of the superconducting wiggler on the SSRF storage ring
    Zhang, Qing-Lei
    Tian, Shun-Qiang
    Jiang, Bo-Cheng
    Xu, Jie-Ping
    Zhao, Zhen-Tang
    CHINESE PHYSICS C, 2016, 40 (03)
  • [49] Beam dynamics of the superconducting wiggler on the SSRF storage ring
    张庆磊
    田顺强
    姜伯承
    许皆平
    赵振堂
    Chinese Physics C, 2016, 40 (03) : 130 - 133
  • [50] Performance of a high-temperature superconducting probe for in vivo microscopy at 2.0 T
    Miller, JR
    Hurlston, SE
    Ma, QY
    Face, DW
    Kountz, DJ
    MacFall, JR
    Hedlund, LW
    Johnson, GA
    MAGNETIC RESONANCE IN MEDICINE, 1999, 41 (01) : 72 - 79